Effect of Cyclic Loading on the Fixture-Abutment Microgap in Short Implants Versus Standard Implants: An In Vitro Study.

IF 1.9 Q2 DENTISTRY, ORAL SURGERY & MEDICINE
International Journal of Dentistry Pub Date : 2024-12-27 eCollection Date: 2024-01-01 DOI:10.1155/ijod/4723112
Gholamreza Esfahanizadeh, Ezatollah Jalalian, Seyyede Niloufar Salehi, Mahsa Ghasemi, Shaghayegh Golalipour
{"title":"Effect of Cyclic Loading on the Fixture-Abutment Microgap in Short Implants Versus Standard Implants: An In Vitro Study.","authors":"Gholamreza Esfahanizadeh, Ezatollah Jalalian, Seyyede Niloufar Salehi, Mahsa Ghasemi, Shaghayegh Golalipour","doi":"10.1155/ijod/4723112","DOIUrl":null,"url":null,"abstract":"<p><p><b>Objectives:</b> This study aimed to assess the effect of cyclic loading on the amount of fixture-abutment microgap in short implants compared to standard implants. <b>Materials and Methods:</b> This in vitro experimental study was conducted on two groups of short and standard implants (<i>n</i> = 10). The microgap at the fixture-abutment interface was measured under a light microscope at ×75 magnification. The implants were mounted in an acrylic resin to simulate the jawbone. They were then subjected to cyclic loading by applying 75 N load with 1 Hz frequency along the longitudinal axis of each implant (perpendicular to the abutment surface). After 500,000 cycles, corresponding to 20 months of mastication in the oral environment, the implants were removed from the acrylic resin, and the microgap at the fixture-abutment interface was measured again under a stereomicroscope by a blinded examiner. Data were then analyzed by <i>t</i>-test using SPSS version 22 (<i>α</i> = 0.05). <b>Results:</b> The mean microgap was 13.59 ± 3.80 µm in the standard and 20.41 ± 11.30 µm in the short implants before cyclic loading (<i>p</i>=0.087). These values changed to 15.22 ± 5.44 and 24.53 ± 21.85 µm, respectively, after cyclic loading. No significant difference was noted in the amount of microgap between the standard and short implants after cyclic loading (<i>p</i>=0.222). <b>Conclusion:</b> Cyclic loading increased the amount of fixture-abutment microgap in both the standard and short implants. However, the difference in this respect was not significant between the two implant lengths. Thus, short implants could be reliably used in patients with limitations for surgery to restore function and esthetics.</p>","PeriodicalId":13947,"journal":{"name":"International Journal of Dentistry","volume":"2024 ","pages":"4723112"},"PeriodicalIF":1.9000,"publicationDate":"2024-12-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11698602/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Dentistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1155/ijod/4723112","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: This study aimed to assess the effect of cyclic loading on the amount of fixture-abutment microgap in short implants compared to standard implants. Materials and Methods: This in vitro experimental study was conducted on two groups of short and standard implants (n = 10). The microgap at the fixture-abutment interface was measured under a light microscope at ×75 magnification. The implants were mounted in an acrylic resin to simulate the jawbone. They were then subjected to cyclic loading by applying 75 N load with 1 Hz frequency along the longitudinal axis of each implant (perpendicular to the abutment surface). After 500,000 cycles, corresponding to 20 months of mastication in the oral environment, the implants were removed from the acrylic resin, and the microgap at the fixture-abutment interface was measured again under a stereomicroscope by a blinded examiner. Data were then analyzed by t-test using SPSS version 22 (α = 0.05). Results: The mean microgap was 13.59 ± 3.80 µm in the standard and 20.41 ± 11.30 µm in the short implants before cyclic loading (p=0.087). These values changed to 15.22 ± 5.44 and 24.53 ± 21.85 µm, respectively, after cyclic loading. No significant difference was noted in the amount of microgap between the standard and short implants after cyclic loading (p=0.222). Conclusion: Cyclic loading increased the amount of fixture-abutment microgap in both the standard and short implants. However, the difference in this respect was not significant between the two implant lengths. Thus, short implants could be reliably used in patients with limitations for surgery to restore function and esthetics.

循环加载对短种植体与标准种植体夹具-基台微间隙的影响:体外研究。
目的:本研究旨在评估与标准种植体相比,循环载荷对短种植体中基牙微间隙量的影响。材料与方法:采用短种植体和标准种植体两组(n = 10)进行体外实验研究。在光学显微镜下以×75放大率测量支架-基台界面的微间隙。植入物安装在丙烯酸树脂中以模拟颌骨。然后沿着每个种植体的纵轴(垂直于基牙表面)施加75 N频率为1 Hz的循环载荷。50万次循环后,即在口腔环境中咀嚼20个月后,将种植体从丙烯酸树脂中取出,盲法检查者在体视显微镜下再次测量固定-基牙界面的微间隙。数据采用SPSS 22版t检验(α = 0.05)。结果:标准种植体的平均微间隙为13.59±3.80µm,短种植体的平均微间隙为20.41±11.30µm (p=0.087)。循环加载后,这两个数值分别为15.22±5.44和24.53±21.85µm。标准种植体和短种植体在循环加载后的微间隙量无显著差异(p=0.222)。结论:循环负荷增加了标准种植体和短种植体的基牙微间隙。然而,两种种植体长度在这方面的差异并不显著。因此,短植入物可以可靠地用于手术受限的患者,以恢复功能和美观。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
International Journal of Dentistry
International Journal of Dentistry DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
3.30
自引率
4.80%
发文量
219
审稿时长
20 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信