Development of a chemiluminescent immunoassay based on magnetic solid phase for quantification of homocysteine in human serum.

IF 3.5 3区 生物学 Q2 BIOTECHNOLOGY & APPLIED MICROBIOLOGY
Yankui Liu, Yaru Li, Rong Wang, Shuning Chen, Ning Sun, Xiaowei Qi
{"title":"Development of a chemiluminescent immunoassay based on magnetic solid phase for quantification of homocysteine in human serum.","authors":"Yankui Liu, Yaru Li, Rong Wang, Shuning Chen, Ning Sun, Xiaowei Qi","doi":"10.1186/s12896-024-00903-z","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Homocysteine (HCY) is a sulfur-containing amino acid that is an independent or important risk factor for the occurrence of many chronic diseases and is one of the most important indicators for determining health risks. However, existing HCY detection methods do not meet the requirements of clinical diagnosis. Therefore, there is an urgent need to establish new detection methods to meet the needs of clinical detection.</p><p><strong>Results: </strong>In this study, we used the principle of competitive method to establish a new method for the determination of HCY in human serum using a chemiluminescent enzyme immunoassay in conjunction with a chemiluminescent assay instrument that uses magnetic microparticles as the solid phase of the immunoreaction. The established method achieved satisfactory results in terms of minimum detection limit, specificity, accuracy, and clinical application. The limit of detection was 0.03 ng/mL. The intra-assay coefficient of variation (CV) was 1.94-5.05%, the inter-assay CV was 2.29-6.88%, and the recovery rate was 88.60-93.27%. Cross-reactivity with L-cysteine ranged from 0.0100 to 0.0200 μmol/L, and cross-reactivity with glutathione ranged from 0.0100 to 0.200 μmol/L, all of which were less than the limit of detection (LoD) of this method. The linear factor R of this method was greater than 0.99.</p><p><strong>Conclusions: </strong>In summary, the developed method showed a good correlation with the product from Abbott. A total of 996 clinical patients with cardiovascular diseases were evaluated using the method developed in this study.</p>","PeriodicalId":8905,"journal":{"name":"BMC Biotechnology","volume":"24 1","pages":"77"},"PeriodicalIF":3.5000,"publicationDate":"2024-10-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11465875/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"BMC Biotechnology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1186/s12896-024-00903-z","RegionNum":3,"RegionCategory":"生物学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"BIOTECHNOLOGY & APPLIED MICROBIOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Homocysteine (HCY) is a sulfur-containing amino acid that is an independent or important risk factor for the occurrence of many chronic diseases and is one of the most important indicators for determining health risks. However, existing HCY detection methods do not meet the requirements of clinical diagnosis. Therefore, there is an urgent need to establish new detection methods to meet the needs of clinical detection.

Results: In this study, we used the principle of competitive method to establish a new method for the determination of HCY in human serum using a chemiluminescent enzyme immunoassay in conjunction with a chemiluminescent assay instrument that uses magnetic microparticles as the solid phase of the immunoreaction. The established method achieved satisfactory results in terms of minimum detection limit, specificity, accuracy, and clinical application. The limit of detection was 0.03 ng/mL. The intra-assay coefficient of variation (CV) was 1.94-5.05%, the inter-assay CV was 2.29-6.88%, and the recovery rate was 88.60-93.27%. Cross-reactivity with L-cysteine ranged from 0.0100 to 0.0200 μmol/L, and cross-reactivity with glutathione ranged from 0.0100 to 0.200 μmol/L, all of which were less than the limit of detection (LoD) of this method. The linear factor R of this method was greater than 0.99.

Conclusions: In summary, the developed method showed a good correlation with the product from Abbott. A total of 996 clinical patients with cardiovascular diseases were evaluated using the method developed in this study.

开发一种基于磁性固相的化学发光免疫测定法,用于定量检测人体血清中的同型半胱氨酸。
背景:同型半胱氨酸(HCY)是一种含硫氨基酸,是许多慢性疾病发生的独立或重要危险因素,也是判断健康风险的重要指标之一。然而,现有的 HCY 检测方法无法满足临床诊断的要求。因此,迫切需要建立新的检测方法来满足临床检测的需要:本研究利用竞争法原理,结合以磁性微粒为免疫反应固相的化学发光检测仪器,建立了一种化学发光酶免疫法测定人血清中 HCY 的新方法。所建立的方法在最低检测限、特异性、准确性和临床应用方面都取得了令人满意的结果。检测限为 0.03 纳克/毫升。测定内变异系数为 1.94-5.05%,测定间变异系数为 2.29-6.88%,回收率为 88.60-93.27%。与L-半胱氨酸的交叉反应为0.0100~0.0200 μmol/L,与谷胱甘肽的交叉反应为0.0100~0.200 μmol/L,均小于方法的检出限(LoD)。该方法的线性系数R大于0.99:总之,所开发的方法与雅培公司的产品具有良好的相关性。本研究使用所开发的方法对996名心血管疾病临床患者进行了评估。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
BMC Biotechnology
BMC Biotechnology 工程技术-生物工程与应用微生物
CiteScore
6.60
自引率
0.00%
发文量
34
审稿时长
2 months
期刊介绍: BMC Biotechnology is an open access, peer-reviewed journal that considers articles on the manipulation of biological macromolecules or organisms for use in experimental procedures, cellular and tissue engineering or in the pharmaceutical, agricultural biotechnology and allied industries.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信