Should respiratory rate be measured on admission over a few seconds by a smartphone application or over 15 minutes by a piezoelectric device? A prospective observational pilot study of acutely ill medical patients admitted to a low-resource Ugandan hospital.
Franck Katembo Sikakulya, Immaculate Nakitende, Joan Nabiryo, Sylivia Namuleme, Alfred Lumala, John Kellett
{"title":"Should respiratory rate be measured on admission over a few seconds by a smartphone application or over 15 minutes by a piezoelectric device? A prospective observational pilot study of acutely ill medical patients admitted to a low-resource Ugandan hospital.","authors":"Franck Katembo Sikakulya, Immaculate Nakitende, Joan Nabiryo, Sylivia Namuleme, Alfred Lumala, John Kellett","doi":"10.52964/AMJA.0986","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>the optimal period over which respiratory rate should be measured is uncertain.</p><p><strong>Setting: </strong>medical ward of low resource Ugandan hospital, Methods: comparison of the first respiratory rates measured on admission over seconds using a smartphone application with measurements over 15 minutes using a piezoelectric device.</p><p><strong>Results: </strong>There was a poor correlation between respiratory rates measured by the piezoelectric device and rates measured by a smartphone application. Although the discrimination for mortality of piezoelectric respiratory rates were slightly higher than those derived from the application, there were not statistically significant differences.</p><p><strong>Conclusion: </strong>this study could not demonstrate potential clinical benefits that justify measuring the first respiratory rate after admission over 15 minutes by a device.</p>","PeriodicalId":39743,"journal":{"name":"Acute Medicine","volume":"23 3","pages":"118-126"},"PeriodicalIF":0.0000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acute Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.52964/AMJA.0986","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Background: the optimal period over which respiratory rate should be measured is uncertain.
Setting: medical ward of low resource Ugandan hospital, Methods: comparison of the first respiratory rates measured on admission over seconds using a smartphone application with measurements over 15 minutes using a piezoelectric device.
Results: There was a poor correlation between respiratory rates measured by the piezoelectric device and rates measured by a smartphone application. Although the discrimination for mortality of piezoelectric respiratory rates were slightly higher than those derived from the application, there were not statistically significant differences.
Conclusion: this study could not demonstrate potential clinical benefits that justify measuring the first respiratory rate after admission over 15 minutes by a device.
期刊介绍:
These are usually commissioned by the editorial team in accordance with a cycle running over several years. Authors wishing to submit a review relevant to Acute Medicine are advised to contact the editor before writing this. Unsolicited review articles received for consideration may be included if the subject matter is considered of interest to the readership, provided the topic has not already been covered in a recent edition. Review articles are usually 3000-5000 words and may include tables, pictures and other figures as required for the text. Include 3 or 4 ‘key points’ summarising the main teaching messages.