{"title":"A novel double disc electrode excitation method for oil elemental analysis in rotating disc electrode-optical emission spectrometry (RDE-OES)†","authors":"Jinfan Huang, Xiao Fu, Cunjun Li, Guangyue Niu, Fa-Jie Duan, Xianlei Chen and Huadong Hao","doi":"10.1039/D4JA00148F","DOIUrl":null,"url":null,"abstract":"<p >Rotating disc electrode-optical emission spectrometry (RDE-OES) is a leading technology for detecting oil elements. However, electrode wear impedes sample combustion and spectral stability, causing errors in acquisition and analysis. In this work, we propose and develop a novel arc excitation method with a double disc electrode structure, replacing the conventional upper rod electrode with a disc electrode of the same type as the lower electrode, and the double electrodes rotate simultaneously. This method minimizes wear and maintains a constant gap, enabling a long-lasting, energy-focused arc and high-intensity, stable spectrum. Our method is validated by analyzing the spectral lines of Al, Mg, Cr and Ca. Compared with the conventional method, the results revealed that our method can improve calibration robustness, with a 2–3 times higher spectral intensity and 1–3 times higher signal-to-noise ratio (SNR) with an 8–10% reduction in the relative standard deviation (RSD) values. <em>R</em><small><sup>2</sup></small> for Al, Ca, Cr and Mg increased from 0.9872, 0.9772, 0.9824, and 0.9832 to 0.9988, 0.9994, 0.9998 and 0.9997, respectively, and the limits of detection (LODs) decreased from 2.0, 1.8, 1.7, and 1.9 ppm to 1.9, 1.6, 1.6 and 1.7 ppm. Our method achieved consistently high and stable spectral intensity and significantly reduced electrode wear during excitation, with the outer diameter change being 44.4% of the rod electrode length change after 50 excitations. The results demonstrate the method's ability for signal stability and accurate quantitative analysis, highlighting RDE-OES's importance in oil analysis.</p>","PeriodicalId":81,"journal":{"name":"Journal of Analytical Atomic Spectrometry","volume":" 2","pages":" 374-383"},"PeriodicalIF":3.1000,"publicationDate":"2025-01-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Analytical Atomic Spectrometry","FirstCategoryId":"92","ListUrlMain":"https://pubs.rsc.org/en/content/articlelanding/2025/ja/d4ja00148f","RegionNum":2,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Rotating disc electrode-optical emission spectrometry (RDE-OES) is a leading technology for detecting oil elements. However, electrode wear impedes sample combustion and spectral stability, causing errors in acquisition and analysis. In this work, we propose and develop a novel arc excitation method with a double disc electrode structure, replacing the conventional upper rod electrode with a disc electrode of the same type as the lower electrode, and the double electrodes rotate simultaneously. This method minimizes wear and maintains a constant gap, enabling a long-lasting, energy-focused arc and high-intensity, stable spectrum. Our method is validated by analyzing the spectral lines of Al, Mg, Cr and Ca. Compared with the conventional method, the results revealed that our method can improve calibration robustness, with a 2–3 times higher spectral intensity and 1–3 times higher signal-to-noise ratio (SNR) with an 8–10% reduction in the relative standard deviation (RSD) values. R2 for Al, Ca, Cr and Mg increased from 0.9872, 0.9772, 0.9824, and 0.9832 to 0.9988, 0.9994, 0.9998 and 0.9997, respectively, and the limits of detection (LODs) decreased from 2.0, 1.8, 1.7, and 1.9 ppm to 1.9, 1.6, 1.6 and 1.7 ppm. Our method achieved consistently high and stable spectral intensity and significantly reduced electrode wear during excitation, with the outer diameter change being 44.4% of the rod electrode length change after 50 excitations. The results demonstrate the method's ability for signal stability and accurate quantitative analysis, highlighting RDE-OES's importance in oil analysis.

Abstract Image

旋转盘电极-光学发射光谱(RDE-OES)† 中用于油元素分析的新型双盘电极激发方法
旋转圆盘电极-光学发射光谱法(RDE-OES)是一种领先的检测油元素的技术。然而,电极磨损阻碍了样品燃烧和光谱稳定性,导致采集和分析误差。在这项工作中,我们提出并开发了一种新的双圆盘电极结构的电弧激励方法,用与下电极相同类型的圆盘电极代替传统的上杆电极,双电极同时旋转。这种方法最大限度地减少了磨损,并保持了恒定的间隙,从而实现了持久的、能量集中的电弧和高强度、稳定的光谱。通过对Al、Mg、Cr和Ca的谱线分析,验证了该方法的稳健性。与传统方法相比,该方法的光谱强度提高了2-3倍,信噪比(SNR)提高了1-3倍,相对标准偏差(RSD)值降低了8-10%。Al、Ca、Cr和Mg的R2分别从0.9872、0.9772、0.9824和0.9832增加到0.9988、0.9994、0.9998和0.9997,检出限(lod)从2.0、1.8、1.7和1.9 ppm降低到1.9、1.6、1.6和1.7 ppm。我们的方法在激振过程中获得了稳定的高光谱强度,显著降低了电极磨损,激振50次后,电极外径变化为电极长度变化的44.4%。结果表明,该方法具有信号稳定性和准确定量分析的能力,突出了RDE-OES在石油分析中的重要性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
6.20
自引率
26.50%
发文量
228
审稿时长
1.7 months
期刊介绍: Innovative research on the fundamental theory and application of spectrometric techniques.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信