{"title":"Improved metal suppression using new generation low-field MRI: a biophantom feasibility study.","authors":"Johanna Luitjens, Katharina Ziegeler, Daehyun Yoon, Felix Gassert, Rupsa Bhattacharjee, Rawee Manatrakul, Chotigar Ngarmsrikam, Amy Becker, Yang Yang, Gabby B Joseph, Pan Su, Pedro Itriago-Leon, Sharmila Majumdar, Thomas M Link","doi":"10.1007/s00256-024-04809-x","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Novel 0.55 MRI scanners have the potential to reduce metal artifacts around orthopedic implants. The purpose of this study was to compare metal artifact size and depiction of anatomy between 0.55 T and 3.0 T MRI in a biophantom.</p><p><strong>Materials and methods: </strong>Steel and titanium screws were implanted in 12 porcine knee specimens and imaging at 0.55 T and 3 T MRI was performed using the following sequences: turbo spin-echo (TSE), TSE with view angle tilting (VAT), and slice encoding for metal artifact correction (SEMAC) with proton-density (PD) and T2-weighted short-tau inversion-recovery (T2w-STIR) contrasts. Artifacts were measured, and visualization of anatomy (cartilage, bone, growth plates, cruciate ligaments) was assessed and compared between groups.</p><p><strong>Results: </strong>Metal artifacts were significantly smaller at 0.55 T. The smallest artifact sizes were achieved with SEMAC at 0.55 T for both PD and T2w-STIR sequences; corresponding relative size reductions vs. 3.0 T were 78.7% and 79.4% (stainless steel) and 45.3% and 1.4% (titanium). Depiction of anatomical structures was superior at 0.55 T.</p><p><strong>Conclusion: </strong>Substantial reduction of artifact size resulting in superior depiction of anatomical structures is possible on novel 0.55 T MRI systems. Further clinical studies are required to elucidate patient-relevant advantages.</p>","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":null,"pages":null},"PeriodicalIF":16.4000,"publicationDate":"2024-10-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00256-024-04809-x","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: Novel 0.55 MRI scanners have the potential to reduce metal artifacts around orthopedic implants. The purpose of this study was to compare metal artifact size and depiction of anatomy between 0.55 T and 3.0 T MRI in a biophantom.
Materials and methods: Steel and titanium screws were implanted in 12 porcine knee specimens and imaging at 0.55 T and 3 T MRI was performed using the following sequences: turbo spin-echo (TSE), TSE with view angle tilting (VAT), and slice encoding for metal artifact correction (SEMAC) with proton-density (PD) and T2-weighted short-tau inversion-recovery (T2w-STIR) contrasts. Artifacts were measured, and visualization of anatomy (cartilage, bone, growth plates, cruciate ligaments) was assessed and compared between groups.
Results: Metal artifacts were significantly smaller at 0.55 T. The smallest artifact sizes were achieved with SEMAC at 0.55 T for both PD and T2w-STIR sequences; corresponding relative size reductions vs. 3.0 T were 78.7% and 79.4% (stainless steel) and 45.3% and 1.4% (titanium). Depiction of anatomical structures was superior at 0.55 T.
Conclusion: Substantial reduction of artifact size resulting in superior depiction of anatomical structures is possible on novel 0.55 T MRI systems. Further clinical studies are required to elucidate patient-relevant advantages.
期刊介绍:
Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance.
Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.