Evaluation of the analytic performance and macroprolactin sensitivity of a new prolactin immunoassay.

Guillaume David, Pauline Perrin, Camille Sergeant, Gérald Raverot, Véronique Raverot
{"title":"Evaluation of the analytic performance and macroprolactin sensitivity of a new prolactin immunoassay.","authors":"Guillaume David, Pauline Perrin, Camille Sergeant, Gérald Raverot, Véronique Raverot","doi":"10.1016/j.ando.2024.101677","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Prolactin measurement is essential in endocrine diagnostics. Challenges such as the hook effect and reactivity to macroprolactin, which varies according to the reagent, complicate accurate measurement. The present study evaluated a newly marketed reagent to detect prolactin, IDS Prolactin, comparing it to an established reagent, Roche Elecsys Prolactin, assessing its behavior toward macroprolactin and polyethylene glycol (PEG) treatment, and establishing reference intervals.</p><p><strong>Methods: </strong>The IDS Prolactin and Roche Elecsys Prolactin assays were compared using 44 samples containing macroprolactin confirmed on gel filtration chromatography (macroprolactin: BBPRL) and 104 samples for which the diagnosis of macroprolactin was excluded (monomeric prolactin: MNPRL). Analytic performance of the IDS Prolactin assay was also assessed.</p><p><strong>Results: </strong>The new reagent showed satisfactory analytic performance, meeting EFLM standards for repeatability and intermediate imprecision. Comparison between the two methods found robust correlation for monomeric samples (y=1.060x-18.28; r2=0.993). Compared to the Roche assay, which is particularly low in its reaction to macroprolactin, the IDS assay displayed a higher level of detection. PEG precipitation effectively separated monomeric and macroprolactin samples when a cut-off of 65% recovery was used or at the threshold of 444 mIU/L (20.9 µg/L) for post-PEG monomeric prolactin upper limit of normal. Reference intervals were established for women, with ROC curve analysis demonstrating high sensitivity and specificity.</p><p><strong>Conclusion: </strong>The IDS Prolactin assay showed excellent analytic performance and satisfactory characteristics on macroprolactinemic samples.</p>","PeriodicalId":93871,"journal":{"name":"Annales d'endocrinologie","volume":" ","pages":"101677"},"PeriodicalIF":0.0000,"publicationDate":"2024-12-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annales d'endocrinologie","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1016/j.ando.2024.101677","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Prolactin measurement is essential in endocrine diagnostics. Challenges such as the hook effect and reactivity to macroprolactin, which varies according to the reagent, complicate accurate measurement. The present study evaluated a newly marketed reagent to detect prolactin, IDS Prolactin, comparing it to an established reagent, Roche Elecsys Prolactin, assessing its behavior toward macroprolactin and polyethylene glycol (PEG) treatment, and establishing reference intervals.

Methods: The IDS Prolactin and Roche Elecsys Prolactin assays were compared using 44 samples containing macroprolactin confirmed on gel filtration chromatography (macroprolactin: BBPRL) and 104 samples for which the diagnosis of macroprolactin was excluded (monomeric prolactin: MNPRL). Analytic performance of the IDS Prolactin assay was also assessed.

Results: The new reagent showed satisfactory analytic performance, meeting EFLM standards for repeatability and intermediate imprecision. Comparison between the two methods found robust correlation for monomeric samples (y=1.060x-18.28; r2=0.993). Compared to the Roche assay, which is particularly low in its reaction to macroprolactin, the IDS assay displayed a higher level of detection. PEG precipitation effectively separated monomeric and macroprolactin samples when a cut-off of 65% recovery was used or at the threshold of 444 mIU/L (20.9 µg/L) for post-PEG monomeric prolactin upper limit of normal. Reference intervals were established for women, with ROC curve analysis demonstrating high sensitivity and specificity.

Conclusion: The IDS Prolactin assay showed excellent analytic performance and satisfactory characteristics on macroprolactinemic samples.

求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信