Tahereh Zare, Peyman Sheikhzadeh, Behnoosh Teimourian Fard, Pardis Ghafarian, Mohammad Reza Ay
{"title":"Design and Performance Evaluation of SiPM-based High-resolution Dedicated Brain Positron Emission Tomography Scanner: A Simulation Study.","authors":"Tahereh Zare, Peyman Sheikhzadeh, Behnoosh Teimourian Fard, Pardis Ghafarian, Mohammad Reza Ay","doi":"10.4103/jmp.jmp_134_24","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose/aim: </strong>The increasing population age highlights the critical need for early brain disease diagnosis, especially in disorders such as dementia. Consequently, a notable focus has been on developing dedicated brain positron emission tomography (PET) scanners, which offer higher resolution and sensitivity than whole-body PET scanners. This study aims to design and performance evaluation of an LYSO-based dedicated brain PET scanner.</p><p><strong>Materials and methods: </strong>We developed a dedicated brain PET using Monte Carlo simulation based on cylindrical geometry. Each detector block consisted of a 23 × 23 array of 2 mm × 2 mm × 15 mm LYSO crystals coupled with SiPM. The performance of this scanner was evaluated based on the NEMA NU-2-2018 standard, focusing on analyzing various energy windows and coincidence time windows (CTWs).</p><p><strong>Results: </strong>The results demonstrated that the noise equivalent count rate (NECR) peaked at each CTW in the 408-680 keV energy window. In addition, increasing the CTWs from 3 ns to 10 ns resulted in a decrease of 9% in sensitivity and an increase of 63% in NECR. Furthermore, the study findings highlight that using a time-of-flight (TOF) resolution of 250 ps can substantially improve image contrast relative to non-TOF reconstruction.</p><p><strong>Conclusions: </strong>We conclude that employing a broader energy window and a narrower CTW can significantly enhance the scanner's performance regarding sensitivity and NECR. Furthermore, incorporating LYSO pixelated crystals with TOF information will facilitate the generation of high-resolution and high-contrast images.</p>","PeriodicalId":51719,"journal":{"name":"Journal of Medical Physics","volume":"49 4","pages":"631-641"},"PeriodicalIF":0.7000,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11801085/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jmp.jmp_134_24","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/12/18 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose/aim: The increasing population age highlights the critical need for early brain disease diagnosis, especially in disorders such as dementia. Consequently, a notable focus has been on developing dedicated brain positron emission tomography (PET) scanners, which offer higher resolution and sensitivity than whole-body PET scanners. This study aims to design and performance evaluation of an LYSO-based dedicated brain PET scanner.
Materials and methods: We developed a dedicated brain PET using Monte Carlo simulation based on cylindrical geometry. Each detector block consisted of a 23 × 23 array of 2 mm × 2 mm × 15 mm LYSO crystals coupled with SiPM. The performance of this scanner was evaluated based on the NEMA NU-2-2018 standard, focusing on analyzing various energy windows and coincidence time windows (CTWs).
Results: The results demonstrated that the noise equivalent count rate (NECR) peaked at each CTW in the 408-680 keV energy window. In addition, increasing the CTWs from 3 ns to 10 ns resulted in a decrease of 9% in sensitivity and an increase of 63% in NECR. Furthermore, the study findings highlight that using a time-of-flight (TOF) resolution of 250 ps can substantially improve image contrast relative to non-TOF reconstruction.
Conclusions: We conclude that employing a broader energy window and a narrower CTW can significantly enhance the scanner's performance regarding sensitivity and NECR. Furthermore, incorporating LYSO pixelated crystals with TOF information will facilitate the generation of high-resolution and high-contrast images.
期刊介绍:
JOURNAL OF MEDICAL PHYSICS is the official journal of Association of Medical Physicists of India (AMPI). The association has been bringing out a quarterly publication since 1976. Till the end of 1993, it was known as Medical Physics Bulletin, which then became Journal of Medical Physics. The main objective of the Journal is to serve as a vehicle of communication to highlight all aspects of the practice of medical radiation physics. The areas covered include all aspects of the application of radiation physics to biological sciences, radiotherapy, radiodiagnosis, nuclear medicine, dosimetry and radiation protection. Papers / manuscripts dealing with the aspects of physics related to cancer therapy / radiobiology also fall within the scope of the journal.