Free-Breathing Respiratory Triggered High-Pitch Lung CT: Insights From Phantom and Patient Scans.

IF 7 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Annette Schwarz, Christian Hofmann, Jannis Dickmann, Arndt Simon, Andreas Maier, Frank K Wacker, Hans-Jürgen Raatschen, Stephan Gleitz, Martina Schmidbauer
{"title":"Free-Breathing Respiratory Triggered High-Pitch Lung CT: Insights From Phantom and Patient Scans.","authors":"Annette Schwarz, Christian Hofmann, Jannis Dickmann, Arndt Simon, Andreas Maier, Frank K Wacker, Hans-Jürgen Raatschen, Stephan Gleitz, Martina Schmidbauer","doi":"10.1097/RLI.0000000000001157","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Respiratory motion can affect image quality and thus affect the diagnostic accuracy of CT images by masking or mimicking relevant lung pathologies. CT examinations are often performed during deep inspiration and breath-hold to achieve optimal image quality. However, this can be challenging for certain patient groups, such as children, the elderly, or sedated patients. The study aimed to validate a dedicated triggering algorithm for initiating respiratory-triggered high-pitch computed tomography (RT-HPCT) scans in end inspiration and end expiration in complex and irregular respiratory patterns using an anthropomorphic dynamic chest phantom. Additionally, a patient study was conducted to compare the image quality and lung expansion between RT-HPCT and standard HPCT.</p><p><strong>Materials and methods: </strong>The study utilized an algorithm that processes the patient's breathing motion in real-time to determine the appropriate time to initiate a scan. This algorithm was tested on a dynamic, tissue-equivalent chest motion phantom to replicate and simulate 3-dimensional target motion using 28 breathing motion patterns taken from patient with irregular breathing. To evaluate the performance on human patients, prospective RT-HPCT was performed in 18 free-breathing patients. As a reference, unenhanced HPCT of the chest was performed in 20 patients without respiratory triggering during free-breathing. The mean CTDI was 1.73 mGy ± 0.1 mGy for HPCT and 1.68 mGy ± 0.1 mGy for RT-HPCT. For phantom tests, the deviation from the target position of the phantom inlay is known. Image quality is approximated by evaluating stationary versus moving acquisitions. For patient scans, respiratory motion artifacts and inspiration depth were analyzed using expert knowledge of lung anatomy and automated lung volume estimation. Statistical analysis was performed to compare image quality and lung volumes between conventional HPCT and RT-HPCT.</p><p><strong>Results: </strong>In phantom scans, the average deviation from the desired excursion phase was 1.6 mm ± 4.7 mm or 15% ± 24% of the phantom movement range. In patients, the overall image quality significantly improved with respiratory triggering compared with conventional HPCT ( P < 0.001). Quantitative average lung volume was 4.0 L ± 1.1 L in the RT group and 3.6 L ± 1.0 L in the control group.</p><p><strong>Conclusions: </strong>This study demonstrated the feasibility of using a patient-adaptive respiratory triggering algorithm for high-pitch lung CT in both phantom and patients. Respiratory-triggered high-pitch CT scanning significantly reduces breathing artifacts compared with conventional nontriggered free-breathing scans.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":""},"PeriodicalIF":7.0000,"publicationDate":"2025-02-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Investigative Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/RLI.0000000000001157","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: Respiratory motion can affect image quality and thus affect the diagnostic accuracy of CT images by masking or mimicking relevant lung pathologies. CT examinations are often performed during deep inspiration and breath-hold to achieve optimal image quality. However, this can be challenging for certain patient groups, such as children, the elderly, or sedated patients. The study aimed to validate a dedicated triggering algorithm for initiating respiratory-triggered high-pitch computed tomography (RT-HPCT) scans in end inspiration and end expiration in complex and irregular respiratory patterns using an anthropomorphic dynamic chest phantom. Additionally, a patient study was conducted to compare the image quality and lung expansion between RT-HPCT and standard HPCT.

Materials and methods: The study utilized an algorithm that processes the patient's breathing motion in real-time to determine the appropriate time to initiate a scan. This algorithm was tested on a dynamic, tissue-equivalent chest motion phantom to replicate and simulate 3-dimensional target motion using 28 breathing motion patterns taken from patient with irregular breathing. To evaluate the performance on human patients, prospective RT-HPCT was performed in 18 free-breathing patients. As a reference, unenhanced HPCT of the chest was performed in 20 patients without respiratory triggering during free-breathing. The mean CTDI was 1.73 mGy ± 0.1 mGy for HPCT and 1.68 mGy ± 0.1 mGy for RT-HPCT. For phantom tests, the deviation from the target position of the phantom inlay is known. Image quality is approximated by evaluating stationary versus moving acquisitions. For patient scans, respiratory motion artifacts and inspiration depth were analyzed using expert knowledge of lung anatomy and automated lung volume estimation. Statistical analysis was performed to compare image quality and lung volumes between conventional HPCT and RT-HPCT.

Results: In phantom scans, the average deviation from the desired excursion phase was 1.6 mm ± 4.7 mm or 15% ± 24% of the phantom movement range. In patients, the overall image quality significantly improved with respiratory triggering compared with conventional HPCT ( P < 0.001). Quantitative average lung volume was 4.0 L ± 1.1 L in the RT group and 3.6 L ± 1.0 L in the control group.

Conclusions: This study demonstrated the feasibility of using a patient-adaptive respiratory triggering algorithm for high-pitch lung CT in both phantom and patients. Respiratory-triggered high-pitch CT scanning significantly reduces breathing artifacts compared with conventional nontriggered free-breathing scans.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Investigative Radiology
Investigative Radiology 医学-核医学
CiteScore
15.10
自引率
16.40%
发文量
188
审稿时长
4-8 weeks
期刊介绍: Investigative Radiology publishes original, peer-reviewed reports on clinical and laboratory investigations in diagnostic imaging, the diagnostic use of radioactive isotopes, computed tomography, positron emission tomography, magnetic resonance imaging, ultrasound, digital subtraction angiography, and related modalities. Emphasis is on early and timely publication. Primarily research-oriented, the journal also includes a wide variety of features of interest to clinical radiologists.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信