{"title":"Validation of Analytical Performance and Diagnostic Effectiveness of Anti-CCP Antibody Based on the Light-Initiated Chemiluminescence Assay.","authors":"Shi-Zhen Guan, Hong-Jing Zhang, Shan-Shan Cheng, Yi Wang, Hong-Wei Yang, Hui-Qiang Li","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To evaluate the analytical and diagnostic performance of anti-cyclic citrullinated peptide (anti-CCP) antibody measured by the light-initiated chemiluminescence assay (LICA).</p><p><strong>Methods: </strong>A total of 193 patients with rheumatoid arthritis (RA) and 211 controls were included. Serum anti-CCP levels were assessed by LICA and enzyme-linked immunosorbent assay (ELSIA). The analytical performance of LICA anti-CCP assay was evaluated according to the clinical guidelines. The diagnostic performance of LICA anti-CCP assay was evaluated by a series of indicators such as sensitivity, specificity, and area under the receiver operating characteristic curve (AUC) compared with ELISA. The dynamic consistency of the two methods was compared by patient follow-up.</p><p><strong>Results: </strong>The LICA anti-CCP assay showed coefficient of variation of 0.6% - 3.3% and linearity within 10-500 U/mL with a correlation coefficient of 0.991. The LICA and ELISA showed similar sensitivity (88.60% vs 92.94%) and specificity (97.60% vs 94.95%) on determination of anti-CCP. The AUCs of anti-CCP detected by LICA and ELISA were 0.971 and 0.955 (<i>P</i>=0.052), respectively. Follow-up showed that the changes of anti-CCP measured by the two methods were consistent.</p><p><strong>Conclusions: </strong>LICA anti-CCP assay has good analytical and diagnostic performance for RA, which is comparable to ELISA, and has dynamic consistency.</p>","PeriodicalId":8228,"journal":{"name":"Annals of clinical and laboratory science","volume":"54 6","pages":"860-866"},"PeriodicalIF":1.1000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of clinical and laboratory science","FirstCategoryId":"3","ListUrlMain":"","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: To evaluate the analytical and diagnostic performance of anti-cyclic citrullinated peptide (anti-CCP) antibody measured by the light-initiated chemiluminescence assay (LICA).

Methods: A total of 193 patients with rheumatoid arthritis (RA) and 211 controls were included. Serum anti-CCP levels were assessed by LICA and enzyme-linked immunosorbent assay (ELSIA). The analytical performance of LICA anti-CCP assay was evaluated according to the clinical guidelines. The diagnostic performance of LICA anti-CCP assay was evaluated by a series of indicators such as sensitivity, specificity, and area under the receiver operating characteristic curve (AUC) compared with ELISA. The dynamic consistency of the two methods was compared by patient follow-up.

Results: The LICA anti-CCP assay showed coefficient of variation of 0.6% - 3.3% and linearity within 10-500 U/mL with a correlation coefficient of 0.991. The LICA and ELISA showed similar sensitivity (88.60% vs 92.94%) and specificity (97.60% vs 94.95%) on determination of anti-CCP. The AUCs of anti-CCP detected by LICA and ELISA were 0.971 and 0.955 (P=0.052), respectively. Follow-up showed that the changes of anti-CCP measured by the two methods were consistent.

Conclusions: LICA anti-CCP assay has good analytical and diagnostic performance for RA, which is comparable to ELISA, and has dynamic consistency.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Annals of clinical and laboratory science
Annals of clinical and laboratory science 医学-医学实验技术
CiteScore
1.60
自引率
0.00%
发文量
112
审稿时长
6-12 weeks
期刊介绍: The Annals of Clinical & Laboratory Science welcomes manuscripts that report research in clinical science, including pathology, clinical chemistry, biotechnology, molecular biology, cytogenetics, microbiology, immunology, hematology, transfusion medicine, organ and tissue transplantation, therapeutics, toxicology, and clinical informatics.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信