Sagar Sabharwal, Geoff Heyes, George S J Tudor, Robert Flintham, Swarupsinh Chavda, Paul Sanghera
{"title":"Calculating Optic Nerve Planning Organ at Risk Volume Margins for Stereotactic Radiosurgery Using Optic Nerve Motion determined using MRI.","authors":"Sagar Sabharwal, Geoff Heyes, George S J Tudor, Robert Flintham, Swarupsinh Chavda, Paul Sanghera","doi":"10.1093/bjr/tqae201","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>The combination of sharp dose gradients in stereotactic radiosurgery (SRS) and minute optic nerve motion may significantly increase dose to the optic nerves when treating perioptic lesions. The aim of this study was to calculate optic nerve planning organ at risk volume (PRV) margins for CyberKnife SRS treatment planning.</p><p><strong>Methods: </strong>MRI scans were taken of 10 healthy volunteers looking left, right, up, down, and straight ahead to measure optic nerve motion. The measured optic nerve motion and the uncertainties in the technical accuracy of CyberKnife were used to calculate optic nerve PRV margins.</p><p><strong>Results: </strong>Two optic nerve PRV margins were calculated: a non-isotropic margin of mL/R,PRV=3mm, mSup/Inf,PRV=2mm and mAnt/Post,PRV=1mm which considers the full range of motion measured in a worst case scenario; and an isotropic margin of mPRV=1mm which considers a scenario where patients are asked to look neutrally during imaging and treatment. Applying these PRVs to 8 historical sphenoid wing meningioma CyberKnife plans showed tolerance levels may be exceeded due to optic nerve motion.</p><p><strong>Conclusions: </strong>Optic nerve PRV margins may be needed in CyberKnife planning to reduce risk to the optic nerves. The use of a mPRV=1mm PRV to account for organ motion, along with instructing patients to hold their gaze neutrally during imaging and treatment, may be a suitable organ sparing strategy.</p><p><strong>Advances in knowledge: </strong>Measured optic nerve motion and the technical accuracy of the CyberKnife system have been used to calculate optic nerve PRV margins.</p>","PeriodicalId":9306,"journal":{"name":"British Journal of Radiology","volume":" ","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2024-10-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"British Journal of Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/bjr/tqae201","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: The combination of sharp dose gradients in stereotactic radiosurgery (SRS) and minute optic nerve motion may significantly increase dose to the optic nerves when treating perioptic lesions. The aim of this study was to calculate optic nerve planning organ at risk volume (PRV) margins for CyberKnife SRS treatment planning.
Methods: MRI scans were taken of 10 healthy volunteers looking left, right, up, down, and straight ahead to measure optic nerve motion. The measured optic nerve motion and the uncertainties in the technical accuracy of CyberKnife were used to calculate optic nerve PRV margins.
Results: Two optic nerve PRV margins were calculated: a non-isotropic margin of mL/R,PRV=3mm, mSup/Inf,PRV=2mm and mAnt/Post,PRV=1mm which considers the full range of motion measured in a worst case scenario; and an isotropic margin of mPRV=1mm which considers a scenario where patients are asked to look neutrally during imaging and treatment. Applying these PRVs to 8 historical sphenoid wing meningioma CyberKnife plans showed tolerance levels may be exceeded due to optic nerve motion.
Conclusions: Optic nerve PRV margins may be needed in CyberKnife planning to reduce risk to the optic nerves. The use of a mPRV=1mm PRV to account for organ motion, along with instructing patients to hold their gaze neutrally during imaging and treatment, may be a suitable organ sparing strategy.
Advances in knowledge: Measured optic nerve motion and the technical accuracy of the CyberKnife system have been used to calculate optic nerve PRV margins.
期刊介绍:
BJR is the international research journal of the British Institute of Radiology and is the oldest scientific journal in the field of radiology and related sciences.
Dating back to 1896, BJR’s history is radiology’s history, and the journal has featured some landmark papers such as the first description of Computed Tomography "Computerized transverse axial tomography" by Godfrey Hounsfield in 1973. A valuable historical resource, the complete BJR archive has been digitized from 1896.
Quick Facts:
- 2015 Impact Factor – 1.840
- Receipt to first decision – average of 6 weeks
- Acceptance to online publication – average of 3 weeks
- ISSN: 0007-1285
- eISSN: 1748-880X
Open Access option