Alexander Sverstad, Olav Kristianslund, Goran Petrovski, Morten Carstens Moe, Øystein Kalsnes Jørstad
{"title":"A proof-of-concept study of a prototype needle that mitigates intraocular pressure rise following intravitreal injection.","authors":"Alexander Sverstad, Olav Kristianslund, Goran Petrovski, Morten Carstens Moe, Øystein Kalsnes Jørstad","doi":"10.1186/s40942-024-00590-0","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To determine the feasibility of a prototype needle that enhances vitreous reflux (VR) to control intraocular pressure (IOP) in intravitreal injection (IVI).</p><p><strong>Methods: </strong>We created an eye model to compare IVI using a standard 30-G needle with four different versions of a 30-G prototype needle with one to four surface grooves that enhanced VR. We injected 50, 70, and 100 µl saline through porcine sclera or 460-µm-thick rubber and measured the peak and 3-second pressure before we extracted the needle and measured the 10-second pressure.</p><p><strong>Results: </strong>50-µl injection through sclera with the standard needle resulted in mean (SD) pressure of 58.6 (3.8) mmHg at peak, 52.8 (4.7) mmHg at 3 s, and 39.6 (18.0) mmHg at 10 s. The prototype needle lowered the pressure; four grooves resulted in mean (SD) pressure of 29.4 (5.6) mmHg at peak, 22.0 (3.7) mmHg at 3 s, and 7.2 (6.6) mmHg at 10 s. 70-µl and 100-µl injections through sclera with the standard needle resulted in mean (SD) pressure of 68.8 (3.6) and 86.0 (6.0) mmHg at peak. Similar to 50-µl injection, the prototype needle lowered the pressure for 70-µl and 100-µl injections. At 10 s, we observed varying leakage at the injection site for sclera but not for rubber.</p><p><strong>Conclusions: </strong>The study provides proof of concept for a needle design for which surface grooves enhance VR and counteract the effect of IVI on IOP. The safety and efficacy of the prototype needle must be studied further in a clinical trial.</p>","PeriodicalId":14289,"journal":{"name":"International Journal of Retina and Vitreous","volume":null,"pages":null},"PeriodicalIF":1.9000,"publicationDate":"2024-10-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11466030/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Retina and Vitreous","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s40942-024-00590-0","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To determine the feasibility of a prototype needle that enhances vitreous reflux (VR) to control intraocular pressure (IOP) in intravitreal injection (IVI).
Methods: We created an eye model to compare IVI using a standard 30-G needle with four different versions of a 30-G prototype needle with one to four surface grooves that enhanced VR. We injected 50, 70, and 100 µl saline through porcine sclera or 460-µm-thick rubber and measured the peak and 3-second pressure before we extracted the needle and measured the 10-second pressure.
Results: 50-µl injection through sclera with the standard needle resulted in mean (SD) pressure of 58.6 (3.8) mmHg at peak, 52.8 (4.7) mmHg at 3 s, and 39.6 (18.0) mmHg at 10 s. The prototype needle lowered the pressure; four grooves resulted in mean (SD) pressure of 29.4 (5.6) mmHg at peak, 22.0 (3.7) mmHg at 3 s, and 7.2 (6.6) mmHg at 10 s. 70-µl and 100-µl injections through sclera with the standard needle resulted in mean (SD) pressure of 68.8 (3.6) and 86.0 (6.0) mmHg at peak. Similar to 50-µl injection, the prototype needle lowered the pressure for 70-µl and 100-µl injections. At 10 s, we observed varying leakage at the injection site for sclera but not for rubber.
Conclusions: The study provides proof of concept for a needle design for which surface grooves enhance VR and counteract the effect of IVI on IOP. The safety and efficacy of the prototype needle must be studied further in a clinical trial.
期刊介绍:
International Journal of Retina and Vitreous focuses on the ophthalmic subspecialty of vitreoretinal disorders. The journal presents original articles on new approaches to diagnosis, outcomes of clinical trials, innovations in pharmacological therapy and surgical techniques, as well as basic science advances that impact clinical practice. Topical areas include, but are not limited to: -Imaging of the retina, choroid and vitreous -Innovations in optical coherence tomography (OCT) -Small-gauge vitrectomy, retinal detachment, chromovitrectomy -Electroretinography (ERG), microperimetry, other functional tests -Intraocular tumors -Retinal pharmacotherapy & drug delivery -Diabetic retinopathy & other vascular diseases -Age-related macular degeneration (AMD) & other macular entities