Validation of the Pangao PG-800B18 upper-arm blood pressure monitor according to the ISO 81060-2:2018+Amd 1:2020 protocol in the general population.

IF 1.2 4区 医学 Q4 PERIPHERAL VASCULAR DISEASE
Wan Chen, Yuhan Xiang, Rui Cheng, Yang Zhang
{"title":"Validation of the Pangao PG-800B18 upper-arm blood pressure monitor according to the ISO 81060-2:2018+Amd 1:2020 protocol in the general population.","authors":"Wan Chen, Yuhan Xiang, Rui Cheng, Yang Zhang","doi":"10.1097/MBP.0000000000000744","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The study aimed to validate the Pangao PG-800B18 oscillometric upper-arm blood pressure monitor in the general population in accordance with the International Organization for Standardization (ISO) 81060-2:2018 and amendment (Amd) 1:2020 protocol.</p><p><strong>Methods: </strong>Participants were recruited, and the same left-arm sequential method was used for blood pressure measurement in accordance with the protocol. Data validation and analysis were performed according to the protocol, and the Bland-Altman scatter plot was used to show the difference between the test device and reference results.</p><p><strong>Results: </strong>A total of 86 participants were included in the validation analysis. For the validation criterion 1 of the protocol, the mean ± SD of the differences between the test device and reference readings was -1.21 ± 6.71 mmHg and -0.73 ± 5.83 mmHg for systolic (SBP) and diastolic blood pressure (DBP), respectively. Furthermore, the SD of the differences between the test device and reference readings per participant was 5.81 and 5.02 mmHg for SBP and DBP, fulfilling the criterion 2 with SD values of ≤6.95 for SBP and ≤5.89 for DBP.</p><p><strong>Conclusion: </strong>The Pangao PG-800B18 upper-arm blood pressure monitor fulfilled the requirements of the ISO validation standard and can be recommended for clinical use and self-measurement in the general population.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":" ","pages":""},"PeriodicalIF":1.2000,"publicationDate":"2025-01-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000744","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: The study aimed to validate the Pangao PG-800B18 oscillometric upper-arm blood pressure monitor in the general population in accordance with the International Organization for Standardization (ISO) 81060-2:2018 and amendment (Amd) 1:2020 protocol.

Methods: Participants were recruited, and the same left-arm sequential method was used for blood pressure measurement in accordance with the protocol. Data validation and analysis were performed according to the protocol, and the Bland-Altman scatter plot was used to show the difference between the test device and reference results.

Results: A total of 86 participants were included in the validation analysis. For the validation criterion 1 of the protocol, the mean ± SD of the differences between the test device and reference readings was -1.21 ± 6.71 mmHg and -0.73 ± 5.83 mmHg for systolic (SBP) and diastolic blood pressure (DBP), respectively. Furthermore, the SD of the differences between the test device and reference readings per participant was 5.81 and 5.02 mmHg for SBP and DBP, fulfilling the criterion 2 with SD values of ≤6.95 for SBP and ≤5.89 for DBP.

Conclusion: The Pangao PG-800B18 upper-arm blood pressure monitor fulfilled the requirements of the ISO validation standard and can be recommended for clinical use and self-measurement in the general population.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Blood Pressure Monitoring
Blood Pressure Monitoring 医学-外周血管病
CiteScore
2.00
自引率
7.70%
发文量
110
审稿时长
>12 weeks
期刊介绍: Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research. This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine. Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信