Ian Pong , Algirdas Baskys , Charlie Sanabria , Michael Naus , Scott Myers , Heng Pan , Jonathan Lee , Li Wang , Jordan Taylor , Marcos Turqueti , Xiaorong Wang
{"title":"Residual resistance ratio measurement system for Nb3Sn wires extracted from Rutherford cables","authors":"Ian Pong , Algirdas Baskys , Charlie Sanabria , Michael Naus , Scott Myers , Heng Pan , Jonathan Lee , Li Wang , Jordan Taylor , Marcos Turqueti , Xiaorong Wang","doi":"10.1016/j.cryogenics.2024.104002","DOIUrl":null,"url":null,"abstract":"<div><div>Residual resistance ratio (RRR) of superconducting strands is an important parameter for magnet electrical stability. RRR serves as a measure of the low-temperature electrical conductivity of the copper within a conductor that has a copper stabilization matrix. For Nb<sub>3</sub>Sn, due to the need of a reaction heat treatment, the technical requirements for high quality measurements of strands extracted from Rutherford cables are particularly demanding. Quality of wire, cabling deformation, heat treatment temperature, heat treatment atmosphere, sample handling, and measurement methods can all affect the RRR. Therefore, as an integral part of the electrical quality control (QC) of Nb<sub>3</sub>Sn Rutherford cables manufactured at the Lawrence Berkeley National Laboratory, it was prudent that we established a RRR measurement system that can isolate the assessment of cable-fabrication-related impacts from sample preparation and measurement factors.</div><div>Here we describe a bespoke cryocooler-based measurement system, capable of measuring RRR of over 80 samples in a single cooldown. The samples are mounted on custom-designed printed circuit boards that accommodate the shape of strands extracted from a Rutherford cable without added deformation, which we will show is critical in ensuring that the measurements accurately represent the RRR values of the conductor within the cable. Using this sample mounting solution, we routinely measure the overall RRR of the strand as well as individual intra-strand sections corresponding to both cable edges and cable broad faces with high reproducibility. Such measurements provide valuable information on the variation of RRR along the length of the strands as well as across strand productions and cable runs over time.</div></div>","PeriodicalId":10812,"journal":{"name":"Cryogenics","volume":"146 ","pages":"Article 104002"},"PeriodicalIF":1.8000,"publicationDate":"2024-12-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cryogenics","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0011227524002224","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"PHYSICS, APPLIED","Score":null,"Total":0}
引用次数: 0
Abstract
Residual resistance ratio (RRR) of superconducting strands is an important parameter for magnet electrical stability. RRR serves as a measure of the low-temperature electrical conductivity of the copper within a conductor that has a copper stabilization matrix. For Nb3Sn, due to the need of a reaction heat treatment, the technical requirements for high quality measurements of strands extracted from Rutherford cables are particularly demanding. Quality of wire, cabling deformation, heat treatment temperature, heat treatment atmosphere, sample handling, and measurement methods can all affect the RRR. Therefore, as an integral part of the electrical quality control (QC) of Nb3Sn Rutherford cables manufactured at the Lawrence Berkeley National Laboratory, it was prudent that we established a RRR measurement system that can isolate the assessment of cable-fabrication-related impacts from sample preparation and measurement factors.
Here we describe a bespoke cryocooler-based measurement system, capable of measuring RRR of over 80 samples in a single cooldown. The samples are mounted on custom-designed printed circuit boards that accommodate the shape of strands extracted from a Rutherford cable without added deformation, which we will show is critical in ensuring that the measurements accurately represent the RRR values of the conductor within the cable. Using this sample mounting solution, we routinely measure the overall RRR of the strand as well as individual intra-strand sections corresponding to both cable edges and cable broad faces with high reproducibility. Such measurements provide valuable information on the variation of RRR along the length of the strands as well as across strand productions and cable runs over time.
期刊介绍:
Cryogenics is the world''s leading journal focusing on all aspects of cryoengineering and cryogenics. Papers published in Cryogenics cover a wide variety of subjects in low temperature engineering and research. Among the areas covered are:
- Applications of superconductivity: magnets, electronics, devices
- Superconductors and their properties
- Properties of materials: metals, alloys, composites, polymers, insulations
- New applications of cryogenic technology to processes, devices, machinery
- Refrigeration and liquefaction technology
- Thermodynamics
- Fluid properties and fluid mechanics
- Heat transfer
- Thermometry and measurement science
- Cryogenics in medicine
- Cryoelectronics