Reproducibility of liver iron concentration measured by R2* method based on 1.5-T MRI: a meta-analysis.

IF 4.7 2区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
European Radiology Pub Date : 2025-07-01 Epub Date: 2025-01-03 DOI:10.1007/s00330-024-11334-x
Sue Cao, Jie Zhang, Chenyu Dong, Yutao Que, Xin Gao, Ruomi Guo
{"title":"Reproducibility of liver iron concentration measured by R2* method based on 1.5-T MRI: a meta-analysis.","authors":"Sue Cao, Jie Zhang, Chenyu Dong, Yutao Que, Xin Gao, Ruomi Guo","doi":"10.1007/s00330-024-11334-x","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To explore the reproducibility of the 1.5-T MR imaging (MRI)-based R2* method in measuring the liver iron concentration (LIC) across different MRI scanners, scan parameters, and postprocessing techniques.</p><p><strong>Materials and methods: </strong>We performed a systematic search of the PubMed, Embase, Medline, Cochrane Library, and Web of Science databases and identified studies that used the 1.5-T MRI-based R2* method to measure the LIC. The original data were extracted from the selected studies. Reproducibility was assessed in terms of the ability to replicate the measured LIC with the 1.5-T MRI-based R2* method across different scanning instruments, scan parameters, and postprocessing techniques. We used the scanner equipment, scan parameters, and postprocessing technique as random effects to determine whether the R2* values obtained in the different studies were significantly different after the effects of the measured LIC were excluded. The calibration curve that best fit the R2* value to the LIC was estimated via univariate regression.</p><p><strong>Results: </strong>Twenty-one studies (1147 participants) were included. The seven studies (435 participants) in which the initial TE was ≤ 1.0 ms and the TE spacing was ≤ 1.4 ms did not have significant differences in the measured R2* value in pairwise comparisons (p > 0.05). Calibration of R2* to LIC was as follows: LIC (mg/g) = 0.042 + 2.85 × 10<sup>-2</sup> R2* (s<sup>-</sup><sup>1</sup>) (R<sup>2</sup> = 0.79).</p><p><strong>Conclusion: </strong>The 1.5-T MRI-based R2* method in estimating LIC was reproducible across different MRI scanners, scan parameters, and postprocessing techniques.</p><p><strong>Key points: </strong>Question The R2* method is widely used to quantify LIC, but the reproducibility of R2*-based LIC estimation remains unknown. Findings In scan sequences with an initial echo time (TE) ≤ 1.0 ms and a TE spacing ≤ 1.4 ms, the R2*-based LIC quantification at 1.5 T achieved good reproducibility. Clinical relevance The 1.5-T MRI-based R2* method enables reproducible quantification of the LIC; its reproducibility spares patients from unnecessary trauma and economic burdens while promoting the extensive clinical dissemination of R2*-based LIC quantification methods.</p>","PeriodicalId":12076,"journal":{"name":"European Radiology","volume":" ","pages":"3768-3780"},"PeriodicalIF":4.7000,"publicationDate":"2025-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00330-024-11334-x","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/3 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: To explore the reproducibility of the 1.5-T MR imaging (MRI)-based R2* method in measuring the liver iron concentration (LIC) across different MRI scanners, scan parameters, and postprocessing techniques.

Materials and methods: We performed a systematic search of the PubMed, Embase, Medline, Cochrane Library, and Web of Science databases and identified studies that used the 1.5-T MRI-based R2* method to measure the LIC. The original data were extracted from the selected studies. Reproducibility was assessed in terms of the ability to replicate the measured LIC with the 1.5-T MRI-based R2* method across different scanning instruments, scan parameters, and postprocessing techniques. We used the scanner equipment, scan parameters, and postprocessing technique as random effects to determine whether the R2* values obtained in the different studies were significantly different after the effects of the measured LIC were excluded. The calibration curve that best fit the R2* value to the LIC was estimated via univariate regression.

Results: Twenty-one studies (1147 participants) were included. The seven studies (435 participants) in which the initial TE was ≤ 1.0 ms and the TE spacing was ≤ 1.4 ms did not have significant differences in the measured R2* value in pairwise comparisons (p > 0.05). Calibration of R2* to LIC was as follows: LIC (mg/g) = 0.042 + 2.85 × 10-2 R2* (s-1) (R2 = 0.79).

Conclusion: The 1.5-T MRI-based R2* method in estimating LIC was reproducible across different MRI scanners, scan parameters, and postprocessing techniques.

Key points: Question The R2* method is widely used to quantify LIC, but the reproducibility of R2*-based LIC estimation remains unknown. Findings In scan sequences with an initial echo time (TE) ≤ 1.0 ms and a TE spacing ≤ 1.4 ms, the R2*-based LIC quantification at 1.5 T achieved good reproducibility. Clinical relevance The 1.5-T MRI-based R2* method enables reproducible quantification of the LIC; its reproducibility spares patients from unnecessary trauma and economic burdens while promoting the extensive clinical dissemination of R2*-based LIC quantification methods.

基于1.5 t MRI的R2*方法测量肝铁浓度的可重复性:荟萃分析。
目的:探讨基于1.5 t磁共振成像(MRI)的R2*方法在不同MRI扫描仪、扫描参数和后处理技术中测量肝铁浓度(LIC)的可重复性。材料和方法:我们对PubMed、Embase、Medline、Cochrane Library和Web of Science数据库进行了系统检索,并确定了使用基于1.5 t mri的R2*方法测量LIC的研究。原始数据是从选定的研究中提取的。通过使用基于1.5 t mri的R2*方法在不同的扫描仪器、扫描参数和后处理技术中复制所测LIC的能力来评估再现性。我们将扫描设备、扫描参数和后处理技术作为随机效应,在排除测量LIC的影响后,确定不同研究中获得的R2*值是否有显著差异。通过单变量回归估计出最适合R2*值与LIC的校准曲线。结果:纳入21项研究(1147名受试者)。初始TE≤1.0 ms、TE间隔≤1.4 ms的7项研究(435名受试者)在两两比较中测量的R2*值无显著差异(p < 0.05)。R2*对LIC的校准方法为:LIC (mg/g) = 0.042 + 2.85 × 10-2 R2* (s-1) (R2 = 0.79)。结论:基于1.5 t MRI的R2*方法在不同的MRI扫描仪、扫描参数和后处理技术中估计LIC是可重复的。R2*方法被广泛用于量化LIC,但基于R2*的LIC估计的可重复性尚不清楚。结果在初始回波时间(TE)≤1.0 ms、TE间隔≤1.4 ms的扫描序列中,基于R2*的1.5 T下的LIC定量具有良好的再现性。基于1.5 t mri的R2*方法可重复量化LIC;其可重复性使患者免于不必要的创伤和经济负担,同时促进了基于R2*的LIC量化方法的广泛临床推广。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
European Radiology
European Radiology 医学-核医学
CiteScore
11.60
自引率
8.50%
发文量
874
审稿时长
2-4 weeks
期刊介绍: European Radiology (ER) continuously updates scientific knowledge in radiology by publication of strong original articles and state-of-the-art reviews written by leading radiologists. A well balanced combination of review articles, original papers, short communications from European radiological congresses and information on society matters makes ER an indispensable source for current information in this field. This is the Journal of the European Society of Radiology, and the official journal of a number of societies. From 2004-2008 supplements to European Radiology were published under its companion, European Radiology Supplements, ISSN 1613-3749.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信