LC-MS/MS confirmation of 11-nor-9-carboxy-tetrahydrocannabinol (Δ8, Δ9, Δ1°) and hexahydrocannabinol (HHC) metabolites in authentic urine specimens.

IF 2.3 3区 医学 Q3 CHEMISTRY, ANALYTICAL
Amy L Patton, Luette Muir, Joshua Z Seither, Jeffrey P Walterscheid, Erin L Karschner
{"title":"LC-MS/MS confirmation of 11-nor-9-carboxy-tetrahydrocannabinol (Δ8, Δ9, Δ1°) and hexahydrocannabinol (HHC) metabolites in authentic urine specimens.","authors":"Amy L Patton, Luette Muir, Joshua Z Seither, Jeffrey P Walterscheid, Erin L Karschner","doi":"10.1093/jat/bkae091","DOIUrl":null,"url":null,"abstract":"<p><p>Recently, tetrahydrocannabinol (THC) isomers and other semi-synthetic cannabinoids have been introduced into the consumer market as alternatives to botanical cannabis. To assess the prevalence of these potential new analytical targets, a liquid chromatography-tandem mass spectrometry confirmation method was developed for the quantitation of seven cannabinoid metabolites and the qualitative identification of four others in urine. The validated method was applied to authentic urine specimens that screened positive by immunoassay (50 ng/mL cutoff; n=1300). The most commonly observed analytes were 11-nor-9-carboxy-Δ8- and Δ9-THC (Δ8- and Δ9-THCCOOH), with the combination of the two seen as the most prominent analyte combination found. In addition to these metabolites, Δ1°-THCCOOH was observed in 77 specimens. This is the first study to report Δ1°-THCCOOH in authentic urine specimens, with this analyte always appearing in combination with Δ9-THCCOOH. Cross-reactivity studies were performed for (6aR,9R)-Δ1°-THCCOOH using the Beckman Coulter Emit® II Plus Cannabinoid immunoassay and demonstrated cross reactivity equivalent to the Δ9-THCCOOH cutoff, providing added confidence in the reported prevalence and detection patterns. Additionally, 11-nor-9(R)-carboxy-hexahydrocannabinol (9(R)-HHCCOOH) was the most abundant stereoisomer (n=12) in specimens containing HHC metabolites alone (n=14). This is in contrast to 9(S)-HHCCOOH, which was the predominant stereoisomer in specimens containing Δ8- and/or Δ9-THCCOOH. Although HHC and Δ1°-THC metabolites are emerging toxicology findings, based on these specimens collected between April 2022 and May 2024, an analytical panel containing Δ8- and Δ9-THCCOOH appears to be sufficient for revealing cannabinoid exposure within workplace monitoring and deterrence programs.</p>","PeriodicalId":14905,"journal":{"name":"Journal of analytical toxicology","volume":" ","pages":""},"PeriodicalIF":2.3000,"publicationDate":"2024-11-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of analytical toxicology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/jat/bkae091","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Recently, tetrahydrocannabinol (THC) isomers and other semi-synthetic cannabinoids have been introduced into the consumer market as alternatives to botanical cannabis. To assess the prevalence of these potential new analytical targets, a liquid chromatography-tandem mass spectrometry confirmation method was developed for the quantitation of seven cannabinoid metabolites and the qualitative identification of four others in urine. The validated method was applied to authentic urine specimens that screened positive by immunoassay (50 ng/mL cutoff; n=1300). The most commonly observed analytes were 11-nor-9-carboxy-Δ8- and Δ9-THC (Δ8- and Δ9-THCCOOH), with the combination of the two seen as the most prominent analyte combination found. In addition to these metabolites, Δ1°-THCCOOH was observed in 77 specimens. This is the first study to report Δ1°-THCCOOH in authentic urine specimens, with this analyte always appearing in combination with Δ9-THCCOOH. Cross-reactivity studies were performed for (6aR,9R)-Δ1°-THCCOOH using the Beckman Coulter Emit® II Plus Cannabinoid immunoassay and demonstrated cross reactivity equivalent to the Δ9-THCCOOH cutoff, providing added confidence in the reported prevalence and detection patterns. Additionally, 11-nor-9(R)-carboxy-hexahydrocannabinol (9(R)-HHCCOOH) was the most abundant stereoisomer (n=12) in specimens containing HHC metabolites alone (n=14). This is in contrast to 9(S)-HHCCOOH, which was the predominant stereoisomer in specimens containing Δ8- and/or Δ9-THCCOOH. Although HHC and Δ1°-THC metabolites are emerging toxicology findings, based on these specimens collected between April 2022 and May 2024, an analytical panel containing Δ8- and Δ9-THCCOOH appears to be sufficient for revealing cannabinoid exposure within workplace monitoring and deterrence programs.

LC-MS/MS 确认真实尿液样本中的 11-去甲-9-羧基四氢大麻酚(Δ8、Δ9、Δ1°)和六氢大麻酚(HHC)代谢物。
最近,四氢大麻酚(THC)异构体和其他半合成大麻素作为植物大麻的替代品被引入消费市场。为了评估这些潜在的新分析目标的普遍性,我们开发了一种液相色谱-串联质谱确认方法,用于定量检测尿液中的七种大麻素代谢物和定性鉴定其他四种代谢物。验证方法适用于经免疫测定呈阳性的真实尿液标本(50 纳克/毫升临界值;n=1300)。最常见的分析物是 11-去甲-9-羧基-Δ8-和Δ9-THC(Δ8-和Δ9-THCCOOH),两者的组合是发现的最主要的分析物组合。除这些代谢物外,在 77 份标本中还发现了 Δ1°-THCCOOH。这是首次在真实尿液标本中检测到 Δ1°-THCCOOH,该分析物总是与 Δ9-THCCOOH同时出现。使用贝克曼库尔特 Emit® II Plus 大麻免疫分析仪对 (6aR,9R)-Δ1°-THCCOOH 进行了交叉反应研究,结果表明其交叉反应性与Δ9-THCCOOH 临界值相当,为报告的流行率和检测模式增加了可信度。此外,11-去甲-9(R)-羧基六氢大麻酚(9(R)-HHCCOOH)是单独含有 HHC 代谢物的标本(14 份)中含量最高的立体异构体(12 份)。而在含有 Δ8- 和/或 Δ9-THCCOOH 的标本中,9(S)-HHCCOOH 是最主要的立体异构体。虽然 HHC 和 Δ1°-THC 代谢物是新出现的毒理学发现,但根据 2022 年 4 月至 2024 年 5 月期间采集的这些样本,包含 Δ8- 和 Δ9-THCCOOH 的分析面板似乎足以揭示工作场所监测和威慑计划中的大麻素暴露情况。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.10
自引率
20.00%
发文量
92
审稿时长
6-12 weeks
期刊介绍: The Journal of Analytical Toxicology (JAT) is an international toxicology journal devoted to the timely dissemination of scientific communications concerning potentially toxic substances and drug identification, isolation, and quantitation. Since its inception in 1977, the Journal of Analytical Toxicology has striven to present state-of-the-art techniques used in toxicology labs. The peer-review process provided by the distinguished members of the Editorial Advisory Board ensures the high-quality and integrity of articles published in the Journal of Analytical Toxicology. Timely presentation of the latest toxicology developments is ensured through Technical Notes, Case Reports, and Letters to the Editor.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信