Alex Lua, Lynnette R L Tan, Frédéric Panthier, Alba Sierra, Eugenio Ventimiglia, Catalina Solano, Vincent De Coninck, Steffi Kar Kei Yuen, Vineet Gauhar, Olivier Traxer, Etienne Xavier Keller, Jia-Lun Kwok
{"title":"Optimal deflection techniques for flexible and navigable suction ureteral access sheaths (FANS): a comparative in vitro PEARLS analysis.","authors":"Alex Lua, Lynnette R L Tan, Frédéric Panthier, Alba Sierra, Eugenio Ventimiglia, Catalina Solano, Vincent De Coninck, Steffi Kar Kei Yuen, Vineet Gauhar, Olivier Traxer, Etienne Xavier Keller, Jia-Lun Kwok","doi":"10.1007/s00345-024-05297-3","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Flexible and navigable suction ureteral access sheaths (FANS) have been introduced without current evidence on how to optimize deflection. Aim was to evaluate in vitro deflection angles with 2 different FANS techniques-sheath advancement and ureteroscope deflection; and effects of sheath size-ureteroscope combinations.</p><p><strong>Methods: </strong>We evaluated in vitro deflection angles of 10/12Fr, 11/13Fr and 12/14Fr FANS (Hunan Reborn Medical Co. Ltd) with six single-use flexible ureteroscopes (Pusen Uscope 7.5Fr, OTU WiScope 7.5Fr, OTU WiScope 8.6Fr, Innovex EU-scope 8.7Fr, Red Pine RP-U-C12 8.7Fr and Boston Scientific Lithovue 9.5Fr). Two deflection techniques were tested: (1) sheath advancement-advancing the sheath forward over a maximally deflected ureteroscope, and (2) ureteroscope deflection-maximally deflecting the ureteroscope from various starting positions relative to tip of the sheath.</p><p><strong>Results: </strong>Intra and inter-scope comparisons of maximum deflection angles were significantly different (all ANOVA p < 0.01). Largest maximum angles for all ureteroscopes were with the sheath advancement technique (range 218°-277°), and second largest for most scopes using the ureteroscope deflection technique at tip (range 111°-212°), mostly deviating from manufacturer specifications (range 270°-275°). 10/12Fr and 11/13Fr sheath sizes were more flexible than 12/14Fr. Largest angles were with 11/13Fr sheath-OTU8.6Fr/Innovex8.7Fr combinations.</p><p><strong>Conclusion: </strong>Optimal deflection with FANS is achieved using either a sheath advancement technique, or ureteroscope deflection technique at tip. Despite using these optimized techniques, deflection angles specified by manufacturers seem hardly achievable. The sheath advancement technique and 11/13Fr sheath-OTU8.6Fr/Innovex8.7Fr combinations may be better suited for lower pole situations. Urologists should be aware of these differences and apply the findings to their FANS technique.</p>","PeriodicalId":23954,"journal":{"name":"World Journal of Urology","volume":"42 1","pages":"606"},"PeriodicalIF":2.8000,"publicationDate":"2024-10-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"World Journal of Urology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00345-024-05297-3","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"UROLOGY & NEPHROLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Flexible and navigable suction ureteral access sheaths (FANS) have been introduced without current evidence on how to optimize deflection. Aim was to evaluate in vitro deflection angles with 2 different FANS techniques-sheath advancement and ureteroscope deflection; and effects of sheath size-ureteroscope combinations.
Methods: We evaluated in vitro deflection angles of 10/12Fr, 11/13Fr and 12/14Fr FANS (Hunan Reborn Medical Co. Ltd) with six single-use flexible ureteroscopes (Pusen Uscope 7.5Fr, OTU WiScope 7.5Fr, OTU WiScope 8.6Fr, Innovex EU-scope 8.7Fr, Red Pine RP-U-C12 8.7Fr and Boston Scientific Lithovue 9.5Fr). Two deflection techniques were tested: (1) sheath advancement-advancing the sheath forward over a maximally deflected ureteroscope, and (2) ureteroscope deflection-maximally deflecting the ureteroscope from various starting positions relative to tip of the sheath.
Results: Intra and inter-scope comparisons of maximum deflection angles were significantly different (all ANOVA p < 0.01). Largest maximum angles for all ureteroscopes were with the sheath advancement technique (range 218°-277°), and second largest for most scopes using the ureteroscope deflection technique at tip (range 111°-212°), mostly deviating from manufacturer specifications (range 270°-275°). 10/12Fr and 11/13Fr sheath sizes were more flexible than 12/14Fr. Largest angles were with 11/13Fr sheath-OTU8.6Fr/Innovex8.7Fr combinations.
Conclusion: Optimal deflection with FANS is achieved using either a sheath advancement technique, or ureteroscope deflection technique at tip. Despite using these optimized techniques, deflection angles specified by manufacturers seem hardly achievable. The sheath advancement technique and 11/13Fr sheath-OTU8.6Fr/Innovex8.7Fr combinations may be better suited for lower pole situations. Urologists should be aware of these differences and apply the findings to their FANS technique.
期刊介绍:
The WORLD JOURNAL OF UROLOGY conveys regularly the essential results of urological research and their practical and clinical relevance to a broad audience of urologists in research and clinical practice. In order to guarantee a balanced program, articles are published to reflect the developments in all fields of urology on an internationally advanced level. Each issue treats a main topic in review articles of invited international experts. Free papers are unrelated articles to the main topic.