Tear meniscus height comparison between AS-OCT and Oculus Keratograph® K5M.

Angeline Lim Pei Yik, Fakhruddin Shamsheer Barodawala
{"title":"Tear meniscus height comparison between AS-OCT and Oculus Keratograph® K5M.","authors":"Angeline Lim Pei Yik, Fakhruddin Shamsheer Barodawala","doi":"10.22336/rjo.2024.72","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To evaluate the agreement between Spectral-Domain Optical Coherence Tomography (AS-OCT) and Oculus Keratograph® 5M in measuring tear meniscus height (TMH) and to assess the impact of contact lens wear on these measurements. <b>Materials and methods</b>: TMH was measured in 54 healthy eyes using AS-OCT (3D OCT-1 Maestro, Topcon, Tokyo, Japan) and Oculus Keratograph® 5M (OCULUS Optikgeräte, Wetzlar, Germany), with and without contact lens wear. Bland-Altman analysis was used to assess agreement between the two devices. Delefilcon A, water-gradient, daily disposable contact lenses were used, and measurements were carried out after a 20-minute adaptation period.</p><p><strong>Results: </strong>The means of TMH without the contact lenses were 0.21 ± 0.06 mm and 0.20 ± 0.05 mm obtained from AS-OCT and Oculus K5M, respectively, and these measurements were not statistically significant (t (53) = 0.99, p = 0.33). No significant differences were observed in TMH compared to contact lenses (t (53) = 1.52, p = 0.13). Agreement between measurements obtained by both the instruments was assessed using Bland-Altman analysis. The limits of agreement were within clinically acceptable ranges (0.10 mm - 0.15 mm), with no evidence of significant bias (t = -0.32, r = 0.22). The results obtained with contact lenses were also not statistically significant (t (53) = 1.52, p < 0.05).</p><p><strong>Discussion: </strong>The present study compared tear meniscus height (TMH) measurements obtained from AS-OCT and Oculus K5M in subjects with and without contact lens wear. Both instruments showed good agreement, with AS-OCT consistently measuring slightly higher TMH values than Oculus K5M. The mean TMH values were similar to those of previous studies, indicating normal tear film in the subjects. Contact lens wear was found to reduce TMH slightly, but it returned to baseline after a short adaptation period. The Bland-Altman analysis confirmed good agreement between the two instruments, with most data points falling within the limits of agreement. These findings suggest that AS-OCT and Oculus K5M can be reliable tools for measuring TMH and can be used interchangeably for clinical practice.</p><p><strong>Conclusion: </strong>AS-OCT and Oculus Keratograph® 5M showed comparable results in measuring TMH, suggesting potential interchangeability in clinical practice. Further validation in broader clinical settings and diverse subject groups may be warranted.</p>","PeriodicalId":94355,"journal":{"name":"Romanian journal of ophthalmology","volume":"68 4","pages":"398-403"},"PeriodicalIF":0.0000,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11809819/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Romanian journal of ophthalmology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22336/rjo.2024.72","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: To evaluate the agreement between Spectral-Domain Optical Coherence Tomography (AS-OCT) and Oculus Keratograph® 5M in measuring tear meniscus height (TMH) and to assess the impact of contact lens wear on these measurements. Materials and methods: TMH was measured in 54 healthy eyes using AS-OCT (3D OCT-1 Maestro, Topcon, Tokyo, Japan) and Oculus Keratograph® 5M (OCULUS Optikgeräte, Wetzlar, Germany), with and without contact lens wear. Bland-Altman analysis was used to assess agreement between the two devices. Delefilcon A, water-gradient, daily disposable contact lenses were used, and measurements were carried out after a 20-minute adaptation period.

Results: The means of TMH without the contact lenses were 0.21 ± 0.06 mm and 0.20 ± 0.05 mm obtained from AS-OCT and Oculus K5M, respectively, and these measurements were not statistically significant (t (53) = 0.99, p = 0.33). No significant differences were observed in TMH compared to contact lenses (t (53) = 1.52, p = 0.13). Agreement between measurements obtained by both the instruments was assessed using Bland-Altman analysis. The limits of agreement were within clinically acceptable ranges (0.10 mm - 0.15 mm), with no evidence of significant bias (t = -0.32, r = 0.22). The results obtained with contact lenses were also not statistically significant (t (53) = 1.52, p < 0.05).

Discussion: The present study compared tear meniscus height (TMH) measurements obtained from AS-OCT and Oculus K5M in subjects with and without contact lens wear. Both instruments showed good agreement, with AS-OCT consistently measuring slightly higher TMH values than Oculus K5M. The mean TMH values were similar to those of previous studies, indicating normal tear film in the subjects. Contact lens wear was found to reduce TMH slightly, but it returned to baseline after a short adaptation period. The Bland-Altman analysis confirmed good agreement between the two instruments, with most data points falling within the limits of agreement. These findings suggest that AS-OCT and Oculus K5M can be reliable tools for measuring TMH and can be used interchangeably for clinical practice.

Conclusion: AS-OCT and Oculus Keratograph® 5M showed comparable results in measuring TMH, suggesting potential interchangeability in clinical practice. Further validation in broader clinical settings and diverse subject groups may be warranted.

求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信