Evaluation of targeting accuracy of cone beam CT guided histotripsy in an in vivo porcine model.

IF 3 3区 医学 Q2 ONCOLOGY
International Journal of Hyperthermia Pub Date : 2025-12-01 Epub Date: 2025-01-22 DOI:10.1080/02656736.2025.2455138
Martin G Wagner, Grace M Minesinger, Katrina L Falk, Ayca Z Kutlu, Meridith A Kisting, Michael A Speidel, Timothy J Ziemlewicz, J Louis Hinshaw, John F Swietlik, Fred T Lee, Paul F Laeseke
{"title":"Evaluation of targeting accuracy of cone beam CT guided histotripsy in an <i>in vivo</i> porcine model.","authors":"Martin G Wagner, Grace M Minesinger, Katrina L Falk, Ayca Z Kutlu, Meridith A Kisting, Michael A Speidel, Timothy J Ziemlewicz, J Louis Hinshaw, John F Swietlik, Fred T Lee, Paul F Laeseke","doi":"10.1080/02656736.2025.2455138","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The application of histotripsy, an emerging noninvasive, non-ionizing, and non-thermal tumor treatment, is currently limited by the inherent limitations of diagnostic ultrasound as the sole targeting modality. This study evaluates the feasibility and accuracy of cone beam computed tomography (CBCT) guidance for histotripsy treatments in an <i>in vivo</i> porcine model.</p><p><strong>Materials and methods: </strong>Histotripsy treatments were performed in the liver of seven healthy swine under the guidance of a C-arm CBCT system that was calibrated to the robotic arm of the histotripsy system. For each treatment, pseudotumors (small histotripsy treatments of 15 mm) were created using conventional US guidance to serve as targets for subsequent CBCT guided treatments. A pretreatment CBCT with intravenous contrast was acquired for each swine and the center of the pseudotumor was selected as the target. The robotic arm automatically aligned the transducer to the selected target location. Ultrasound based aberration offset correction was performed when possible, and a 25 mm diameter treatment was performed. A post-treatment CBCT with intravenous contrast was then acquired to evaluate coverage, treatment size, and distance between the pseudotumor target and actual treatment zone center.</p><p><strong>Results: </strong>Treatments were technically successful and pseudotumors were completely covered in all seven treatments (7/7). The average treatment diameter was 39.3 ± 4.2 mm. The center-to-center distance between pseudotumor and actual treatments was 3.8 ± 1.3 mm.</p><p><strong>Conclusion: </strong>CBCT provides accurate targeting for histotripsy treatment <i>in vivo</i>. While future work is required to assess safety and efficacy in the presence of obstructions, the proposed approach could supplement ultrasound imaging for targeting.</p>","PeriodicalId":14137,"journal":{"name":"International Journal of Hyperthermia","volume":"42 1","pages":"2455138"},"PeriodicalIF":3.0000,"publicationDate":"2025-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11784921/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Hyperthermia","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/02656736.2025.2455138","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/22 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: The application of histotripsy, an emerging noninvasive, non-ionizing, and non-thermal tumor treatment, is currently limited by the inherent limitations of diagnostic ultrasound as the sole targeting modality. This study evaluates the feasibility and accuracy of cone beam computed tomography (CBCT) guidance for histotripsy treatments in an in vivo porcine model.

Materials and methods: Histotripsy treatments were performed in the liver of seven healthy swine under the guidance of a C-arm CBCT system that was calibrated to the robotic arm of the histotripsy system. For each treatment, pseudotumors (small histotripsy treatments of 15 mm) were created using conventional US guidance to serve as targets for subsequent CBCT guided treatments. A pretreatment CBCT with intravenous contrast was acquired for each swine and the center of the pseudotumor was selected as the target. The robotic arm automatically aligned the transducer to the selected target location. Ultrasound based aberration offset correction was performed when possible, and a 25 mm diameter treatment was performed. A post-treatment CBCT with intravenous contrast was then acquired to evaluate coverage, treatment size, and distance between the pseudotumor target and actual treatment zone center.

Results: Treatments were technically successful and pseudotumors were completely covered in all seven treatments (7/7). The average treatment diameter was 39.3 ± 4.2 mm. The center-to-center distance between pseudotumor and actual treatments was 3.8 ± 1.3 mm.

Conclusion: CBCT provides accurate targeting for histotripsy treatment in vivo. While future work is required to assess safety and efficacy in the presence of obstructions, the proposed approach could supplement ultrasound imaging for targeting.

猪体内模型锥束CT引导下组织切片瞄准精度评价。
目的:组织切片术是一种新兴的非侵入性、非电离性和非热性肿瘤治疗方法,目前由于超声诊断作为唯一靶向方式的固有局限性,其应用受到限制。本研究评估了锥束计算机断层扫描(CBCT)指导活体猪组织切片治疗的可行性和准确性。材料和方法:在c臂CBCT系统的指导下,对7头健康猪的肝脏进行了组织切片治疗,该系统被校准到组织切片系统的机械臂。对于每种治疗,使用传统的US引导创建假肿瘤(15 mm的小组织切片治疗),作为后续CBCT引导治疗的靶标。对每头猪进行预处理CBCT加静脉造影剂,选择假肿瘤中心为靶点。机械臂自动将换能器对准选定的目标位置。在可能的情况下进行基于超声的像差偏移校正,并进行直径为25 mm的处理。然后获得治疗后CBCT静脉造影剂,以评估覆盖范围、治疗大小以及假肿瘤靶点与实际治疗区中心之间的距离。结果:7次治疗均取得了技术上的成功,假瘤完全被覆盖(7/7)。平均处理直径为39.3±4.2 mm。假瘤与实际治疗的中心距离为3.8±1.3 mm。结论:CBCT为活体组织切片治疗提供了准确的靶向性。虽然未来的工作需要评估存在障碍物的安全性和有效性,但所提出的方法可以补充超声成像的靶向性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.90
自引率
12.90%
发文量
153
审稿时长
6-12 weeks
期刊介绍: The International Journal of Hyperthermia
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信