Andre Jun Hui Wong, Ryan Ruiyang Ling, Chong Boon Teo, Jeremy Chee, Raymond Yeow Seng Ngo, Woei Shyang Loh, Eunice Dawn Kwa
{"title":"Evaluating the accuracy of a self-administered smartphone hearing test application in a geriatric population.","authors":"Andre Jun Hui Wong, Ryan Ruiyang Ling, Chong Boon Teo, Jeremy Chee, Raymond Yeow Seng Ngo, Woei Shyang Loh, Eunice Dawn Kwa","doi":"10.1007/s00405-024-08989-z","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>As the global population ages, hearing loss becomes increasingly prevalent, and is associated with neurocognitive and psychiatric comorbidities, impacting quality of life. Early screening and timely intervention might prevent or delay cognitive decline, a gap in care that can potentially be addressed by self-administered smartphone hearing tests.</p><p><strong>Objective: </strong>This study aims to evaluate the accuracy of Mimi™ (Berlin, Germany), a commercially available self-administered smartphone hearing test compared to pure tone audiogram (PTA) in terms of both hearing levels and hearing thresholds in our local geriatric population > 65 years-old.</p><p><strong>Method: </strong>Fifty-two participants above 65 years of age requiring conventional audiograms were recruited from a National Referral University Hospital Otolaryngology clinic from March to June 2022. All participants were administered the conventional PTA tests in a sound-proof booth conducted by audiology technicians followed by Mimi™ Hearing Test in a quiet clinic room. Comparisons between the hearing levels of both tests were analyzed using Spearman's rank correlation coefficient, Bland-Altman plots and Gwet's Kappa which looked at concordance. Hearing thresholds were then analysed using the Wilcoxon signed rank (SR) test.</p><p><strong>Results: </strong>Mimi™ showed strong to very strong correlation with good agreement compared to readings obtained from formal PTA. Concordance in determining hearing loss also showed substantial to almost perfect agreement at each individual frequency, with values of kappa falling between 0.735-0.857. In terms of thresholds, there were no significant differences in thresholds given by both tests except for 2.0 kHz, HFPTA and 4FPTA (p < 0.05).</p><p><strong>Conclusion: </strong>Mimi™ serves as a good screening tool for detection of moderate hearing loss for early pickup and treatment except at higher frequencies. The smartphone hearing test is also less accurate in determining the extent of hearing loss and formal PTA after hearing loss is detected on screening should still be standard of care.</p>","PeriodicalId":11952,"journal":{"name":"European Archives of Oto-Rhino-Laryngology","volume":null,"pages":null},"PeriodicalIF":1.9000,"publicationDate":"2024-10-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Archives of Oto-Rhino-Laryngology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00405-024-08989-z","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OTORHINOLARYNGOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: As the global population ages, hearing loss becomes increasingly prevalent, and is associated with neurocognitive and psychiatric comorbidities, impacting quality of life. Early screening and timely intervention might prevent or delay cognitive decline, a gap in care that can potentially be addressed by self-administered smartphone hearing tests.
Objective: This study aims to evaluate the accuracy of Mimi™ (Berlin, Germany), a commercially available self-administered smartphone hearing test compared to pure tone audiogram (PTA) in terms of both hearing levels and hearing thresholds in our local geriatric population > 65 years-old.
Method: Fifty-two participants above 65 years of age requiring conventional audiograms were recruited from a National Referral University Hospital Otolaryngology clinic from March to June 2022. All participants were administered the conventional PTA tests in a sound-proof booth conducted by audiology technicians followed by Mimi™ Hearing Test in a quiet clinic room. Comparisons between the hearing levels of both tests were analyzed using Spearman's rank correlation coefficient, Bland-Altman plots and Gwet's Kappa which looked at concordance. Hearing thresholds were then analysed using the Wilcoxon signed rank (SR) test.
Results: Mimi™ showed strong to very strong correlation with good agreement compared to readings obtained from formal PTA. Concordance in determining hearing loss also showed substantial to almost perfect agreement at each individual frequency, with values of kappa falling between 0.735-0.857. In terms of thresholds, there were no significant differences in thresholds given by both tests except for 2.0 kHz, HFPTA and 4FPTA (p < 0.05).
Conclusion: Mimi™ serves as a good screening tool for detection of moderate hearing loss for early pickup and treatment except at higher frequencies. The smartphone hearing test is also less accurate in determining the extent of hearing loss and formal PTA after hearing loss is detected on screening should still be standard of care.
期刊介绍:
Official Journal of
European Union of Medical Specialists – ORL Section and Board
Official Journal of Confederation of European Oto-Rhino-Laryngology Head and Neck Surgery
"European Archives of Oto-Rhino-Laryngology" publishes original clinical reports and clinically relevant experimental studies, as well as short communications presenting new results of special interest. With peer review by a respected international editorial board and prompt English-language publication, the journal provides rapid dissemination of information by authors from around the world. This particular feature makes it the journal of choice for readers who want to be informed about the continuing state of the art concerning basic sciences and the diagnosis and management of diseases of the head and neck on an international level.
European Archives of Oto-Rhino-Laryngology was founded in 1864 as "Archiv für Ohrenheilkunde" by A. von Tröltsch, A. Politzer and H. Schwartze.