First Dimension Trap-and-Elute Combined with Second Dimension Reversed-Phase Liquid Chromatography Separation Using a Two-Dimensional-Liquid Chromatography-Tandem Mass Spectrometry System for Sensitive Quantification of Human Insulin and Six Insulin Analogs in Plasma: Improved Chromatographic Resolution and Stability Testing

IF 2.8 3区 工程技术 Q2 CHEMISTRY, ANALYTICAL
Pavel Sistik, Romana Urinovska, Klara Handlosova, Petr Handlos, Katerina Andelova, Jan Jurica, David Stejskal
{"title":"First Dimension Trap-and-Elute Combined with Second Dimension Reversed-Phase Liquid Chromatography Separation Using a Two-Dimensional-Liquid Chromatography-Tandem Mass Spectrometry System for Sensitive Quantification of Human Insulin and Six Insulin Analogs in Plasma: Improved Chromatographic Resolution and Stability Testing","authors":"Pavel Sistik,&nbsp;Romana Urinovska,&nbsp;Klara Handlosova,&nbsp;Petr Handlos,&nbsp;Katerina Andelova,&nbsp;Jan Jurica,&nbsp;David Stejskal","doi":"10.1002/jssc.70092","DOIUrl":null,"url":null,"abstract":"<p>Multidimensional chromatography coupled to tandem mass spectrometry (MS/MS), including simple sample preparation with protein precipitation, anion conversion with ammonium hydroxide, and solid-phase extraction using mixed-mode anion exchange in a 96-well plate format, has been validated for rapid simultaneous analysis of human insulin and its six analogs (lispro, glulisine, glargine, degludec, detemir, and aspart) in human plasma. This method is critical for clinical diagnostics, forensic investigations, and anti-doping efforts due to the widespread use of these substances. In the present study, improved chromatographic resolution was achieved using a first-dimension trap-and-elute configuration with an XBridge C18 (2.1 × 20 mm, 3.5 µm) trap column combined with second dimension separation on a Cortecs Ultra-High-Performance Liquid Chromatography (UHPLC) C18+ (2.1 × 100 mm, 1.6 µm) analytical column implemented within a two-dimensional-LC-MS/MS system. The total chromatographic run time was 11 min. This setup increases both the resolution and sensitivity of the method. A mobile phase consisting of 0.8% formic acid (FA) in water and 0.7% FA in acetonitrile was used for gradient elution. Bovine insulin was used as the internal standard. MS detection was performed in positive electrospray ionization mode, and the ion suppression due to matrix effects was evaluated. Validation criteria included linearity, precision, accuracy, recovery, lower limit of quantitation, matrix effect, and stability tests with and without protease inhibitor cocktail under different conditions (short-term stability, long-term stability, and freeze-thaw stability). The concentration range for all insulins was 50–15 000 pg/mL, with limits of quantification below the therapeutic reference range for all analytes. Intra-run precision ranged from 1.1% to 5.7%, inter-run precision from 0.7% to 5.9%, and overall recovery from 96.9% to 114.3%. The validated method has been implemented successfully by the Department of Forensic Medicine at our hospital for the investigation of unexplained deaths.</p>","PeriodicalId":17098,"journal":{"name":"Journal of separation science","volume":"48 2","pages":""},"PeriodicalIF":2.8000,"publicationDate":"2025-02-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/jssc.70092","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of separation science","FirstCategoryId":"5","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/jssc.70092","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Multidimensional chromatography coupled to tandem mass spectrometry (MS/MS), including simple sample preparation with protein precipitation, anion conversion with ammonium hydroxide, and solid-phase extraction using mixed-mode anion exchange in a 96-well plate format, has been validated for rapid simultaneous analysis of human insulin and its six analogs (lispro, glulisine, glargine, degludec, detemir, and aspart) in human plasma. This method is critical for clinical diagnostics, forensic investigations, and anti-doping efforts due to the widespread use of these substances. In the present study, improved chromatographic resolution was achieved using a first-dimension trap-and-elute configuration with an XBridge C18 (2.1 × 20 mm, 3.5 µm) trap column combined with second dimension separation on a Cortecs Ultra-High-Performance Liquid Chromatography (UHPLC) C18+ (2.1 × 100 mm, 1.6 µm) analytical column implemented within a two-dimensional-LC-MS/MS system. The total chromatographic run time was 11 min. This setup increases both the resolution and sensitivity of the method. A mobile phase consisting of 0.8% formic acid (FA) in water and 0.7% FA in acetonitrile was used for gradient elution. Bovine insulin was used as the internal standard. MS detection was performed in positive electrospray ionization mode, and the ion suppression due to matrix effects was evaluated. Validation criteria included linearity, precision, accuracy, recovery, lower limit of quantitation, matrix effect, and stability tests with and without protease inhibitor cocktail under different conditions (short-term stability, long-term stability, and freeze-thaw stability). The concentration range for all insulins was 50–15 000 pg/mL, with limits of quantification below the therapeutic reference range for all analytes. Intra-run precision ranged from 1.1% to 5.7%, inter-run precision from 0.7% to 5.9%, and overall recovery from 96.9% to 114.3%. The validated method has been implemented successfully by the Department of Forensic Medicine at our hospital for the investigation of unexplained deaths.

Abstract Image

求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of separation science
Journal of separation science 化学-分析化学
CiteScore
6.30
自引率
16.10%
发文量
408
审稿时长
1.8 months
期刊介绍: The Journal of Separation Science (JSS) is the most comprehensive source in separation science, since it covers all areas of chromatographic and electrophoretic separation methods in theory and practice, both in the analytical and in the preparative mode, solid phase extraction, sample preparation, and related techniques. Manuscripts on methodological or instrumental developments, including detection aspects, in particular mass spectrometry, as well as on innovative applications will also be published. Manuscripts on hyphenation, automation, and miniaturization are particularly welcome. Pre- and post-separation facets of a total analysis may be covered as well as the underlying logic of the development or application of a method.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信