Supervised Automated Kinetic Perimetry (SAKP) Using Simulated Visual Field Data - Presentation of a New Examination Technique.

IF 0.8 4区 医学 Q4 OPHTHALMOLOGY
Ulrich Schiefer, Michael Wörner, Ditta Zobor
{"title":"Supervised Automated Kinetic Perimetry (SAKP) Using Simulated Visual Field Data - Presentation of a New Examination Technique.","authors":"Ulrich Schiefer, Michael Wörner, Ditta Zobor","doi":"10.1055/a-2427-3556","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The aim of this study was to develop, optimise, train, and evaluate an algorithm for performing Supervised Automated Kinetic Perimetry (SAKP) using digitalised perimetric simulation data.</p><p><strong>Methods: </strong>The original SAKP algorithm was based on findings from a multicentre study to establish reference values by semi-automated kinetic perimetry (SKP) combined with an automated examination method with moving stimuli (\"Program K\", developed in Japan). The algorithm evaluated the outer angles of isopter segments and responded to deviations from expected values by placing examination vectors to measure the outer boundaries of the visual field (VF). Specialised interpolation methods were also used to create individual 3D hills of vision and local \"probing vectors\" to optimise the eccentricity of the vector origins. This algorithm was trained iteratively on seven representative digitalised 3D VF results from five typical classes and optimised in each step: (1) Normal VF, (2) Central scotoma, (3) Concentric VF constriction, (4) Retinal nerve fibre layer defects in the visual field (VFDs), (5) VFDs with respect to the vertical meridian. The optimised SAKP algorithm was then applied to a new set of twenty 3D VF results of varying origin and severity. The primary targets were measured in agreement between actual calculated VF expressed as accuracy (A), that is, the ratio between the area containing correct predictions and total area of predictions measured between 0 = worst and 1 = best, and examination duration (T). The results are given as median (and interquartile range). We also verified the test's robustness by varying individual error rates (ERs) and error magnitudes (EMs).</p><p><strong>Results: </strong>The median and interquartile range (IQR, in brackets) for the total of representative VFs were 0.93 (0.02) for A and 7.0 min (5.2 min) for T, respectively. A gave the best result for altitudinal VFDs and VFDs with hemianopic character and macular sparing (0.98 each) and worst in superior wedge-shaped VFDs (0.78); T was shortest in blind spot displacement (3.9 min) and longest in hemianopic VFDs with hemianopic character and macular sparing with preserved temporal crescent (12.1 min). Error rate and magnitude (up to 30% each) only showed a comparatively low influence on A and T.</p><p><strong>Conclusion: </strong>The SAKP algorithm presented here achieves a comparatively high degree of accuracy and robustness for actual, simulated visual field data within acceptable examination times. This algorithm is currently being prepared for application in real patient examinations under clinical conditions.</p>","PeriodicalId":17904,"journal":{"name":"Klinische Monatsblatter fur Augenheilkunde","volume":" ","pages":""},"PeriodicalIF":0.8000,"publicationDate":"2024-12-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Klinische Monatsblatter fur Augenheilkunde","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1055/a-2427-3556","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: The aim of this study was to develop, optimise, train, and evaluate an algorithm for performing Supervised Automated Kinetic Perimetry (SAKP) using digitalised perimetric simulation data.

Methods: The original SAKP algorithm was based on findings from a multicentre study to establish reference values by semi-automated kinetic perimetry (SKP) combined with an automated examination method with moving stimuli ("Program K", developed in Japan). The algorithm evaluated the outer angles of isopter segments and responded to deviations from expected values by placing examination vectors to measure the outer boundaries of the visual field (VF). Specialised interpolation methods were also used to create individual 3D hills of vision and local "probing vectors" to optimise the eccentricity of the vector origins. This algorithm was trained iteratively on seven representative digitalised 3D VF results from five typical classes and optimised in each step: (1) Normal VF, (2) Central scotoma, (3) Concentric VF constriction, (4) Retinal nerve fibre layer defects in the visual field (VFDs), (5) VFDs with respect to the vertical meridian. The optimised SAKP algorithm was then applied to a new set of twenty 3D VF results of varying origin and severity. The primary targets were measured in agreement between actual calculated VF expressed as accuracy (A), that is, the ratio between the area containing correct predictions and total area of predictions measured between 0 = worst and 1 = best, and examination duration (T). The results are given as median (and interquartile range). We also verified the test's robustness by varying individual error rates (ERs) and error magnitudes (EMs).

Results: The median and interquartile range (IQR, in brackets) for the total of representative VFs were 0.93 (0.02) for A and 7.0 min (5.2 min) for T, respectively. A gave the best result for altitudinal VFDs and VFDs with hemianopic character and macular sparing (0.98 each) and worst in superior wedge-shaped VFDs (0.78); T was shortest in blind spot displacement (3.9 min) and longest in hemianopic VFDs with hemianopic character and macular sparing with preserved temporal crescent (12.1 min). Error rate and magnitude (up to 30% each) only showed a comparatively low influence on A and T.

Conclusion: The SAKP algorithm presented here achieves a comparatively high degree of accuracy and robustness for actual, simulated visual field data within acceptable examination times. This algorithm is currently being prepared for application in real patient examinations under clinical conditions.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
1.30
自引率
0.00%
发文量
235
审稿时长
4-8 weeks
期刊介绍: -Konzentriertes Fachwissen aus Klinik und Praxis: Die entscheidenden Ergebnisse der internationalen Forschung - für Sie auf den Punkt gebracht und kritisch kommentiert, Übersichtsarbeiten zu den maßgeblichen Themen der täglichen Praxis, Top informiert - breite klinische Berichterstattung. -CME-Punkte sammeln mit dem Refresher: Effiziente, CME-zertifizierte Fortbildung, mit dem Refresher, 3 CME-Punkte pro Ausgabe - bis zu 36 CME-Punkte im Jahr!. -Aktuelle Rubriken mit echtem Nutzwert: Kurzreferate zu den wichtigsten Artikeln internationaler Zeitschriften, Schwerpunktthema in jedem Heft: Ausführliche Übersichtsarbeiten zu den wichtigsten Themen der Ophthalmologie – so behalten Sie das gesamte Fach im Blick!, Originalien mit den neuesten Entwicklungen, Übersichten zu den relevanten Themen.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信