Recurrent disconnection of Gore Iliac Branch Endoprosthesis: Case report and review of contributing factors

Catherine Boudreau, Caroline Caradu, Xavier Bérard, Eric Ducasse
{"title":"Recurrent disconnection of Gore Iliac Branch Endoprosthesis: Case report and review of contributing factors","authors":"Catherine Boudreau,&nbsp;Caroline Caradu,&nbsp;Xavier Bérard,&nbsp;Eric Ducasse","doi":"10.1016/j.avsurg.2025.100363","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><div>Iliac artery aneurysms (IAAs) which may coexist with abdominal aortic aneurysms (AAAs) or appear as isolated lesions, represent around 2 % of intra-abdominal aneurysms. Endovascular repair using iliac branch devices (IBDs) helps preserve internal iliac artery (IIA) perfusion, reducing complications like buttock claudication. However, device displacement and type III endoleaks are significant challenges, especially in tortuous anatomy.</div></div><div><h3>Report</h3><div>We report two cases of type III endoleak due to displacement of Gore Iliac Branch Endoprosthesis (IBE). In the first case, a 60-year-old male with a left IAA experienced immediate postoperative IBE migration and type III endoleak, necessitating urgent reintervention with bridging stents. In the second, a 74-year-old male treated for a right IAA had recurrent IBE displacements over four years, with repeated type III endoleaks, managed by repeated stenting. Severe iliac tortuosity contributed to both cases.</div></div><div><h3>Discussion</h3><div>These cases highlight the complexities of IAA endovascular repair with IBDs, particularly the risk of device migration and endoleaks, with severe iliac tortuosity, device properties (e.g. conformability, low friction coefficient), and response to cardiac pulsatility increasing migration risk. Future studies should address these challenges, focusing on device design improvements and predictive imaging techniques, including artificial intelligence, to enhance long-term outcomes in IAA repair.</div></div>","PeriodicalId":72235,"journal":{"name":"Annals of vascular surgery. Brief reports and innovations","volume":"5 1","pages":"Article 100363"},"PeriodicalIF":0.0000,"publicationDate":"2025-01-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of vascular surgery. Brief reports and innovations","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2772687825000042","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction

Iliac artery aneurysms (IAAs) which may coexist with abdominal aortic aneurysms (AAAs) or appear as isolated lesions, represent around 2 % of intra-abdominal aneurysms. Endovascular repair using iliac branch devices (IBDs) helps preserve internal iliac artery (IIA) perfusion, reducing complications like buttock claudication. However, device displacement and type III endoleaks are significant challenges, especially in tortuous anatomy.

Report

We report two cases of type III endoleak due to displacement of Gore Iliac Branch Endoprosthesis (IBE). In the first case, a 60-year-old male with a left IAA experienced immediate postoperative IBE migration and type III endoleak, necessitating urgent reintervention with bridging stents. In the second, a 74-year-old male treated for a right IAA had recurrent IBE displacements over four years, with repeated type III endoleaks, managed by repeated stenting. Severe iliac tortuosity contributed to both cases.

Discussion

These cases highlight the complexities of IAA endovascular repair with IBDs, particularly the risk of device migration and endoleaks, with severe iliac tortuosity, device properties (e.g. conformability, low friction coefficient), and response to cardiac pulsatility increasing migration risk. Future studies should address these challenges, focusing on device design improvements and predictive imaging techniques, including artificial intelligence, to enhance long-term outcomes in IAA repair.
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
0.20
自引率
0.00%
发文量
0
审稿时长
62 days
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信