Rishi Puri MD, PhD , Julianne Spencer PhD , Didier Tchétché MD , Nicolas M. Van Mieghem MD, PhD , John K. Forrest MD , Michael J. Reardon MD , Jorge Zhingre Sanchez PhD , Andres Caballero PhD , Philipp Blanke MD , Jonathon A. Leipsic MD , Paul Sorajja MD , G. Michael Deeb MD , Shinichi Fukuhara MD , Lindsay M. Lucas MSc , Taofik Oyekunle MSc , Gilbert H.L. Tang MD, MSc, MBA
{"title":"Inflow-to-Outflow Stent Frame Expansion, Ellipticity, and Decoupling in Evolut TAVR: Implications for Mid-term Hemodynamic Performance","authors":"Rishi Puri MD, PhD , Julianne Spencer PhD , Didier Tchétché MD , Nicolas M. Van Mieghem MD, PhD , John K. Forrest MD , Michael J. Reardon MD , Jorge Zhingre Sanchez PhD , Andres Caballero PhD , Philipp Blanke MD , Jonathon A. Leipsic MD , Paul Sorajja MD , G. Michael Deeb MD , Shinichi Fukuhara MD , Lindsay M. Lucas MSc , Taofik Oyekunle MSc , Gilbert H.L. Tang MD, MSc, MBA","doi":"10.1016/j.jscai.2024.102448","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>The native aortic annulus for self-expanding transcatheter aortic valve replacement (TAVR) has variable ellipticity. A noncircular and underexpanded transcatheter aortic valve (TAV) may impact hemodynamic performance. This study aimed to quantify Evolut TAV (Medtronic) frame ellipticity and expansion 30 days post-TAVR and evaluate their impact on 1-year hypoattenuating leaflet thickening and 4-year hemodynamics.</div></div><div><h3>Methods</h3><div>We retrospectively evaluated 184 patients from the Evolut Low Risk substudy with high-quality computed tomography images. Frame ellipticity ratio and percent expansion were quantified at each frame node level 30 days after TAVR. Variables associated with frame deformation, 1-year hypoattenuating leaflet thickening, and 4-year hemodynamics were identified.</div></div><div><h3>Results</h3><div>Mean Evolut frame ellipticity was highest at the inflow (1.18 ± 0.08) and lowest at the functional leaflet region (1.05 ± 0.03) and frame outflow (1.04 ± 0.03). Frame expansion was lowest at the inflow (83.8% ± 4.9%) and highest at the functional leaflet region (97.8% ± 1.7%). TAV frame circularity and expansion significantly increased from the annular level to the leaflet region (<em>P</em> < .001). Mean gradient, effective orifice area, and paravalvular regurgitation at 4 years were not affected by Evolut TAV's relative noncircularity and underexpansion at the frame inflow. Frame underexpansion at the leaflet region, however, was associated with a smaller effective orifice area at 4 years.</div></div><div><h3>Conclusions</h3><div>Evolut frame deformation at the inflow did not affect the circularity and expansion of the stent at the functional leaflet region. Mid-term (4-year) Evolut hemodynamic performance does not appear to be impacted by frame inflow geometry.</div></div>","PeriodicalId":73990,"journal":{"name":"Journal of the Society for Cardiovascular Angiography & Interventions","volume":"4 1","pages":"Article 102448"},"PeriodicalIF":0.0000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of the Society for Cardiovascular Angiography & Interventions","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2772930324021379","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Background
The native aortic annulus for self-expanding transcatheter aortic valve replacement (TAVR) has variable ellipticity. A noncircular and underexpanded transcatheter aortic valve (TAV) may impact hemodynamic performance. This study aimed to quantify Evolut TAV (Medtronic) frame ellipticity and expansion 30 days post-TAVR and evaluate their impact on 1-year hypoattenuating leaflet thickening and 4-year hemodynamics.
Methods
We retrospectively evaluated 184 patients from the Evolut Low Risk substudy with high-quality computed tomography images. Frame ellipticity ratio and percent expansion were quantified at each frame node level 30 days after TAVR. Variables associated with frame deformation, 1-year hypoattenuating leaflet thickening, and 4-year hemodynamics were identified.
Results
Mean Evolut frame ellipticity was highest at the inflow (1.18 ± 0.08) and lowest at the functional leaflet region (1.05 ± 0.03) and frame outflow (1.04 ± 0.03). Frame expansion was lowest at the inflow (83.8% ± 4.9%) and highest at the functional leaflet region (97.8% ± 1.7%). TAV frame circularity and expansion significantly increased from the annular level to the leaflet region (P < .001). Mean gradient, effective orifice area, and paravalvular regurgitation at 4 years were not affected by Evolut TAV's relative noncircularity and underexpansion at the frame inflow. Frame underexpansion at the leaflet region, however, was associated with a smaller effective orifice area at 4 years.
Conclusions
Evolut frame deformation at the inflow did not affect the circularity and expansion of the stent at the functional leaflet region. Mid-term (4-year) Evolut hemodynamic performance does not appear to be impacted by frame inflow geometry.