Accuracy of the smartphone blood pressure measurement solution OptiBP to track blood pressure changes in pregnant women.

IF 3.3 2区 医学 Q1 PERIPHERAL VASCULAR DISEASE
Pedro Almeida, Alexia Cuénoud, Harry Hoang, Alexandra Othenin-Girard, Nadia Salhi, Andreas Köthe, Urvan Christen, Patrick Schoettker
{"title":"Accuracy of the smartphone blood pressure measurement solution OptiBP to track blood pressure changes in pregnant women.","authors":"Pedro Almeida, Alexia Cuénoud, Harry Hoang, Alexandra Othenin-Girard, Nadia Salhi, Andreas Köthe, Urvan Christen, Patrick Schoettker","doi":"10.1097/HJH.0000000000003956","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Hypertensive disorders present significant morbidity and mortality during pregnancy. Although ambulatory blood pressure measurement remains the standard of care for normotensive women, self-monitoring at home is increasingly prevalent. The widespread use of smartphones worldwide has sparked interest in mobile applications that leverage the built-in hardware for blood pressure estimation, yet few trials have assessed their accuracy.</p><p><strong>Methods: </strong>This prospective, longitudinal and monocentric study evaluated the accuracy of the OptiBP algorithm against standard oscillometric blood pressure measurements in a sample of pregnant women. Patients scheduled for elective caesarean sections were enrolled during the preoperative anesthesia consultations. Paired blood pressure measurements using OptiBP and the reference method were obtained at multiple time-points in late pregnancy and the postpartum period. Agreement between methods was assessed using the AAMI/ESH/ISO 81060-2:2018 standard thresholds of 5 ± 8 mmHg for mean ± standard deviation of the error (criterion 1) and patient-specific standard deviation of the mean error (criterion 2) and represented graphically by Bland-Altman scatterplots.</p><p><strong>Results: </strong>Forty-eight women were enrolled of which 32 completed the protocol, yielding 338 total valid measurement pairs. Mean and standard deviation of the error were -1.78 ± 7.94 and 1.19 ± 7.59, and the patient-specific standard deviation of the mean error was 4.68 and 4.52, for SBP and DBP, respectively.</p><p><strong>Conclusion: </strong>Compared with blood pressure measurements taken with an oscillometric device, OptiBP's blood pressure estimates meet the AAMI/ESH/ISO 81060-2:2018 criteria.</p>","PeriodicalId":16043,"journal":{"name":"Journal of Hypertension","volume":" ","pages":""},"PeriodicalIF":3.3000,"publicationDate":"2025-02-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Hypertension","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/HJH.0000000000003956","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Hypertensive disorders present significant morbidity and mortality during pregnancy. Although ambulatory blood pressure measurement remains the standard of care for normotensive women, self-monitoring at home is increasingly prevalent. The widespread use of smartphones worldwide has sparked interest in mobile applications that leverage the built-in hardware for blood pressure estimation, yet few trials have assessed their accuracy.

Methods: This prospective, longitudinal and monocentric study evaluated the accuracy of the OptiBP algorithm against standard oscillometric blood pressure measurements in a sample of pregnant women. Patients scheduled for elective caesarean sections were enrolled during the preoperative anesthesia consultations. Paired blood pressure measurements using OptiBP and the reference method were obtained at multiple time-points in late pregnancy and the postpartum period. Agreement between methods was assessed using the AAMI/ESH/ISO 81060-2:2018 standard thresholds of 5 ± 8 mmHg for mean ± standard deviation of the error (criterion 1) and patient-specific standard deviation of the mean error (criterion 2) and represented graphically by Bland-Altman scatterplots.

Results: Forty-eight women were enrolled of which 32 completed the protocol, yielding 338 total valid measurement pairs. Mean and standard deviation of the error were -1.78 ± 7.94 and 1.19 ± 7.59, and the patient-specific standard deviation of the mean error was 4.68 and 4.52, for SBP and DBP, respectively.

Conclusion: Compared with blood pressure measurements taken with an oscillometric device, OptiBP's blood pressure estimates meet the AAMI/ESH/ISO 81060-2:2018 criteria.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Hypertension
Journal of Hypertension 医学-外周血管病
CiteScore
7.90
自引率
6.10%
发文量
1389
审稿时长
3 months
期刊介绍: The Journal of Hypertension publishes papers reporting original clinical and experimental research which are of a high standard and which contribute to the advancement of knowledge in the field of hypertension. The Journal publishes full papers, reviews or editorials (normally by invitation), and correspondence.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信