{"title":"Validation of the ANDON KD-7920 wrist blood pressure monitor in general population according to the ISO 81060-2:2018+AMD1:2020 protocol.","authors":"Duan Li, Yiming Li, Jiaheng Li, Yuan Gu, Fujun Shang","doi":"10.1097/MBP.0000000000000740","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>This study aimed to evaluate the accuracy of the ANDON KD-7920 oscillometric wrist blood pressure monitor in general population according to the ISO 81060-2:2018+AMD1:2020.</p><p><strong>Methods: </strong>The same arm sequential method was used for blood pressure measurement among qualified participants according to the ISO 81060-2:2018+AMD1:2020. The validation results were assessed following the criteria of the protocol, and the Bland-Altman scatter plot was used to present the difference between the test device and reference results.</p><p><strong>Results: </strong>A total of 90 participants were included in the study. According to criterion 1 of the validation protocol, the mean ± SD of the differences between the test device and reference readings was 0.76 ± 6.23 mmHg and -0.13 ± 5.57 mmHg for systolic and diastolic blood pressure, respectively, which met the requirements of less than 5 ± 8 mmHg. According to the criterion 2, the average differences between the test device and reference readings per participant were 0.76 ± 5.56 mmHg and -0.13 ± 5.27 mmHg for systolic and diastolic blood pressure, respectively, which also met the requirements.</p><p><strong>Conclusion: </strong>The ANDON KD-7920 wrist blood pressure monitor passed all the requirements of the ISO 81060-2:2018+AMD1:2020 and can be recommended for clinical use and self-measurement in general population.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":" ","pages":"93-96"},"PeriodicalIF":1.2000,"publicationDate":"2025-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000740","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/12/24 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: This study aimed to evaluate the accuracy of the ANDON KD-7920 oscillometric wrist blood pressure monitor in general population according to the ISO 81060-2:2018+AMD1:2020.
Methods: The same arm sequential method was used for blood pressure measurement among qualified participants according to the ISO 81060-2:2018+AMD1:2020. The validation results were assessed following the criteria of the protocol, and the Bland-Altman scatter plot was used to present the difference between the test device and reference results.
Results: A total of 90 participants were included in the study. According to criterion 1 of the validation protocol, the mean ± SD of the differences between the test device and reference readings was 0.76 ± 6.23 mmHg and -0.13 ± 5.57 mmHg for systolic and diastolic blood pressure, respectively, which met the requirements of less than 5 ± 8 mmHg. According to the criterion 2, the average differences between the test device and reference readings per participant were 0.76 ± 5.56 mmHg and -0.13 ± 5.27 mmHg for systolic and diastolic blood pressure, respectively, which also met the requirements.
Conclusion: The ANDON KD-7920 wrist blood pressure monitor passed all the requirements of the ISO 81060-2:2018+AMD1:2020 and can be recommended for clinical use and self-measurement in general population.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.