Reliability of isokinetic dynamometer for isometric assessment of ankle plantar flexor strength

IF 2.2 3区 医学 Q1 REHABILITATION
Fanny-May Santy , Anthony Pernoud , Simon Barrué-Belou , François Fourchet , Hugo Bothorel , Pierre Samozino
{"title":"Reliability of isokinetic dynamometer for isometric assessment of ankle plantar flexor strength","authors":"Fanny-May Santy ,&nbsp;Anthony Pernoud ,&nbsp;Simon Barrué-Belou ,&nbsp;François Fourchet ,&nbsp;Hugo Bothorel ,&nbsp;Pierre Samozino","doi":"10.1016/j.ptsp.2024.11.001","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><div>To evaluate isokinetic dynamometer reliability for isometric assessment of plantar flexor (PF) strength.</div></div><div><h3>Design</h3><div>Cross-sectional.</div></div><div><h3>Setting</h3><div>Testing by the same physiotherapist twice during a first session (repeatability) and once during a second session (reproducibility).</div></div><div><h3>Participants</h3><div>Twenty-two healthy subjects (44 ankles, 11 men/11 women).</div></div><div><h3>Main outcome measures</h3><div>Isometric PF peak torque, with and without body mass normalization, at 0° and +20° of plantar flexion. Measurement reliability was evaluated using intraclass correlation coefficient (ICC), standard error of measurement (SEM) and minimal detectable change (MDC).</div></div><div><h3>Results</h3><div>Without normalization, measurement repeatability was excellent at 0° of plantar flexion (ICC, 0.94; SEM, 6.6%; MDC, 18.4%) compared with good repeatability at +20° (ICC, 0.85; SEM, 11.1%; MDC 30.6%). Measurement repeatability following normalization was good at 0° (ICC, 0.88; SEM, 5.2%; MDC, 14.4%) and +20° (ICC, 0.79; SEM, 10.2%; MDC, 28.1%). While reproducibility was good at 0° with normalization (ICC, 0.84; SEM, 5.9%; MDC, 16.3%) or excellent without (ICC 0.92; SEM 7.5%; MDC, 20.8%), it was moderate at +20° with normalization (ICC 0.71; SEM 11.3%; MDC, 31.3%) or good without (ICC 0.78; SEM 13.0%; MDC, 36.1%).</div></div><div><h3>Conclusion</h3><div>The reliability of PF maximal isometric strength is good/excellent at 0° of plantar flexion but moderate/good at +20°.</div></div>","PeriodicalId":49698,"journal":{"name":"Physical Therapy in Sport","volume":"71 ","pages":"Pages 36-42"},"PeriodicalIF":2.2000,"publicationDate":"2024-11-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physical Therapy in Sport","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1466853X24001147","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"REHABILITATION","Score":null,"Total":0}
引用次数: 0

Abstract

Objective

To evaluate isokinetic dynamometer reliability for isometric assessment of plantar flexor (PF) strength.

Design

Cross-sectional.

Setting

Testing by the same physiotherapist twice during a first session (repeatability) and once during a second session (reproducibility).

Participants

Twenty-two healthy subjects (44 ankles, 11 men/11 women).

Main outcome measures

Isometric PF peak torque, with and without body mass normalization, at 0° and +20° of plantar flexion. Measurement reliability was evaluated using intraclass correlation coefficient (ICC), standard error of measurement (SEM) and minimal detectable change (MDC).

Results

Without normalization, measurement repeatability was excellent at 0° of plantar flexion (ICC, 0.94; SEM, 6.6%; MDC, 18.4%) compared with good repeatability at +20° (ICC, 0.85; SEM, 11.1%; MDC 30.6%). Measurement repeatability following normalization was good at 0° (ICC, 0.88; SEM, 5.2%; MDC, 14.4%) and +20° (ICC, 0.79; SEM, 10.2%; MDC, 28.1%). While reproducibility was good at 0° with normalization (ICC, 0.84; SEM, 5.9%; MDC, 16.3%) or excellent without (ICC 0.92; SEM 7.5%; MDC, 20.8%), it was moderate at +20° with normalization (ICC 0.71; SEM 11.3%; MDC, 31.3%) or good without (ICC 0.78; SEM 13.0%; MDC, 36.1%).

Conclusion

The reliability of PF maximal isometric strength is good/excellent at 0° of plantar flexion but moderate/good at +20°.
求助全文
约1分钟内获得全文 求助全文
来源期刊
Physical Therapy in Sport
Physical Therapy in Sport 医学-康复医学
CiteScore
4.50
自引率
8.30%
发文量
125
审稿时长
39 days
期刊介绍: Physical Therapy in Sport is an international peer-reviewed journal that provides a forum for the publication of research and clinical practice material relevant to the healthcare professions involved in sports and exercise medicine, and rehabilitation. The journal publishes material that is indispensable for day-to-day practice and continuing professional development. Physical Therapy in Sport covers topics dealing with the diagnosis, treatment, and prevention of injuries, as well as more general areas of sports and exercise medicine and related sports science. The journal publishes original research, case studies, reviews, masterclasses, papers on clinical approaches, and book reviews, as well as occasional reports from conferences. Papers are double-blind peer-reviewed by our international advisory board and other international experts, and submissions from a broad range of disciplines are actively encouraged.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信