Evaluation of marginal accuracy of polyetheretherketone and zirconia copings fabricated using computer-aided design/computer-aided manufacturing technique: An in vitro study.
{"title":"Evaluation of marginal accuracy of polyetheretherketone and zirconia copings fabricated using computer-aided design/computer-aided manufacturing technique: An in vitro study.","authors":"Manish Jangade, Ranu Sharma, Nishanth A Sudharson, Raj Jain, Shailendra Kumar Sahu, Anurag Dani","doi":"10.4103/jips.jips_8_24","DOIUrl":null,"url":null,"abstract":"<p><strong>Aim: </strong>The aim of this study was to compare the marginal accuracy of polyetheretherketone (PEEK) and zirconia copings fabricated using computer-aided design/computer-aided manufacturing (CAD/CAM) technology, and to assess the impact of their material properties on accuracy when produced with a 4-axis milling system under controlled conditions.</p><p><strong>Settings and design: </strong>The study employed an in vitro design with a stainless steel die model featuring a 6 mm axial wall height, a 6-degree total occlusal convergence, and a radial shoulder finish line.</p><p><strong>Materials and methods: </strong>Thirty stone dies were created from silicone impressions of the metal die and poured using type-IV dental stone. The dies were divided into two groups: Group-A (PEEK) copings and Group-B (zirconia) copings. All copings were fabricated using a CAD/CAM system. Vertical marginal accuracy was assessed with a stereomicroscope and image analysis software at ×20 magnification. The marginal gaP values were subjected to a student (independent) t-test for statistical analysis.</p><p><strong>Statistical analysis used: </strong>The statistical analysis involved a student (independent) t-test to compare the marginal gaP values between Group A (PEEK) and Group B (zirconia).</p><p><strong>Results: </strong>The mean marginal discrepancy for Group A (PEEK) and Group B (zirconia) was 92.84 μm ± 3.48 μm and 63.12 μm ± 31.47 μm, respectively. A statistically significant variation (t = 3.635, P = 0.001) between the groups was observed, indicating better marginal accuracy with zirconia copings compared to PEEK copings.</p><p><strong>Conclusion: </strong>Both PEEK and zirconia copings demonstrated vertical marginal discrepancies within the clinically acceptable limit of <120 μm. However, zirconia copings exhibited superior marginal accuracy in this in vitro study.</p>","PeriodicalId":22669,"journal":{"name":"The Journal of Indian Prosthodontic Society","volume":"25 1","pages":"95-101"},"PeriodicalIF":1.0000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Journal of Indian Prosthodontic Society","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jips.jips_8_24","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/3 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0
Abstract
Aim: The aim of this study was to compare the marginal accuracy of polyetheretherketone (PEEK) and zirconia copings fabricated using computer-aided design/computer-aided manufacturing (CAD/CAM) technology, and to assess the impact of their material properties on accuracy when produced with a 4-axis milling system under controlled conditions.
Settings and design: The study employed an in vitro design with a stainless steel die model featuring a 6 mm axial wall height, a 6-degree total occlusal convergence, and a radial shoulder finish line.
Materials and methods: Thirty stone dies were created from silicone impressions of the metal die and poured using type-IV dental stone. The dies were divided into two groups: Group-A (PEEK) copings and Group-B (zirconia) copings. All copings were fabricated using a CAD/CAM system. Vertical marginal accuracy was assessed with a stereomicroscope and image analysis software at ×20 magnification. The marginal gaP values were subjected to a student (independent) t-test for statistical analysis.
Statistical analysis used: The statistical analysis involved a student (independent) t-test to compare the marginal gaP values between Group A (PEEK) and Group B (zirconia).
Results: The mean marginal discrepancy for Group A (PEEK) and Group B (zirconia) was 92.84 μm ± 3.48 μm and 63.12 μm ± 31.47 μm, respectively. A statistically significant variation (t = 3.635, P = 0.001) between the groups was observed, indicating better marginal accuracy with zirconia copings compared to PEEK copings.