{"title":"Validation of the Hanvon FY730 upper-arm blood pressure monitor according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018).","authors":"Hao Chen","doi":"10.1097/MBP.0000000000000742","DOIUrl":null,"url":null,"abstract":"<p><p>This study evaluates the measurement accuracy of the Hanvon FY730 upper-arm blood pressure (BP) monitor, which uses the Korotkoff sound method, in compliance with the AAMI/ESH/ISO universal standard (ISO 81060-2:2018). A clinical trial involving 86 eligible adults was conducted, and data were analyzed based on standardized protocols. The mean differences between the device and reference measurements for systolic blood pressure (SBP) and diastolic blood pressure (DBP) were -0.37 mmHg (SD: 2.25 mmHg) and -0.17 mmHg (SD: 2.02 mmHg), respectively. Standard deviations for SBP (≤6.93 mmHg) and DBP (≤6.95 mmHg) met the required thresholds. A total of 258 effective paired measurements demonstrated that the Hanvon FY730 provides highly consistent results compared with reference methods, with all deviations falling within acceptable limits. The study also confirmed the device's reliability across a wide range of arm circumferences (22-36 cm) and BP levels (SBP: 90-188 mmHg, DBP: 53-107 mmHg). These findings confirm that the Hanvon FY730 complies with ISO 81060-2:2018 standards, demonstrating its suitability for accurate BP monitoring in clinical, personal, and home settings.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":" ","pages":""},"PeriodicalIF":1.2000,"publicationDate":"2025-01-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000742","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0
Abstract
This study evaluates the measurement accuracy of the Hanvon FY730 upper-arm blood pressure (BP) monitor, which uses the Korotkoff sound method, in compliance with the AAMI/ESH/ISO universal standard (ISO 81060-2:2018). A clinical trial involving 86 eligible adults was conducted, and data were analyzed based on standardized protocols. The mean differences between the device and reference measurements for systolic blood pressure (SBP) and diastolic blood pressure (DBP) were -0.37 mmHg (SD: 2.25 mmHg) and -0.17 mmHg (SD: 2.02 mmHg), respectively. Standard deviations for SBP (≤6.93 mmHg) and DBP (≤6.95 mmHg) met the required thresholds. A total of 258 effective paired measurements demonstrated that the Hanvon FY730 provides highly consistent results compared with reference methods, with all deviations falling within acceptable limits. The study also confirmed the device's reliability across a wide range of arm circumferences (22-36 cm) and BP levels (SBP: 90-188 mmHg, DBP: 53-107 mmHg). These findings confirm that the Hanvon FY730 complies with ISO 81060-2:2018 standards, demonstrating its suitability for accurate BP monitoring in clinical, personal, and home settings.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.
This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine.
Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.