{"title":"Reproducibility of liver iron concentration measured by R2* method based on 1.5-T MRI: a meta-analysis.","authors":"Sue Cao, Jie Zhang, Chenyu Dong, Yutao Que, Xin Gao, Ruomi Guo","doi":"10.1007/s00330-024-11334-x","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To explore the reproducibility of the 1.5-T MR imaging (MRI)-based R2* method in measuring the liver iron concentration (LIC) across different MRI scanners, scan parameters, and postprocessing techniques.</p><p><strong>Materials and methods: </strong>We performed a systematic search of the PubMed, Embase, Medline, Cochrane Library, and Web of Science databases and identified studies that used the 1.5-T MRI-based R2* method to measure the LIC. The original data were extracted from the selected studies. Reproducibility was assessed in terms of the ability to replicate the measured LIC with the 1.5-T MRI-based R2* method across different scanning instruments, scan parameters, and postprocessing techniques. We used the scanner equipment, scan parameters, and postprocessing technique as random effects to determine whether the R2* values obtained in the different studies were significantly different after the effects of the measured LIC were excluded. The calibration curve that best fit the R2* value to the LIC was estimated via univariate regression.</p><p><strong>Results: </strong>Twenty-one studies (1147 participants) were included. The seven studies (435 participants) in which the initial TE was ≤ 1.0 ms and the TE spacing was ≤ 1.4 ms did not have significant differences in the measured R2* value in pairwise comparisons (p > 0.05). Calibration of R2* to LIC was as follows: LIC (mg/g) = 0.042 + 2.85 × 10<sup>-2</sup> R2* (s<sup>-</sup><sup>1</sup>) (R<sup>2</sup> = 0.79).</p><p><strong>Conclusion: </strong>The 1.5-T MRI-based R2* method in estimating LIC was reproducible across different MRI scanners, scan parameters, and postprocessing techniques.</p><p><strong>Key points: </strong>Question The R2* method is widely used to quantify LIC, but the reproducibility of R2*-based LIC estimation remains unknown. Findings In scan sequences with an initial echo time (TE) ≤ 1.0 ms and a TE spacing ≤ 1.4 ms, the R2*-based LIC quantification at 1.5 T achieved good reproducibility. Clinical relevance The 1.5-T MRI-based R2* method enables reproducible quantification of the LIC; its reproducibility spares patients from unnecessary trauma and economic burdens while promoting the extensive clinical dissemination of R2*-based LIC quantification methods.</p>","PeriodicalId":12076,"journal":{"name":"European Radiology","volume":" ","pages":"3768-3780"},"PeriodicalIF":4.7000,"publicationDate":"2025-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00330-024-11334-x","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/3 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: To explore the reproducibility of the 1.5-T MR imaging (MRI)-based R2* method in measuring the liver iron concentration (LIC) across different MRI scanners, scan parameters, and postprocessing techniques.
Materials and methods: We performed a systematic search of the PubMed, Embase, Medline, Cochrane Library, and Web of Science databases and identified studies that used the 1.5-T MRI-based R2* method to measure the LIC. The original data were extracted from the selected studies. Reproducibility was assessed in terms of the ability to replicate the measured LIC with the 1.5-T MRI-based R2* method across different scanning instruments, scan parameters, and postprocessing techniques. We used the scanner equipment, scan parameters, and postprocessing technique as random effects to determine whether the R2* values obtained in the different studies were significantly different after the effects of the measured LIC were excluded. The calibration curve that best fit the R2* value to the LIC was estimated via univariate regression.
Results: Twenty-one studies (1147 participants) were included. The seven studies (435 participants) in which the initial TE was ≤ 1.0 ms and the TE spacing was ≤ 1.4 ms did not have significant differences in the measured R2* value in pairwise comparisons (p > 0.05). Calibration of R2* to LIC was as follows: LIC (mg/g) = 0.042 + 2.85 × 10-2 R2* (s-1) (R2 = 0.79).
Conclusion: The 1.5-T MRI-based R2* method in estimating LIC was reproducible across different MRI scanners, scan parameters, and postprocessing techniques.
Key points: Question The R2* method is widely used to quantify LIC, but the reproducibility of R2*-based LIC estimation remains unknown. Findings In scan sequences with an initial echo time (TE) ≤ 1.0 ms and a TE spacing ≤ 1.4 ms, the R2*-based LIC quantification at 1.5 T achieved good reproducibility. Clinical relevance The 1.5-T MRI-based R2* method enables reproducible quantification of the LIC; its reproducibility spares patients from unnecessary trauma and economic burdens while promoting the extensive clinical dissemination of R2*-based LIC quantification methods.
期刊介绍:
European Radiology (ER) continuously updates scientific knowledge in radiology by publication of strong original articles and state-of-the-art reviews written by leading radiologists. A well balanced combination of review articles, original papers, short communications from European radiological congresses and information on society matters makes ER an indispensable source for current information in this field.