Mo Wang , Jie Shi , Yichuan Song, Shunli Zhang, Rui Zhang
{"title":"An accurate measurement method for serum homocysteine measurement by ID-LC/MS/MS and the application of external quality assessment","authors":"Mo Wang , Jie Shi , Yichuan Song, Shunli Zhang, Rui Zhang","doi":"10.1016/j.cca.2024.120021","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Serum homocysteine (Hcy) measurement accuracy is essential for detecting and diagnosing cardio-cerebrovascular illnesses. Although many different methods have been developed to determine the concentration of Hcy, those different assays showed nonequivalent results. This study aimed to develop an accurate and precise isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC/MS/MS) method for serum homocysteine (Hcy) and assign values for reference materials used in external quality assessment (EQA) programs.</div></div><div><h3>Methods</h3><div>Different concentrations of homocysteine calibration solutions were prepared with d4-Hcy as the internal standard. This method employed DTT to reduce protein-bound Hcy, followed by protein precipitation, and gradient elution on a Supelcosil LC-CN column for chromatographic separation, and multiple reaction monitor (MRM) mode with electrospray ionization (ESI) for mass spectrometric detection. After optimized ID-LC/MS/MS parameters, imprecision, trueness, the limit of quantification (LoQ), the limit of detection (LoD), and measurement uncertainty were evaluated to check the methodological performance. The established method was employed in assigning values to EQA samples, which were sent to 63 clinical laboratories in Beijing to measure Hcy.</div></div><div><h3>Results</h3><div>At 10.69,15.99, and 37.80 μmol/L levels, the present method demonstrated analytical imprecision of 0.57 %, 0.65 %, and 0.57 %, and recoveries of 99.67 % to 100.21 %, respectively. The bias between the target values of GBW (Guojia Biaozhun Wuzhi) were − 0.79 % to 0.62 %. The LoQ and LoD for Hcy were 0.36 nmol/L and 0.27 nmol/L. The method had an uncertainty (U 95 %) of 1.34 % to 1.48 %. For the three levels of EQA samples, the percentage of laboratories meeting the trueness evaluation criteria (±10 %) was 81.67 %, 83.33 %, and 71.67 % respectively.</div></div><div><h3>Conclusions</h3><div>With optimal method precision and trueness, the ID-LC/MS/MS method to measure serum Hcy can be used for value assignment of EQA samples, which can provide reliable data for monitoring the accuracy of clinical laboratory for Hcy measurement.</div></div>","PeriodicalId":10205,"journal":{"name":"Clinica Chimica Acta","volume":null,"pages":null},"PeriodicalIF":3.2000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinica Chimica Acta","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0009898124022745","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Serum homocysteine (Hcy) measurement accuracy is essential for detecting and diagnosing cardio-cerebrovascular illnesses. Although many different methods have been developed to determine the concentration of Hcy, those different assays showed nonequivalent results. This study aimed to develop an accurate and precise isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC/MS/MS) method for serum homocysteine (Hcy) and assign values for reference materials used in external quality assessment (EQA) programs.
Methods
Different concentrations of homocysteine calibration solutions were prepared with d4-Hcy as the internal standard. This method employed DTT to reduce protein-bound Hcy, followed by protein precipitation, and gradient elution on a Supelcosil LC-CN column for chromatographic separation, and multiple reaction monitor (MRM) mode with electrospray ionization (ESI) for mass spectrometric detection. After optimized ID-LC/MS/MS parameters, imprecision, trueness, the limit of quantification (LoQ), the limit of detection (LoD), and measurement uncertainty were evaluated to check the methodological performance. The established method was employed in assigning values to EQA samples, which were sent to 63 clinical laboratories in Beijing to measure Hcy.
Results
At 10.69,15.99, and 37.80 μmol/L levels, the present method demonstrated analytical imprecision of 0.57 %, 0.65 %, and 0.57 %, and recoveries of 99.67 % to 100.21 %, respectively. The bias between the target values of GBW (Guojia Biaozhun Wuzhi) were − 0.79 % to 0.62 %. The LoQ and LoD for Hcy were 0.36 nmol/L and 0.27 nmol/L. The method had an uncertainty (U 95 %) of 1.34 % to 1.48 %. For the three levels of EQA samples, the percentage of laboratories meeting the trueness evaluation criteria (±10 %) was 81.67 %, 83.33 %, and 71.67 % respectively.
Conclusions
With optimal method precision and trueness, the ID-LC/MS/MS method to measure serum Hcy can be used for value assignment of EQA samples, which can provide reliable data for monitoring the accuracy of clinical laboratory for Hcy measurement.
期刊介绍:
The Official Journal of the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)
Clinica Chimica Acta is a high-quality journal which publishes original Research Communications in the field of clinical chemistry and laboratory medicine, defined as the diagnostic application of chemistry, biochemistry, immunochemistry, biochemical aspects of hematology, toxicology, and molecular biology to the study of human disease in body fluids and cells.
The objective of the journal is to publish novel information leading to a better understanding of biological mechanisms of human diseases, their prevention, diagnosis, and patient management. Reports of an applied clinical character are also welcome. Papers concerned with normal metabolic processes or with constituents of normal cells or body fluids, such as reports of experimental or clinical studies in animals, are only considered when they are clearly and directly relevant to human disease. Evaluation of commercial products have a low priority for publication, unless they are novel or represent a technological breakthrough. Studies dealing with effects of drugs and natural products and studies dealing with the redox status in various diseases are not within the journal''s scope. Development and evaluation of novel analytical methodologies where applicable to diagnostic clinical chemistry and laboratory medicine, including point-of-care testing, and topics on laboratory management and informatics will also be considered. Studies focused on emerging diagnostic technologies and (big) data analysis procedures including digitalization, mobile Health, and artificial Intelligence applied to Laboratory Medicine are also of interest.