Comparison of dosimetric variability in the cervical spine between carbon fiber and titanium instrumentation.

IF 2.9 2区 医学 Q2 CLINICAL NEUROLOGY
Ahmed Mansi, Elhaum G Rezaii, Shuo Wang, Chi Zhang, Miki Katzir
{"title":"Comparison of dosimetric variability in the cervical spine between carbon fiber and titanium instrumentation.","authors":"Ahmed Mansi, Elhaum G Rezaii, Shuo Wang, Chi Zhang, Miki Katzir","doi":"10.3171/2024.10.SPINE24584","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Carbon fiber instrumentation has gained popularity in spine oncology for its radiographic advantage of reduced artifact on imaging. With its increased use in patients who undergo radiation therapy, the dosimetric accuracy of postoperative radiation with carbon fiber constructs compared to classic titanium instrumentation becomes an important question. The purpose of this study was to compare the dosimetric accuracy of postoperative radiation in carbon fiber-instrumented versus titanium-instrumented cadaveric cervical spines after corpectomy.</p><p><strong>Methods: </strong>Two cadaveric specimens underwent two-level corpectomy using either titanium or carbon fiber instrumentation. Dosimeter chips were placed circumferentially around the constructs to calculate the dose of radiation to surrounding areas. The cadavers underwent one round of radiation with their respective constructs, and the dose of radiation was calculated and compared to the measured dose in each chip. After the first round of radiation, the instrumentation was switched between cadavers and the radiation therapy was repeated. The difference between the calculated and measured dose in carbon fiber versus that in titanium instrumentation in each cadaveric model was subsequently compared.</p><p><strong>Results: </strong>There was a significant difference in the dosimetry calculated at the area of the spinal cord dorsal to the corpectomy cage, with 68% less variability between the calculated and measured dose in the carbon fiber construct compared to that of the titanium construct. The mean variation of the measured dose at the spinal cord was 7.73% in titanium versus 4.6% in carbon fiber (p = 0.024, 1-tail; p = 0.048, 2-tail). There was also 30% less variability between the measured and calculated dose in the carbon fiber construct at the dosimeter chips lateral to the spinal cord, with a mean variation in the carbon fiber-instrumented cadaver of 4.94% compared to 6.45% with titanium (p = 0.01, 1-tail; p = 0.02, 2-tail). When all the dosimeters were combined without the control group, there was a statistically significant 14.5% difference in the mean variation of the calculated versus measured doses between the carbon fiber-instrumented cadavers versus titanium-instrumented cadavers (4.8% vs 5.5%, respectively; p = 0.015, 1-tail; p = 0.03, 2-tail).</p><p><strong>Conclusions: </strong>There is less variation in the calculated versus measured doses of radiation therapy in carbon fiber instrumentation compared to titanium instrumentation. This suggests improved radiation therapy delivering accuracy and complication avoidance in carbon fiber-instrumented spines.</p>","PeriodicalId":16562,"journal":{"name":"Journal of neurosurgery. Spine","volume":" ","pages":"1-9"},"PeriodicalIF":2.9000,"publicationDate":"2025-02-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of neurosurgery. Spine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3171/2024.10.SPINE24584","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: Carbon fiber instrumentation has gained popularity in spine oncology for its radiographic advantage of reduced artifact on imaging. With its increased use in patients who undergo radiation therapy, the dosimetric accuracy of postoperative radiation with carbon fiber constructs compared to classic titanium instrumentation becomes an important question. The purpose of this study was to compare the dosimetric accuracy of postoperative radiation in carbon fiber-instrumented versus titanium-instrumented cadaveric cervical spines after corpectomy.

Methods: Two cadaveric specimens underwent two-level corpectomy using either titanium or carbon fiber instrumentation. Dosimeter chips were placed circumferentially around the constructs to calculate the dose of radiation to surrounding areas. The cadavers underwent one round of radiation with their respective constructs, and the dose of radiation was calculated and compared to the measured dose in each chip. After the first round of radiation, the instrumentation was switched between cadavers and the radiation therapy was repeated. The difference between the calculated and measured dose in carbon fiber versus that in titanium instrumentation in each cadaveric model was subsequently compared.

Results: There was a significant difference in the dosimetry calculated at the area of the spinal cord dorsal to the corpectomy cage, with 68% less variability between the calculated and measured dose in the carbon fiber construct compared to that of the titanium construct. The mean variation of the measured dose at the spinal cord was 7.73% in titanium versus 4.6% in carbon fiber (p = 0.024, 1-tail; p = 0.048, 2-tail). There was also 30% less variability between the measured and calculated dose in the carbon fiber construct at the dosimeter chips lateral to the spinal cord, with a mean variation in the carbon fiber-instrumented cadaver of 4.94% compared to 6.45% with titanium (p = 0.01, 1-tail; p = 0.02, 2-tail). When all the dosimeters were combined without the control group, there was a statistically significant 14.5% difference in the mean variation of the calculated versus measured doses between the carbon fiber-instrumented cadavers versus titanium-instrumented cadavers (4.8% vs 5.5%, respectively; p = 0.015, 1-tail; p = 0.03, 2-tail).

Conclusions: There is less variation in the calculated versus measured doses of radiation therapy in carbon fiber instrumentation compared to titanium instrumentation. This suggests improved radiation therapy delivering accuracy and complication avoidance in carbon fiber-instrumented spines.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of neurosurgery. Spine
Journal of neurosurgery. Spine 医学-临床神经学
CiteScore
5.10
自引率
10.70%
发文量
396
审稿时长
6 months
期刊介绍: Primarily publish original works in neurosurgery but also include studies in clinical neurophysiology, organic neurology, ophthalmology, radiology, pathology, and molecular biology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信