Proof of concept testing of a vascular closure device for use in fetal surgery.

IF 1.7 4区 医学 Q3 OBSTETRICS & GYNECOLOGY
Braxton Forde, Samuel Martin, Marc Oria, Jordan Kapke, Eyal Krispin, Jose L Peiro
{"title":"Proof of concept testing of a vascular closure device for use in fetal surgery.","authors":"Braxton Forde, Samuel Martin, Marc Oria, Jordan Kapke, Eyal Krispin, Jose L Peiro","doi":"10.1080/14767058.2024.2435468","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Prior clinical findings have demonstrated that maternal laparotomy with trans-amniotic trans-uterine suturing of the fetoscopic port site during in utero myelomeningocele repair reduces the risk of membrane rupture. However, due to laparotomy-associated morbidity, we aimed to explore the feasibility of using a vascular closure device for percutaneous trans-amniotic trans-uterine suturing.</p><p><strong>Methods: </strong>This IRB and IACUC-exempt study utilized 2 strategies for proof-of-concept testing of using the Abbott Perclose<sup>™</sup> ProStyle<sup>™</sup> Device for suture placement; 1. Ultrasound guided application on a high fidelity maternal abdominal uterus model used for fetal procedures and 2. Placement under direct visualization with sheep undergoing cesarean delivery for other research purposes. In the high-fidelity uterus model, the Abbott Perclose<sup>™</sup> ProStyle<sup>™</sup> device was used to place a transuterine/transamniotic stitch with accompanying video recording of the approach (https://go.screenpal.com/watch/cZfhoDVsYvW password: perclose). Regarding the second approach, at the time of a cesarean section, 12 French Checkflo<sup>®</sup> cannulas were inserted into the sheep amniotic space <i>via</i> different approaches: (1) Seldinger technique, (2) Seldinger technique insertion of Checkflo<sup>®</sup> cannula and subsequent use of the Abbott Perclose<sup>™</sup> ProStyle<sup>™</sup> device to suture the port site after check flow removal, (3) Abbott Perclose<sup>™</sup> ProStyle<sup>™</sup> device utilized in what is described as a \"pre-close\" technique, where prior to cannula placement, trans-uterine trans-amniotic stitches are placed followed by the insertion of a 12 French Checkflo<sup>®</sup> cannula over the same guidewire. Samples of the sutured uterine wall were sent to pathology and H&E staining was performed to assess uterine hole closure and amnion-to-uterus fixation.</p><p><strong>Results: </strong>The high-fidelity model confirmed that the Perclose<sup>™</sup> ProStyle<sup>™</sup> Device was easily visualized by ultrasound and suture deployment was without complication. In the animal model, the Perclose<sup>™</sup> ProStyle<sup>™</sup> device effectively sutured the amnion to the uterus in both the pre- and post-close approach. The pre-close technique achieved better amnion-to-uterus approximation and more appropriate uterine hole closure. H&E staining revealed that without suturing, amnion separation from the chorion layer occurred, and the uterine hole persisted. The post-close technique showed partial connection between the amnion and chorion, but inadequate uterine hole closure with amnion shift into the defect. Optimal closure, with secure amnion-to-chorion fixation and uterine closure, was achieved through the pre-close technique.</p><p><strong>Conclusion: </strong>The Abbott Perclose<sup>™</sup> ProStyle<sup>™</sup> Device seems to be a feasible device for use of uterine port closure in maternal-fetal surgery, larger animal studies with mid-pregnancy application are needed to further validate or refute these findings.</p>","PeriodicalId":50146,"journal":{"name":"Journal of Maternal-Fetal & Neonatal Medicine","volume":"38 1","pages":"2435468"},"PeriodicalIF":1.7000,"publicationDate":"2025-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Maternal-Fetal & Neonatal Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/14767058.2024.2435468","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/12/8 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"OBSTETRICS & GYNECOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: Prior clinical findings have demonstrated that maternal laparotomy with trans-amniotic trans-uterine suturing of the fetoscopic port site during in utero myelomeningocele repair reduces the risk of membrane rupture. However, due to laparotomy-associated morbidity, we aimed to explore the feasibility of using a vascular closure device for percutaneous trans-amniotic trans-uterine suturing.

Methods: This IRB and IACUC-exempt study utilized 2 strategies for proof-of-concept testing of using the Abbott Perclose ProStyle Device for suture placement; 1. Ultrasound guided application on a high fidelity maternal abdominal uterus model used for fetal procedures and 2. Placement under direct visualization with sheep undergoing cesarean delivery for other research purposes. In the high-fidelity uterus model, the Abbott Perclose ProStyle device was used to place a transuterine/transamniotic stitch with accompanying video recording of the approach (https://go.screenpal.com/watch/cZfhoDVsYvW password: perclose). Regarding the second approach, at the time of a cesarean section, 12 French Checkflo® cannulas were inserted into the sheep amniotic space via different approaches: (1) Seldinger technique, (2) Seldinger technique insertion of Checkflo® cannula and subsequent use of the Abbott Perclose ProStyle device to suture the port site after check flow removal, (3) Abbott Perclose ProStyle device utilized in what is described as a "pre-close" technique, where prior to cannula placement, trans-uterine trans-amniotic stitches are placed followed by the insertion of a 12 French Checkflo® cannula over the same guidewire. Samples of the sutured uterine wall were sent to pathology and H&E staining was performed to assess uterine hole closure and amnion-to-uterus fixation.

Results: The high-fidelity model confirmed that the Perclose ProStyle Device was easily visualized by ultrasound and suture deployment was without complication. In the animal model, the Perclose ProStyle device effectively sutured the amnion to the uterus in both the pre- and post-close approach. The pre-close technique achieved better amnion-to-uterus approximation and more appropriate uterine hole closure. H&E staining revealed that without suturing, amnion separation from the chorion layer occurred, and the uterine hole persisted. The post-close technique showed partial connection between the amnion and chorion, but inadequate uterine hole closure with amnion shift into the defect. Optimal closure, with secure amnion-to-chorion fixation and uterine closure, was achieved through the pre-close technique.

Conclusion: The Abbott Perclose ProStyle Device seems to be a feasible device for use of uterine port closure in maternal-fetal surgery, larger animal studies with mid-pregnancy application are needed to further validate or refute these findings.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
4.40
自引率
0.00%
发文量
217
审稿时长
2-3 weeks
期刊介绍: The official journal of The European Association of Perinatal Medicine, The Federation of Asia and Oceania Perinatal Societies and The International Society of Perinatal Obstetricians. The journal publishes a wide range of peer-reviewed research on the obstetric, medical, genetic, mental health and surgical complications of pregnancy and their effects on the mother, fetus and neonate. Research on audit, evaluation and clinical care in maternal-fetal and perinatal medicine is also featured.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信