Influence of Continuous Rotation and Optimal Torque Reverse Kinematics on the Cyclic Fatigue Strength of Endodontic NiTi Clockwise Cutting Rotary Instruments.
Jorge N R Martins, Emmanuel J N L Silva, Duarte Marques, Marco A Versiani
{"title":"Influence of Continuous Rotation and Optimal Torque Reverse Kinematics on the Cyclic Fatigue Strength of Endodontic NiTi Clockwise Cutting Rotary Instruments.","authors":"Jorge N R Martins, Emmanuel J N L Silva, Duarte Marques, Marco A Versiani","doi":"10.3390/dj12100317","DOIUrl":null,"url":null,"abstract":"<p><p><b>Objectives:</b> The objective of the present study was to evaluate the cyclic fatigue strength of clockwise cutting rotary endodontic instruments when subjected to two different kinematics: continuous clockwise rotation and clockwise reciprocation movement under optimum torque reverse (OTR) motion. <b>Methods:</b> New ProTaper Next X1 (n = 20) and X2 (n = 20) instruments were randomly divided into two subgroups (n = 10) based on kinematics (continuous rotation or OTR). The specimens were tested using a custom-made device with a non-tapered stainless-steel artificial canal measuring 19 mm in length, featuring a 6 mm radius and an 86-degree curvature. All instruments were tested with a lubricant at room temperature until a fracture occurred. The time to fracture and the length of the separated fragment were recorded. Subsequently, the fractured instruments were inspected under a scanning electron microscope for signs of cyclic fatigue failure, plastic deformation, and/or crack propagation. The subgroup comparisons for time to fracture and instrument length were performed using the independent samples <i>t</i>-test, with the level of statistical significance set at 0.05. <b>Results:</b> When using OTR movement, the ProTaper Next X1 increased the time to fracture from 52.9 to 125.8 s (<i>p</i> < 0.001), while the ProTaper Next X2 increased from 45.4 to 66.0 s (<i>p</i> < 0.001). No subgroup exhibited plastic deformations, but both showed dimpling marks indicative of cyclic fatigue as the primary mode of failure. Additionally, OTR movement resulted in more metal alloy microcracks. <b>Conclusions:</b> The use of OTR motion extended the lifespan of the tested instruments and resulted in a higher number of metal microcracks. This suggests that OTR motion helped to distribute the mechanical stress more evenly across the instrument, thereby relieving localized tension. As a result, it delayed the formation of a single catastrophic crack, enhancing the overall performance of the instruments during the experimental procedures.</p>","PeriodicalId":11269,"journal":{"name":"Dentistry Journal","volume":"12 10","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2024-09-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11506770/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Dentistry Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3390/dj12100317","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: The objective of the present study was to evaluate the cyclic fatigue strength of clockwise cutting rotary endodontic instruments when subjected to two different kinematics: continuous clockwise rotation and clockwise reciprocation movement under optimum torque reverse (OTR) motion. Methods: New ProTaper Next X1 (n = 20) and X2 (n = 20) instruments were randomly divided into two subgroups (n = 10) based on kinematics (continuous rotation or OTR). The specimens were tested using a custom-made device with a non-tapered stainless-steel artificial canal measuring 19 mm in length, featuring a 6 mm radius and an 86-degree curvature. All instruments were tested with a lubricant at room temperature until a fracture occurred. The time to fracture and the length of the separated fragment were recorded. Subsequently, the fractured instruments were inspected under a scanning electron microscope for signs of cyclic fatigue failure, plastic deformation, and/or crack propagation. The subgroup comparisons for time to fracture and instrument length were performed using the independent samples t-test, with the level of statistical significance set at 0.05. Results: When using OTR movement, the ProTaper Next X1 increased the time to fracture from 52.9 to 125.8 s (p < 0.001), while the ProTaper Next X2 increased from 45.4 to 66.0 s (p < 0.001). No subgroup exhibited plastic deformations, but both showed dimpling marks indicative of cyclic fatigue as the primary mode of failure. Additionally, OTR movement resulted in more metal alloy microcracks. Conclusions: The use of OTR motion extended the lifespan of the tested instruments and resulted in a higher number of metal microcracks. This suggests that OTR motion helped to distribute the mechanical stress more evenly across the instrument, thereby relieving localized tension. As a result, it delayed the formation of a single catastrophic crack, enhancing the overall performance of the instruments during the experimental procedures.