Reverse Phase-High-Performance Liquid Chromatography (RP-HPLC) Method Development and Validation Using Analytical Quality-by-Design Approach for Determination of Isoliquiritigenin in Bulk and Biological Sample.
Mohamed Nihal P, Debasish Mohapatra, Alam Mohd Adil Alam Manir, Vancha Harish, Sachin Kumar Singh, Sakshi Upendra Lad, Srinivas Sutrapu, Sumant Saini, Sharfuddin Mohd
{"title":"Reverse Phase-High-Performance Liquid Chromatography (RP-HPLC) Method Development and Validation Using Analytical Quality-by-Design Approach for Determination of Isoliquiritigenin in Bulk and Biological Sample.","authors":"Mohamed Nihal P, Debasish Mohapatra, Alam Mohd Adil Alam Manir, Vancha Harish, Sachin Kumar Singh, Sakshi Upendra Lad, Srinivas Sutrapu, Sumant Saini, Sharfuddin Mohd","doi":"10.1089/adt.2024.050","DOIUrl":null,"url":null,"abstract":"<p><p>The primary objective of the present investigation is to develop and validate a simple, robust, and cost-effective isocratic reverse phase-high-performance liquid chromatography (RP-HPLC) method for determining isoliquiritigenin (ISL) in both bulk and biological samples using an analytical quality-by-design (AQbD) approach. The central composite design was employed for method optimization using Design Expert® software, by taking mobile phase ratio and flow rate as independent variables and peak area, retention time, tailing factor, and theoretical plates as dependent variables. The design suggested the use of a mobile phase consisting of acetonitrile:0.2% ortho-phosphoric acid (75:25, v/v) and a flow rate of 0.9 mL/min as optimal chromatographic conditions. The detection of ISL was performed at 364 nm. The optimized method was validated in accordance with International Conference on Harmonization (ICH) Q2(R1) guidelines. The method showed excellent linearity, limit of detection, limit of quantification, accuracy, precision, robustness, and system suitability. All validation parameters fell within the acceptable limits set by ICH. Additionally, the applicability of the method in biological samples were analyzed. In conclusion, the results suggest that the developed and validated AQbD-based RP-HPLC method was well-suited for the estimation of ISL in bulk and biological sample.</p>","PeriodicalId":8586,"journal":{"name":"Assay and drug development technologies","volume":" ","pages":"409-424"},"PeriodicalIF":1.6000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Assay and drug development technologies","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1089/adt.2024.050","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/11/25 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0
Abstract
The primary objective of the present investigation is to develop and validate a simple, robust, and cost-effective isocratic reverse phase-high-performance liquid chromatography (RP-HPLC) method for determining isoliquiritigenin (ISL) in both bulk and biological samples using an analytical quality-by-design (AQbD) approach. The central composite design was employed for method optimization using Design Expert® software, by taking mobile phase ratio and flow rate as independent variables and peak area, retention time, tailing factor, and theoretical plates as dependent variables. The design suggested the use of a mobile phase consisting of acetonitrile:0.2% ortho-phosphoric acid (75:25, v/v) and a flow rate of 0.9 mL/min as optimal chromatographic conditions. The detection of ISL was performed at 364 nm. The optimized method was validated in accordance with International Conference on Harmonization (ICH) Q2(R1) guidelines. The method showed excellent linearity, limit of detection, limit of quantification, accuracy, precision, robustness, and system suitability. All validation parameters fell within the acceptable limits set by ICH. Additionally, the applicability of the method in biological samples were analyzed. In conclusion, the results suggest that the developed and validated AQbD-based RP-HPLC method was well-suited for the estimation of ISL in bulk and biological sample.
期刊介绍:
ASSAY and Drug Development Technologies provides access to novel techniques and robust tools that enable critical advances in early-stage screening. This research published in the Journal leads to important therapeutics and platforms for drug discovery and development. This reputable peer-reviewed journal features original papers application-oriented technology reviews, topical issues on novel and burgeoning areas of research, and reports in methodology and technology application.
ASSAY and Drug Development Technologies coverage includes:
-Assay design, target development, and high-throughput technologies-
Hit to Lead optimization and medicinal chemistry through preclinical candidate selection-
Lab automation, sample management, bioinformatics, data mining, virtual screening, and data analysis-
Approaches to assays configured for gene families, inherited, and infectious diseases-
Assays and strategies for adapting model organisms to drug discovery-
The use of stem cells as models of disease-
Translation of phenotypic outputs to target identification-
Exploration and mechanistic studies of the technical basis for assay and screening artifacts