{"title":"Inhaled corticosteroid delivery is markedly affected by breathing pattern and valved holding chamber model.","authors":"Laura Ojanperä, Lauri Lehtimäki, Heini Huhtala, Péter Csonka","doi":"10.1002/ppul.27300","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>There is a scarcity of high-quality research on the efficient delivery of inhaled corticosteroids using valved holding chambers (VHCs) in children.</p><p><strong>Methods: </strong>The delivered dose (DD) of fluticasone from a metered dose inhaler (pMDI) was tested using four VHCs: AeroChamber plus Flow-Vu (AC), Babyhaler (BH), EasyChamber (EC), and Optichamber Diamond (OD). The in vitro setup included an anatomical child throat model, Next Generation Impactor, and a breathing simulator to generate tidal breathing of a four and a 6-year-old child, and adult type single inhalation.</p><p><strong>Results: </strong>OD showed the lowest proportion of fluticasone trapped in the throat with all breathing patterns. AC showed similar fine particle dose (FPD) in the respirable range (1-5 µm) irrespective of the breathing pattern. For BH, the median FPD 1-5 µm was highest during adult breathing. OD and EC showed higher overall DD and higher doses in the 1-5 µm range with paediatric breathing profiles compared to adult inhalation. The median DD and FPD 1-5 µm were significantly lower with BH compared to any other VHCs during tidal breathing. Compared to EC, the FPD of the other VHCs were skewed towards <2 µm particles.</p><p><strong>Conclusion: </strong>Fluticasone delivery is markedly affected by breathing pattern and VHC model. The observed differences in throat deposition and FPD delivered may have significant clinical implications for side effects and controlling airway inflammation. All VHCs intended for paediatric use should undergo testing using internationally recognised standardised methods incorporating clinically relevant paediatric breathing patterns.</p>","PeriodicalId":19932,"journal":{"name":"Pediatric Pulmonology","volume":" ","pages":"e27300"},"PeriodicalIF":2.7000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11715152/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pediatric Pulmonology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/ppul.27300","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/9/30 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"PEDIATRICS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: There is a scarcity of high-quality research on the efficient delivery of inhaled corticosteroids using valved holding chambers (VHCs) in children.
Methods: The delivered dose (DD) of fluticasone from a metered dose inhaler (pMDI) was tested using four VHCs: AeroChamber plus Flow-Vu (AC), Babyhaler (BH), EasyChamber (EC), and Optichamber Diamond (OD). The in vitro setup included an anatomical child throat model, Next Generation Impactor, and a breathing simulator to generate tidal breathing of a four and a 6-year-old child, and adult type single inhalation.
Results: OD showed the lowest proportion of fluticasone trapped in the throat with all breathing patterns. AC showed similar fine particle dose (FPD) in the respirable range (1-5 µm) irrespective of the breathing pattern. For BH, the median FPD 1-5 µm was highest during adult breathing. OD and EC showed higher overall DD and higher doses in the 1-5 µm range with paediatric breathing profiles compared to adult inhalation. The median DD and FPD 1-5 µm were significantly lower with BH compared to any other VHCs during tidal breathing. Compared to EC, the FPD of the other VHCs were skewed towards <2 µm particles.
Conclusion: Fluticasone delivery is markedly affected by breathing pattern and VHC model. The observed differences in throat deposition and FPD delivered may have significant clinical implications for side effects and controlling airway inflammation. All VHCs intended for paediatric use should undergo testing using internationally recognised standardised methods incorporating clinically relevant paediatric breathing patterns.
期刊介绍:
Pediatric Pulmonology (PPUL) is the foremost global journal studying the respiratory system in disease and in health as it develops from intrauterine life though adolescence to adulthood. Combining explicit and informative analysis of clinical as well as basic scientific research, PPUL provides a look at the many facets of respiratory system disorders in infants and children, ranging from pathological anatomy, developmental issues, and pathophysiology to infectious disease, asthma, cystic fibrosis, and airborne toxins. Focused attention is given to the reporting of diagnostic and therapeutic methods for neonates, preschool children, and adolescents, the enduring effects of childhood respiratory diseases, and newly described infectious diseases.
PPUL concentrates on subject matters of crucial interest to specialists preparing for the Pediatric Subspecialty Examinations in the United States and other countries. With its attentive coverage and extensive clinical data, this journal is a principle source for pediatricians in practice and in training and a must have for all pediatric pulmonologists.