Novel Real-Time Fluoroscopic Assessment Method of THV Expansion Following Balloon-Expandable TAVR.

IF 6.1 1区 医学 Q1 CARDIAC & CARDIOVASCULAR SYSTEMS
Ali Husain, Georgios Tzimas, Mariama Akodad, Julius Jelisejevas, Sophie Offen, Kevin Millar, Jonathon A Leipsic, Philipp Blanke, David A Wood, Stephanie L Sellers, John G Webb, David Meier, Janarthanan Sathananthan
{"title":"Novel Real-Time Fluoroscopic Assessment Method of THV Expansion Following Balloon-Expandable TAVR.","authors":"Ali Husain, Georgios Tzimas, Mariama Akodad, Julius Jelisejevas, Sophie Offen, Kevin Millar, Jonathon A Leipsic, Philipp Blanke, David A Wood, Stephanie L Sellers, John G Webb, David Meier, Janarthanan Sathananthan","doi":"10.1161/CIRCINTERVENTIONS.124.014617","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Transcatheter heart valve (THV) underexpansion after transcatheter aortic valve replacement may be associated with worse outcomes. THV expansion can be assessed fluoroscopically using a pigtail for calibration; however, the accuracy of this technique specific to transcatheter aortic valve replacement is unknown. We assessed the accuracy and reproducibility of a novel fluoroscopic method to assess THV expansion using the THV commissural post for calibration.</p><p><strong>Methods: </strong>Patients who underwent transcatheter aortic valve replacement with a SAPIEN 3 (S3) THV had documented 3 cusp and cusp overlap views, and post-implant computed tomography was identified. THV expansion was fluoroscopically assessed in the 3 cusp and cusp overlap views using the S3 commissural post height and the pigtail as a reference for calibration. The correlation between the 2 methods and computed tomography was evaluated.</p><p><strong>Results: </strong>Forty patients were included. On the bench, the commissural post height measured 3.3, 3.5, 4.0, and 4.5 mm for the 20 mm S3, 23 mm S3, 26 mm S3, and 29 mm S3, respectively. The Pearson correlation coefficient (r) with computed tomography for measuring the inflow, mid-portion, and outflow THV diameter was 0.98, 0.97, and 0.98 for the commissural post height method and 0.82, 0.81, and 0.78 for the pigtail method, respectively. Unlike the pigtail method, the correlation between the commissural post height method and computed tomography remained strong across all THV sizes and in both the 3 cusp and cusp overlap views.</p><p><strong>Conclusions: </strong>The commissural post height method is a novel real-time fluoroscopic tool that has the potential to assess THV expansion and guide further optimization after balloon-expandable transcatheter aortic valve replacement.</p>","PeriodicalId":10330,"journal":{"name":"Circulation: Cardiovascular Interventions","volume":" ","pages":"e014617"},"PeriodicalIF":6.1000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Circulation: Cardiovascular Interventions","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1161/CIRCINTERVENTIONS.124.014617","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/24 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Transcatheter heart valve (THV) underexpansion after transcatheter aortic valve replacement may be associated with worse outcomes. THV expansion can be assessed fluoroscopically using a pigtail for calibration; however, the accuracy of this technique specific to transcatheter aortic valve replacement is unknown. We assessed the accuracy and reproducibility of a novel fluoroscopic method to assess THV expansion using the THV commissural post for calibration.

Methods: Patients who underwent transcatheter aortic valve replacement with a SAPIEN 3 (S3) THV had documented 3 cusp and cusp overlap views, and post-implant computed tomography was identified. THV expansion was fluoroscopically assessed in the 3 cusp and cusp overlap views using the S3 commissural post height and the pigtail as a reference for calibration. The correlation between the 2 methods and computed tomography was evaluated.

Results: Forty patients were included. On the bench, the commissural post height measured 3.3, 3.5, 4.0, and 4.5 mm for the 20 mm S3, 23 mm S3, 26 mm S3, and 29 mm S3, respectively. The Pearson correlation coefficient (r) with computed tomography for measuring the inflow, mid-portion, and outflow THV diameter was 0.98, 0.97, and 0.98 for the commissural post height method and 0.82, 0.81, and 0.78 for the pigtail method, respectively. Unlike the pigtail method, the correlation between the commissural post height method and computed tomography remained strong across all THV sizes and in both the 3 cusp and cusp overlap views.

Conclusions: The commissural post height method is a novel real-time fluoroscopic tool that has the potential to assess THV expansion and guide further optimization after balloon-expandable transcatheter aortic valve replacement.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Circulation: Cardiovascular Interventions
Circulation: Cardiovascular Interventions CARDIAC & CARDIOVASCULAR SYSTEMS-
CiteScore
10.30
自引率
1.80%
发文量
221
审稿时长
6-12 weeks
期刊介绍: Circulation: Cardiovascular Interventions, an American Heart Association journal, focuses on interventional techniques pertaining to coronary artery disease, structural heart disease, and vascular disease, with priority placed on original research and on randomized trials and large registry studies. In addition, pharmacological, diagnostic, and pathophysiological aspects of interventional cardiology are given special attention in this online-only journal.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信