Utility of a 13C-Spirulina Stable Isotope Gastric Emptying Breath Test in Diabetes Mellitus.

IF 3.5 3区 医学 Q1 CLINICAL NEUROLOGY
Revati Varma, Catherine E Williams, Ethan S McClain, Kent R Bailey, Tamas Ordog, Adil E Bharucha
{"title":"Utility of a <sup>13</sup>C-Spirulina Stable Isotope Gastric Emptying Breath Test in Diabetes Mellitus.","authors":"Revati Varma, Catherine E Williams, Ethan S McClain, Kent R Bailey, Tamas Ordog, Adil E Bharucha","doi":"10.1111/nmo.15008","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The carbon-13 spirulina gastric emptying breath test (GEBT) is approved to identify delayed, but not accelerated, gastric emptying (GE). We compared the utility of the GEBT to scintigraphy for diagnosing abnormal GE in patients with diabetes mellitus.</p><p><strong>Methods: </strong>Twenty-eight patients with diabetes ate a 230-kcal test meal labeled with technetium 99 m and <sup>13</sup>C-spirulina, after which 10 scintigraphic images and breath samples (baseline, 15, 30, 45, 60, 90, 120, 150, 180, 210, and 240 min) were collected on 2 occasions 1 week apart. We assessed the accuracy of <sup>13</sup>C-spirulina GEBT excretion rate (percent dose multiplied by 1000 [kPCD] min<sup>-1</sup>) values to predict scintigraphic half-life and distinguish between normal, delayed, and accelerated GE and the intraindividual reproducibility of the GEBT.</p><p><strong>Key results: </strong>Scintigraphy revealed normal, delayed, and accelerated GE, respectively, in 17 (30%), 29 (52%), and 10 (18%) test results. GE T½ values measured with scintigraphy and GEBT were highly concordant within individuals; the intraindividual reproducibility was 34% (scintigraphy) and 15% (GEBT). Compared to current criteria, the kPCD150 (150 min) and kPCD180 values provided equally sensitive (90%) and more specific (81% vs. 67%) approach for distinguishing between delayed versus normal/accelerated GE. A new metric (kPCD60-kPCD15 min) was 90% sensitive and 83% specific for distinguishing between accelerated versus normal/delayed GE. These findings were used to create nomograms and an algorithm for interpreting GEBT results.</p><p><strong>Conclusions and inferences: </strong>Among patients with poorly controlled diabetes, the <sup>13</sup>C-spirulina GEBT can accurately and precisely assess GE and effectively distinguish between normal, delayed, and accelerated GE.</p>","PeriodicalId":19123,"journal":{"name":"Neurogastroenterology and Motility","volume":" ","pages":"e15008"},"PeriodicalIF":3.5000,"publicationDate":"2025-01-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Neurogastroenterology and Motility","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/nmo.15008","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The carbon-13 spirulina gastric emptying breath test (GEBT) is approved to identify delayed, but not accelerated, gastric emptying (GE). We compared the utility of the GEBT to scintigraphy for diagnosing abnormal GE in patients with diabetes mellitus.

Methods: Twenty-eight patients with diabetes ate a 230-kcal test meal labeled with technetium 99 m and 13C-spirulina, after which 10 scintigraphic images and breath samples (baseline, 15, 30, 45, 60, 90, 120, 150, 180, 210, and 240 min) were collected on 2 occasions 1 week apart. We assessed the accuracy of 13C-spirulina GEBT excretion rate (percent dose multiplied by 1000 [kPCD] min-1) values to predict scintigraphic half-life and distinguish between normal, delayed, and accelerated GE and the intraindividual reproducibility of the GEBT.

Key results: Scintigraphy revealed normal, delayed, and accelerated GE, respectively, in 17 (30%), 29 (52%), and 10 (18%) test results. GE T½ values measured with scintigraphy and GEBT were highly concordant within individuals; the intraindividual reproducibility was 34% (scintigraphy) and 15% (GEBT). Compared to current criteria, the kPCD150 (150 min) and kPCD180 values provided equally sensitive (90%) and more specific (81% vs. 67%) approach for distinguishing between delayed versus normal/accelerated GE. A new metric (kPCD60-kPCD15 min) was 90% sensitive and 83% specific for distinguishing between accelerated versus normal/delayed GE. These findings were used to create nomograms and an algorithm for interpreting GEBT results.

Conclusions and inferences: Among patients with poorly controlled diabetes, the 13C-spirulina GEBT can accurately and precisely assess GE and effectively distinguish between normal, delayed, and accelerated GE.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Neurogastroenterology and Motility
Neurogastroenterology and Motility 医学-临床神经学
CiteScore
7.80
自引率
8.60%
发文量
178
审稿时长
3-6 weeks
期刊介绍: Neurogastroenterology & Motility (NMO) is the official Journal of the European Society of Neurogastroenterology & Motility (ESNM) and the American Neurogastroenterology and Motility Society (ANMS). It is edited by James Galligan, Albert Bredenoord, and Stephen Vanner. The editorial and peer review process is independent of the societies affiliated to the journal and publisher: Neither the ANMS, the ESNM or the Publisher have editorial decision-making power. Whenever these are relevant to the content being considered or published, the editors, journal management committee and editorial board declare their interests and affiliations.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信