{"title":"Repeatability, Reproducibility, and Observer Variability of Cortical T1 Mapping for Renal Tissue Characterization.","authors":"Magdalena Nowak, Markus Henningsson, Tom Davis, Najib Chowdhury, Andrea Dennis, Carolina Fernandes, Helena Thomaides Brears, Matthew D Robson","doi":"10.1002/jmri.29602","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The global rise in kidney diseases underscores the need for reliable, noninvasive imaging biomarkers. Among these, renal cortical T1 has shown promise but further technical validation is still required.</p><p><strong>Purpose: </strong>To evaluate the repeatability, reproducibility, and observer variability of kidney cortical T1 mapping in human volunteers without known renal disease.</p><p><strong>Study type: </strong>Prospective.</p><p><strong>Subjects: </strong>Three cohorts without renal disease: 1) 25 volunteers (median age 38 [interquartile range, IQR: 28-42] years, female N = 11) for scan-rescan assessments on GE 1.5 T and Siemens 1.5 T; 2) 29 volunteers (median age 29 [IQR: 24-40] years, female N = 15) for scan-rescan assessments on Siemens 3 T; and 3) 16 volunteers (median age 34 [IQR: 31-42] years, female N = 8) for cross-scanner reproducibility.</p><p><strong>Field strength/sequences: </strong>1.5 T and 3 T, a modified Look-Locker imaging (MOLLI) sequence with a balanced steady-state free precession (bSSFP) readout.</p><p><strong>Assessment: </strong>Kidney cortical T1 data was acquired on GE 1.5 T scanner, Siemens 1.5 T and 3 T scanners. Within-scanner repeatability and inter/intra-observer variability: GE 1.5 T and Siemens 1.5 T, and cross-scanner manufacturer reproducibility: Siemens 1.5 T-GE 1.5 T.</p><p><strong>Statistical tests: </strong>Bland Altman analysis, coefficient of variation (CoV), intra-class coefficient (ICC), and repeatability coefficient (RC).</p><p><strong>Results: </strong>Renal cortical T1 mapping showed high repeatability and reliability across scanner field strengths and manufacturers (repeatability: CoV 1.9%-2.8%, ICC 0.79-0.88, pooled RC 73 msec; reproducibility: CoV 3.0%, ICC 0.75, RC 90 msec). The method also showed robust observer variability (CoV 0.6%-1.4%, ICC 0.93-0.98, RC 22-48 msec).</p><p><strong>Data conclusion: </strong>Kidney cortical T1 mapping is a highly repeatable and reproducible method across MRI manufacturers, field strengths, and observer conditions.</p><p><strong>Evidence level: </strong>2 TECHNICAL EFFICACY: Stage 2.</p>","PeriodicalId":3,"journal":{"name":"ACS Applied Electronic Materials","volume":null,"pages":null},"PeriodicalIF":4.3000,"publicationDate":"2024-10-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Electronic Materials","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/jmri.29602","RegionNum":3,"RegionCategory":"材料科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ENGINEERING, ELECTRICAL & ELECTRONIC","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The global rise in kidney diseases underscores the need for reliable, noninvasive imaging biomarkers. Among these, renal cortical T1 has shown promise but further technical validation is still required.
Purpose: To evaluate the repeatability, reproducibility, and observer variability of kidney cortical T1 mapping in human volunteers without known renal disease.
Study type: Prospective.
Subjects: Three cohorts without renal disease: 1) 25 volunteers (median age 38 [interquartile range, IQR: 28-42] years, female N = 11) for scan-rescan assessments on GE 1.5 T and Siemens 1.5 T; 2) 29 volunteers (median age 29 [IQR: 24-40] years, female N = 15) for scan-rescan assessments on Siemens 3 T; and 3) 16 volunteers (median age 34 [IQR: 31-42] years, female N = 8) for cross-scanner reproducibility.
Field strength/sequences: 1.5 T and 3 T, a modified Look-Locker imaging (MOLLI) sequence with a balanced steady-state free precession (bSSFP) readout.
Assessment: Kidney cortical T1 data was acquired on GE 1.5 T scanner, Siemens 1.5 T and 3 T scanners. Within-scanner repeatability and inter/intra-observer variability: GE 1.5 T and Siemens 1.5 T, and cross-scanner manufacturer reproducibility: Siemens 1.5 T-GE 1.5 T.
Statistical tests: Bland Altman analysis, coefficient of variation (CoV), intra-class coefficient (ICC), and repeatability coefficient (RC).
Results: Renal cortical T1 mapping showed high repeatability and reliability across scanner field strengths and manufacturers (repeatability: CoV 1.9%-2.8%, ICC 0.79-0.88, pooled RC 73 msec; reproducibility: CoV 3.0%, ICC 0.75, RC 90 msec). The method also showed robust observer variability (CoV 0.6%-1.4%, ICC 0.93-0.98, RC 22-48 msec).
Data conclusion: Kidney cortical T1 mapping is a highly repeatable and reproducible method across MRI manufacturers, field strengths, and observer conditions.