Re-sheathing failure with Navitor during transcatheter aortic valve implantation: a case report.

IF 0.8 Q4 CARDIAC & CARDIOVASCULAR SYSTEMS
European Heart Journal: Case Reports Pub Date : 2025-01-21 eCollection Date: 2025-01-01 DOI:10.1093/ehjcr/ytaf015
Hiroshi Onoda, Hiroshi Ueno, Yuki Hida, Teruhiko Imamura, Koichiro Kinugawa
{"title":"Re-sheathing failure with Navitor during transcatheter aortic valve implantation: a case report.","authors":"Hiroshi Onoda, Hiroshi Ueno, Yuki Hida, Teruhiko Imamura, Koichiro Kinugawa","doi":"10.1093/ehjcr/ytaf015","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Self-expanding valves used in transcatheter aortic valve implantation (TAVI) are designed to allow recapture and repositioning, facilitating optimal placement and mitigating conduction disturbances and paravalvular leakage. Here, we present a rare case in which the Navitor (Abbott Structural Heart, Santa Clara, CA, USA) could not be recaptured.</p><p><strong>Case summary: </strong>An 81-year-old Japanese woman with very severe aortic stenosis and a massively calcified nodule at the non-coronary cusp (NCC) underwent TAVI with a 25 mm Navitor valve. During the initial deployment attempt, non-uniform expansion (NUE) was observed on the NCC side when the valve was 80% deployed. An attempt was made to recapture and reposition the valve, but the delivery system capsule failed to fully re-sheath the prosthesis, leaving approximately one-third of the valve outside the capsule and preventing complete recapture. The Navitor was promptly redeployed while still within the basal ring. Following redeployment, the NUE resolved, and the valve was successfully positioned 3 mm below the basal ring on the NCC side and 4 mm below the left coronary cusp. We hypothesized that interference between the capsule tip and the calcified nodule on the NCC leaflet inhibited the re-sheathing process.</p><p><strong>Conclusion: </strong>This report documents a rare complication involving the failure to recapture the Navitor valve. In cases with large calcified nodules on the leaflet, caution is essential during the re-sheathing process. We strongly recommend re-deploying the prosthesis rather than attempting to remove it from the basal ring to minimize procedural risks and ensure proper valve placement.</p>","PeriodicalId":11910,"journal":{"name":"European Heart Journal: Case Reports","volume":"9 1","pages":"ytaf015"},"PeriodicalIF":0.8000,"publicationDate":"2025-01-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11770386/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Heart Journal: Case Reports","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1093/ehjcr/ytaf015","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Self-expanding valves used in transcatheter aortic valve implantation (TAVI) are designed to allow recapture and repositioning, facilitating optimal placement and mitigating conduction disturbances and paravalvular leakage. Here, we present a rare case in which the Navitor (Abbott Structural Heart, Santa Clara, CA, USA) could not be recaptured.

Case summary: An 81-year-old Japanese woman with very severe aortic stenosis and a massively calcified nodule at the non-coronary cusp (NCC) underwent TAVI with a 25 mm Navitor valve. During the initial deployment attempt, non-uniform expansion (NUE) was observed on the NCC side when the valve was 80% deployed. An attempt was made to recapture and reposition the valve, but the delivery system capsule failed to fully re-sheath the prosthesis, leaving approximately one-third of the valve outside the capsule and preventing complete recapture. The Navitor was promptly redeployed while still within the basal ring. Following redeployment, the NUE resolved, and the valve was successfully positioned 3 mm below the basal ring on the NCC side and 4 mm below the left coronary cusp. We hypothesized that interference between the capsule tip and the calcified nodule on the NCC leaflet inhibited the re-sheathing process.

Conclusion: This report documents a rare complication involving the failure to recapture the Navitor valve. In cases with large calcified nodules on the leaflet, caution is essential during the re-sheathing process. We strongly recommend re-deploying the prosthesis rather than attempting to remove it from the basal ring to minimize procedural risks and ensure proper valve placement.

求助全文
约1分钟内获得全文 求助全文
来源期刊
European Heart Journal: Case Reports
European Heart Journal: Case Reports Medicine-Cardiology and Cardiovascular Medicine
CiteScore
1.30
自引率
10.00%
发文量
451
审稿时长
14 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信