XN-1000 Hematology Analyzer as an Alternative to Flow Cytometry for Measuring Residual Cells in Blood Components.

IF 4 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY
Anita Siller, Lisa Seekircher, Daniela Schmidt, Lena Tschiderer, Peter Willeit, Harald Schennach, Marco Amato
{"title":"XN-1000 Hematology Analyzer as an Alternative to Flow Cytometry for Measuring Residual Cells in Blood Components.","authors":"Anita Siller, Lisa Seekircher, Daniela Schmidt, Lena Tschiderer, Peter Willeit, Harald Schennach, Marco Amato","doi":"10.3343/alm.2024.0448","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Measuring residual cells in blood products is legally required for monitoring the manufacturing process and ensuring recipient safety. We compared the accuracy and performance of the two methodologies.</p><p><strong>Methods: </strong>Residual white blood cells (rWBCs), red blood cells (rRBCs), and platelets (rPLTs) were measured in RBC concentrates (rWBCs), fresh frozen plasma (rWBCs, rRBCs, and rPLTs), and PLT concentrates (rWBCs and rRBCs) using the Sysmex XN-1000 hematology analyzer (Sysmex, Kobe, Japan) equipped with Blood Bank mode and standard flow cytometry (fluorescence-activated cell sorting; FACS).</p><p><strong>Results: </strong>rWBC counts in RBC concentrates and plasma were similar between XN-1000 and FACS. In pooled pathogen-inactivated PLT concentrates, XN-1000 yielded higher rWBC counts. Correlations between XN-1000 and FACS were moderate for rWBCs (0.42, 95% confidence interval: 0.15-0.69) in RBC inline-filtrated WBC-depleted RBC concentrates. In plasma, correlations were high for rWBC, rRBC, and rPLT counts, with Spearman correlation coefficients of 0.82-0.97. In pathogen-inactivated PLT concentrates, correlations were moderate for rWBCs (0.58, 0.33-0.84) and rRBCs (0.61, 0.35-0.87) in pooled samples but not significant in apheresis-derived samples. Median differences between FACS and XN-1000 were generally low, but XN-1000 overestimated residual cell counts in a subset of measurements. Residual cell cut-off values were surpassed in ⟩90% of RBC concentrates, plasma, and apheresis pathogen-inactivated PLT concentrates using both methods. In pooled pathogen-inactivated PLT concentrates, 91.2% and 70.6% surpassed the cut-off using FACS and XN-1000, respectively.</p><p><strong>Conclusions: </strong>Sysmex XN-1000 is suitable for residual cell measurements in RBC concentrates and plasma, with some limitations for PLT concentrates.</p>","PeriodicalId":8421,"journal":{"name":"Annals of Laboratory Medicine","volume":" ","pages":""},"PeriodicalIF":4.0000,"publicationDate":"2025-02-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Laboratory Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3343/alm.2024.0448","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Measuring residual cells in blood products is legally required for monitoring the manufacturing process and ensuring recipient safety. We compared the accuracy and performance of the two methodologies.

Methods: Residual white blood cells (rWBCs), red blood cells (rRBCs), and platelets (rPLTs) were measured in RBC concentrates (rWBCs), fresh frozen plasma (rWBCs, rRBCs, and rPLTs), and PLT concentrates (rWBCs and rRBCs) using the Sysmex XN-1000 hematology analyzer (Sysmex, Kobe, Japan) equipped with Blood Bank mode and standard flow cytometry (fluorescence-activated cell sorting; FACS).

Results: rWBC counts in RBC concentrates and plasma were similar between XN-1000 and FACS. In pooled pathogen-inactivated PLT concentrates, XN-1000 yielded higher rWBC counts. Correlations between XN-1000 and FACS were moderate for rWBCs (0.42, 95% confidence interval: 0.15-0.69) in RBC inline-filtrated WBC-depleted RBC concentrates. In plasma, correlations were high for rWBC, rRBC, and rPLT counts, with Spearman correlation coefficients of 0.82-0.97. In pathogen-inactivated PLT concentrates, correlations were moderate for rWBCs (0.58, 0.33-0.84) and rRBCs (0.61, 0.35-0.87) in pooled samples but not significant in apheresis-derived samples. Median differences between FACS and XN-1000 were generally low, but XN-1000 overestimated residual cell counts in a subset of measurements. Residual cell cut-off values were surpassed in ⟩90% of RBC concentrates, plasma, and apheresis pathogen-inactivated PLT concentrates using both methods. In pooled pathogen-inactivated PLT concentrates, 91.2% and 70.6% surpassed the cut-off using FACS and XN-1000, respectively.

Conclusions: Sysmex XN-1000 is suitable for residual cell measurements in RBC concentrates and plasma, with some limitations for PLT concentrates.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Annals of Laboratory Medicine
Annals of Laboratory Medicine MEDICAL LABORATORY TECHNOLOGY-
CiteScore
8.30
自引率
12.20%
发文量
100
审稿时长
6-12 weeks
期刊介绍: Annals of Laboratory Medicine is the official journal of Korean Society for Laboratory Medicine. The journal title has been recently changed from the Korean Journal of Laboratory Medicine (ISSN, 1598-6535) from the January issue of 2012. The JCR 2017 Impact factor of Ann Lab Med was 1.916.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信