Karsten Sperlich, Sebastian Bohn, Mario Gerlach, Julia Schubert, Heinrich Stolz, Rudolf Guthoff, Oliver Stachs
{"title":"In Vivo Acquisition of Human Retinal Double-Pass Images during Simulated Intraocular Lens Implantation.","authors":"Karsten Sperlich, Sebastian Bohn, Mario Gerlach, Julia Schubert, Heinrich Stolz, Rudolf Guthoff, Oliver Stachs","doi":"10.1055/a-2453-8253","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The aim of the study was to capture images that form on the human retina after the simulated implantation of an intraocular lens (IOL). White light was used rather than the commonly used near-infrared light, which is unsuitable for the examination of diffractive IOLs. For this purpose, a special optical setup was developed to investigate the influence of the IOL design on two-dimensional retinal images <i>in vivo</i>.</p><p><strong>Materials and methods: </strong>A double-pass ophthalmoscopic setup with a scientific CCD camera system was developed. Imaging the retinal image of a white LED located at infinity provides access to the double-pass point spread function of the natural eye. Subsequently, a see-through device for simulated IOL implantation (VirtIOL, 10Lens S. L. U., Terrassa, Spain) was integrated to investigate the influence of the IOL design on the retinal image quality of complex scenarios.</p><p><strong>Results: </strong>Retinal images were acquired from an incoherent white point light source. Combined with simulated IOL implantation, retinal images were acquired from the point light source, letters, and a United States Airforce target on a 6-m distant monitor. As expected, the double-pass images obtained with a monofocal IOL were sharper than those obtained with a multifocal IOL.</p><p><strong>Conclusion: </strong>The method opens up access to double-pass point spread function for white light, thus solving the problem of infrared light-based methods providing incorrect results when examining diffractive IOLs. This approach may be helpful for the investigation of perception in the future.</p>","PeriodicalId":17904,"journal":{"name":"Klinische Monatsblatter fur Augenheilkunde","volume":"241 12","pages":"1302-1308"},"PeriodicalIF":0.8000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Klinische Monatsblatter fur Augenheilkunde","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1055/a-2453-8253","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/12/5 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The aim of the study was to capture images that form on the human retina after the simulated implantation of an intraocular lens (IOL). White light was used rather than the commonly used near-infrared light, which is unsuitable for the examination of diffractive IOLs. For this purpose, a special optical setup was developed to investigate the influence of the IOL design on two-dimensional retinal images in vivo.
Materials and methods: A double-pass ophthalmoscopic setup with a scientific CCD camera system was developed. Imaging the retinal image of a white LED located at infinity provides access to the double-pass point spread function of the natural eye. Subsequently, a see-through device for simulated IOL implantation (VirtIOL, 10Lens S. L. U., Terrassa, Spain) was integrated to investigate the influence of the IOL design on the retinal image quality of complex scenarios.
Results: Retinal images were acquired from an incoherent white point light source. Combined with simulated IOL implantation, retinal images were acquired from the point light source, letters, and a United States Airforce target on a 6-m distant monitor. As expected, the double-pass images obtained with a monofocal IOL were sharper than those obtained with a multifocal IOL.
Conclusion: The method opens up access to double-pass point spread function for white light, thus solving the problem of infrared light-based methods providing incorrect results when examining diffractive IOLs. This approach may be helpful for the investigation of perception in the future.
期刊介绍:
-Konzentriertes Fachwissen aus Klinik und Praxis:
Die entscheidenden Ergebnisse der internationalen Forschung - für Sie auf den Punkt gebracht und kritisch kommentiert,
Übersichtsarbeiten zu den maßgeblichen Themen der täglichen Praxis,
Top informiert - breite klinische Berichterstattung.
-CME-Punkte sammeln mit dem Refresher:
Effiziente, CME-zertifizierte Fortbildung, mit dem Refresher,
3 CME-Punkte pro Ausgabe - bis zu 36 CME-Punkte im Jahr!.
-Aktuelle Rubriken mit echtem Nutzwert:
Kurzreferate zu den wichtigsten Artikeln internationaler Zeitschriften,
Schwerpunktthema in jedem Heft: Ausführliche Übersichtsarbeiten zu den wichtigsten Themen der Ophthalmologie – so behalten Sie das gesamte Fach im Blick!,
Originalien mit den neuesten Entwicklungen,
Übersichten zu den relevanten Themen.