{"title":"Establishing quality assurance for COVID-19 antigen tests in the Indo Pacific Region: A multi-site implementation study","authors":"Jacqueline Prestedge , Marilyn Ninan , Chan Leakhena Phoeung , Kennarey Seang , Garrett Young , Bouaphanh Khamphaphongphane , Boualay Norchaleun , Pamela J Toliman , Liza Cabuang , Vonthanak Saphonn , Jessica Markby , Kirsten Ridley , Andrew Vallely , Philip Cunningham , Deborah A Williamson","doi":"10.1016/j.diagmicrobio.2024.116677","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Quality assurance programs (QAPs) are used to evaluate the analytical quality of a diagnostic test and provide feedback to improve quality processes in testing. Rapid diagnostic tests were used in both laboratory and non-laboratory settings to diagnose COVID-19, although varied in reported performance. We aimed to design and implement a QAP for antigen rapid diagnostic tests (Ag-RDTs) for COVID-19 in Cambodia, Lao PDR, and Papua New Guinea.</div></div><div><h3>Methods</h3><div>Inactivated SARS-CoV-2 material derived from cell culture of Wildtype, Delta, and Omicron isolates were used to manufacture 435 quality control (QC) panels (consisting of a single positive and negative sample) and 36 external quality assessment (EQA) panels for use across 15 sites in accordance with provided training and protocols.</div></div><div><h3>Results</h3><div>In total, 369 QC results and 112 EQA results were submitted by electronic reporting or paper forms by two countries. 19·3 % (34/176) positive and 99·5 % (192/193) negative samples were correctly reported in the QC, and 37·5 % (30/80) positive and 97·5 % (39/40) negative samples were correctly reported in the EQA.</div></div><div><h3>Conclusion</h3><div>This QAP demonstrates the importance of operator training and the design of QAP materials, which may have contributed to the high rate of false-negative interpretations in known COVID-19 positive samples. The implementation of this project, to our knowledge the first of its kind in the region and one of only a few globally, emphasised the importance of quality assurance principles, including non-laboratory community testing sites, to promote diagnostic quality for COVID-19 testing.</div></div>","PeriodicalId":11329,"journal":{"name":"Diagnostic microbiology and infectious disease","volume":"111 3","pages":"Article 116677"},"PeriodicalIF":2.1000,"publicationDate":"2024-12-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Diagnostic microbiology and infectious disease","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0732889324005017","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"INFECTIOUS DISEASES","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Quality assurance programs (QAPs) are used to evaluate the analytical quality of a diagnostic test and provide feedback to improve quality processes in testing. Rapid diagnostic tests were used in both laboratory and non-laboratory settings to diagnose COVID-19, although varied in reported performance. We aimed to design and implement a QAP for antigen rapid diagnostic tests (Ag-RDTs) for COVID-19 in Cambodia, Lao PDR, and Papua New Guinea.
Methods
Inactivated SARS-CoV-2 material derived from cell culture of Wildtype, Delta, and Omicron isolates were used to manufacture 435 quality control (QC) panels (consisting of a single positive and negative sample) and 36 external quality assessment (EQA) panels for use across 15 sites in accordance with provided training and protocols.
Results
In total, 369 QC results and 112 EQA results were submitted by electronic reporting or paper forms by two countries. 19·3 % (34/176) positive and 99·5 % (192/193) negative samples were correctly reported in the QC, and 37·5 % (30/80) positive and 97·5 % (39/40) negative samples were correctly reported in the EQA.
Conclusion
This QAP demonstrates the importance of operator training and the design of QAP materials, which may have contributed to the high rate of false-negative interpretations in known COVID-19 positive samples. The implementation of this project, to our knowledge the first of its kind in the region and one of only a few globally, emphasised the importance of quality assurance principles, including non-laboratory community testing sites, to promote diagnostic quality for COVID-19 testing.
期刊介绍:
Diagnostic Microbiology and Infectious Disease keeps you informed of the latest developments in clinical microbiology and the diagnosis and treatment of infectious diseases. Packed with rigorously peer-reviewed articles and studies in bacteriology, immunology, immunoserology, infectious diseases, mycology, parasitology, and virology, the journal examines new procedures, unusual cases, controversial issues, and important new literature. Diagnostic Microbiology and Infectious Disease distinguished independent editorial board, consisting of experts from many medical specialties, ensures you extensive and authoritative coverage.