Prediction of Thrombus Formation within an Oxygenator via Bioimpedance Analysis.

IF 4.9 3区 工程技术 Q1 CHEMISTRY, ANALYTICAL
Jan Korte, Tobias Lauwigi, Lisa Herzog, Alexander Theißen, Kai Suchorski, Lasse J Strudthoff, Jannis Focke, Sebastian V Jansen, Thomas Gries, Rolf Rossaint, Christian Bleilevens, Patrick Winnersbach
{"title":"Prediction of Thrombus Formation within an Oxygenator via Bioimpedance Analysis.","authors":"Jan Korte, Tobias Lauwigi, Lisa Herzog, Alexander Theißen, Kai Suchorski, Lasse J Strudthoff, Jannis Focke, Sebastian V Jansen, Thomas Gries, Rolf Rossaint, Christian Bleilevens, Patrick Winnersbach","doi":"10.3390/bios14100511","DOIUrl":null,"url":null,"abstract":"<p><p>Blood clot formation inside the membrane oxygenator (MO) remains a risk in extracorporeal membrane oxygenation (ECMO). It is associated with thromboembolic complications and normally detectable only at an advanced stage. Established clinical monitoring techniques lack predictive capabilities, emphasizing the need for refinement in MO monitoring towards an early warning system. In this study, an MO was modified by integrating four sensor fibers in the middle of the hollow fiber mat bundle, allowing for bioimpedance measurement within the MO. The modified MO was perfused with human blood in an <i>in vitro</i> test circuit until fulminant clot formation. The optical analysis of clot residues on the extracted hollow fibers showed a clot deposition area of 51.88% ± 14.25%. This was detectable via an increased bioimpedance signal with a significant increase 5 min in advance to fulminant clot formation inside the MO, which was monitored by the clinical gold standard (pressure difference across the MO (dp-MO)). This study demonstrates the feasibility of detecting clot growth early and effectively by measuring bioimpedance within an MO using integrated sensor fibers. Thus, bioimpedance may even outperform the clinical gold standard of dp-MO as a monitoring method by providing earlier clot detection.</p>","PeriodicalId":48608,"journal":{"name":"Biosensors-Basel","volume":"14 10","pages":""},"PeriodicalIF":4.9000,"publicationDate":"2024-10-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11506726/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biosensors-Basel","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.3390/bios14100511","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Blood clot formation inside the membrane oxygenator (MO) remains a risk in extracorporeal membrane oxygenation (ECMO). It is associated with thromboembolic complications and normally detectable only at an advanced stage. Established clinical monitoring techniques lack predictive capabilities, emphasizing the need for refinement in MO monitoring towards an early warning system. In this study, an MO was modified by integrating four sensor fibers in the middle of the hollow fiber mat bundle, allowing for bioimpedance measurement within the MO. The modified MO was perfused with human blood in an in vitro test circuit until fulminant clot formation. The optical analysis of clot residues on the extracted hollow fibers showed a clot deposition area of 51.88% ± 14.25%. This was detectable via an increased bioimpedance signal with a significant increase 5 min in advance to fulminant clot formation inside the MO, which was monitored by the clinical gold standard (pressure difference across the MO (dp-MO)). This study demonstrates the feasibility of detecting clot growth early and effectively by measuring bioimpedance within an MO using integrated sensor fibers. Thus, bioimpedance may even outperform the clinical gold standard of dp-MO as a monitoring method by providing earlier clot detection.

通过生物阻抗分析预测氧合器内血栓的形成。
在体外膜氧合(ECMO)中,膜氧合器(MO)内血凝块的形成仍然是一个风险。它与血栓栓塞并发症有关,通常只有在晚期才能检测到。现有的临床监测技术缺乏预测能力,因此需要对 MO 监测进行改进,以建立早期预警系统。在这项研究中,对 MO 进行了改进,在中空纤维毡束中间集成了四根传感器纤维,从而可以在 MO 内进行生物阻抗测量。在体外测试电路中,用人血灌注改进后的 MO,直至形成恶性血凝块。对提取的中空纤维上残留血块的光学分析表明,血块沉积面积为 51.88% ± 14.25%。这可通过生物阻抗信号的增加检测到,在 MO 内血栓形成前 5 分钟,生物阻抗信号显著增加,临床金标准(MO 两端的压差(dp-MO))可监测到这一情况。这项研究证明了通过使用集成传感器纤维测量 MO 内的生物阻抗来及早、有效地检测血栓生长的可行性。因此,作为一种监测方法,生物阻抗甚至可能优于临床金标准(dp-MO),能更早地检测到血栓。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Biosensors-Basel
Biosensors-Basel Biochemistry, Genetics and Molecular Biology-Clinical Biochemistry
CiteScore
6.60
自引率
14.80%
发文量
983
审稿时长
11 weeks
期刊介绍: Biosensors (ISSN 2079-6374) provides an advanced forum for studies related to the science and technology of biosensors and biosensing. It publishes original research papers, comprehensive reviews and communications. Our aim is to encourage scientists to publish their experimental and theoretical results in as much detail as possible. There is no restriction on the length of the papers. The full experimental details must be provided so that the results can be reproduced. Electronic files and software regarding the full details of the calculation or experimental procedure, if unable to be published in a normal way, can be deposited as supplementary electronic material.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信