{"title":"Backside Irradiation of Ultraviolet-A for Correcting Nonuniformity Error of Gafchromic XR-QA2 Films.","authors":"Nobuyoshi Tanki, Sachiko Goto, Toshizo Katsuda, Rumi Gotanda, Tatsuhiro Gotanda, Tadao Kuwano","doi":"10.4103/jmp.jmp_87_24","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Radiochromic film is used for quality assurance and quality control of X-ray equipment in the diagnostic radiology. In addition, three-dimensional dose distribution of computed tomography (CT) is measured. To correct the nonuniformity and uncertainty of radiochromic films for dose measurement of CT, the films are preirradiated ultraviolet (UV)-A rays. There is a difference in the UV protection strength of radiochromic films. A concern exists about the effects of the UV-A irradiation intensity. We thus irradiated with UV-A rays from the backsides of the films to assess if backside irradiation was possible.</p><p><strong>Materials and methods: </strong>Gafchromic XR-QA2 and RTQA2 were used in this study. The UV-A rays were simultaneously irradiated on the front and backsides of each film for 12 h. The yellow layer of each film was scanned and imaged. The average pixel values ± standard deviations (SDs) were compared. In the statistical analysis, a paired t-test was performed. To compare, the active-layer densities engendered by the UV-A rays. Calibration curve was created with 48 h of preirradiation of UV-A.</p><p><strong>Results: </strong>The mean pixel values ± SD for Gafchromic XR-QA2 on the front and backsides were 130.776 ± 0.812 and 81.015 ± 1.128, respectively. On the other hand, the mean pixel values ± SD for Gafchromic RTQA2 on the front and backsides were 62.299 ± 1.077 and 133.761 ± 1.365, respectively. The statistical results of the paired t-test were significantly different (P < 0.01) between both films. Fitting equation of the calibration curve is shown below. y = -390.47 ± 200 + (443.45 ± 10x80).5068 ± 0.0434.</p><p><strong>Conclusion: </strong>Based on the relationship between the sensitivity of the active layer to UV-A rays and the strength of UV protection on the surface, we concluded that backside irradiation is recommended for Gafchromic XR-QA2, and frontside irradiation is recommended for Gafchromic RTQA2.</p>","PeriodicalId":51719,"journal":{"name":"Journal of Medical Physics","volume":"49 4","pages":"563-567"},"PeriodicalIF":0.7000,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11801077/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jmp.jmp_87_24","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/12/18 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Radiochromic film is used for quality assurance and quality control of X-ray equipment in the diagnostic radiology. In addition, three-dimensional dose distribution of computed tomography (CT) is measured. To correct the nonuniformity and uncertainty of radiochromic films for dose measurement of CT, the films are preirradiated ultraviolet (UV)-A rays. There is a difference in the UV protection strength of radiochromic films. A concern exists about the effects of the UV-A irradiation intensity. We thus irradiated with UV-A rays from the backsides of the films to assess if backside irradiation was possible.
Materials and methods: Gafchromic XR-QA2 and RTQA2 were used in this study. The UV-A rays were simultaneously irradiated on the front and backsides of each film for 12 h. The yellow layer of each film was scanned and imaged. The average pixel values ± standard deviations (SDs) were compared. In the statistical analysis, a paired t-test was performed. To compare, the active-layer densities engendered by the UV-A rays. Calibration curve was created with 48 h of preirradiation of UV-A.
Results: The mean pixel values ± SD for Gafchromic XR-QA2 on the front and backsides were 130.776 ± 0.812 and 81.015 ± 1.128, respectively. On the other hand, the mean pixel values ± SD for Gafchromic RTQA2 on the front and backsides were 62.299 ± 1.077 and 133.761 ± 1.365, respectively. The statistical results of the paired t-test were significantly different (P < 0.01) between both films. Fitting equation of the calibration curve is shown below. y = -390.47 ± 200 + (443.45 ± 10x80).5068 ± 0.0434.
Conclusion: Based on the relationship between the sensitivity of the active layer to UV-A rays and the strength of UV protection on the surface, we concluded that backside irradiation is recommended for Gafchromic XR-QA2, and frontside irradiation is recommended for Gafchromic RTQA2.
期刊介绍:
JOURNAL OF MEDICAL PHYSICS is the official journal of Association of Medical Physicists of India (AMPI). The association has been bringing out a quarterly publication since 1976. Till the end of 1993, it was known as Medical Physics Bulletin, which then became Journal of Medical Physics. The main objective of the Journal is to serve as a vehicle of communication to highlight all aspects of the practice of medical radiation physics. The areas covered include all aspects of the application of radiation physics to biological sciences, radiotherapy, radiodiagnosis, nuclear medicine, dosimetry and radiation protection. Papers / manuscripts dealing with the aspects of physics related to cancer therapy / radiobiology also fall within the scope of the journal.