Rapid prospective motion correction using free induction decay and stationary field probe navigators at 7T.

IF 3 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Matthias Serger, Rüdiger Stirnberg, Philipp Ehses, Tony Stöcker
{"title":"Rapid prospective motion correction using free induction decay and stationary field probe navigators at 7T.","authors":"Matthias Serger, Rüdiger Stirnberg, Philipp Ehses, Tony Stöcker","doi":"10.1002/mrm.30441","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>MR-based FID navigators (FIDnavs) do not require gradient pulses and are attractive for prospective motion correction (PMC) due to short acquisition times and high sampling rates. However, accuracy and precision are limited and depend on a separate calibration measurement. Besides FIDnavs, stationary NMR field probes are also capable of measuring local, motion-induced field changes. In this work, a linear model is calibrated between field probe data and motion parameters analog to FIDnav calibration and both tracking methods are compared and combined for PMC.</p><p><strong>Methods: </strong>FIDnavs and field probe navigators were implemented in a fast 3D-EPI sequence and calibrated by a linear model to realignment motion parameters of the 3D-EPI time series. A workflow was established to correct head motion prospectively by FIDnavs, field probe navigators or a combination of both. Large motions were instructed to test the accuracy and the impact on image quality in <math> <semantics><mrow><mn>1</mn> <mspace></mspace> <msup><mtext>mm</mtext> <mn>3</mn></msup> </mrow> <annotation>$$ 1\\kern0.1667em {\\mathrm{mm}}^3 $$</annotation></semantics> </math> EPI data.</p><p><strong>Results: </strong>In a group of five subjects, FIDnavs demonstrated approximately doubled accuracy and precision in comparison with field probe navigators for large motions, especially nodding motions were tracked less accurately by field probes. A combination of both methods could not improve the accuracy consistently. Motion artifacts in high-resolution data were reduced similarly by both PMC methods, although artifacts remained due to susceptibility-induced B0 changes.</p><p><strong>Conclusion: </strong>Stationary field probe navigators can be calibrated equivalently as FIDnavs and enable rapid PMC of large and fast motions. Although they reveal decreased accuracy, their contrast-independence facilitates the potential insertion into many sequences.</p>","PeriodicalId":18065,"journal":{"name":"Magnetic Resonance in Medicine","volume":" ","pages":""},"PeriodicalIF":3.0000,"publicationDate":"2025-01-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic Resonance in Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/mrm.30441","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: MR-based FID navigators (FIDnavs) do not require gradient pulses and are attractive for prospective motion correction (PMC) due to short acquisition times and high sampling rates. However, accuracy and precision are limited and depend on a separate calibration measurement. Besides FIDnavs, stationary NMR field probes are also capable of measuring local, motion-induced field changes. In this work, a linear model is calibrated between field probe data and motion parameters analog to FIDnav calibration and both tracking methods are compared and combined for PMC.

Methods: FIDnavs and field probe navigators were implemented in a fast 3D-EPI sequence and calibrated by a linear model to realignment motion parameters of the 3D-EPI time series. A workflow was established to correct head motion prospectively by FIDnavs, field probe navigators or a combination of both. Large motions were instructed to test the accuracy and the impact on image quality in 1 mm 3 $$ 1\kern0.1667em {\mathrm{mm}}^3 $$ EPI data.

Results: In a group of five subjects, FIDnavs demonstrated approximately doubled accuracy and precision in comparison with field probe navigators for large motions, especially nodding motions were tracked less accurately by field probes. A combination of both methods could not improve the accuracy consistently. Motion artifacts in high-resolution data were reduced similarly by both PMC methods, although artifacts remained due to susceptibility-induced B0 changes.

Conclusion: Stationary field probe navigators can be calibrated equivalently as FIDnavs and enable rapid PMC of large and fast motions. Although they reveal decreased accuracy, their contrast-independence facilitates the potential insertion into many sequences.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
6.70
自引率
24.20%
发文量
376
审稿时长
2-4 weeks
期刊介绍: Magnetic Resonance in Medicine (Magn Reson Med) is an international journal devoted to the publication of original investigations concerned with all aspects of the development and use of nuclear magnetic resonance and electron paramagnetic resonance techniques for medical applications. Reports of original investigations in the areas of mathematics, computing, engineering, physics, biophysics, chemistry, biochemistry, and physiology directly relevant to magnetic resonance will be accepted, as well as methodology-oriented clinical studies.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信