Transcatheter ventilation with a modified Rapid-O2 oxygen insufflation device.

IF 4.2 4区 医学 Q1 ANESTHESIOLOGY
Jaewon Jang, Hye Jin Kim, Hyun Joo Kim, Wyun Kon Park
{"title":"Transcatheter ventilation with a modified Rapid-O2 oxygen insufflation device.","authors":"Jaewon Jang, Hye Jin Kim, Hyun Joo Kim, Wyun Kon Park","doi":"10.4097/kja.24095","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The Rapid-O2 oxygen insufflation device® (Rapid-O2) was designed primarily for rescue oxygenation in cannot intubate, cannot oxygenate (CICO) events; thus, hypercapnia is inevitable. Rapid-O2 was modified to enhance ventilation using the Venturi effect during expiration.</p><p><strong>Methods: </strong>To determine the most effective combination of inner catheters (20 G, 18 G, 16 G, 14 G, and 2-mm inner diameter [ID] transtracheal catheter [TTC]) and insufflation catheters (16 G, 14 G, and 2-mm ID TTC) for achieving optimum ventilation, insufflating and expiratory flows were measured at an oxygen flow rate of 15 L/min. The insufflating and expiratory pressures were measured at 6-15 L/min. The flows and pressures were measured using a gas flow analyzer. The insufflating and expiratory times were measured using a trachea-lung model to obtain minute volumes. To assess the improvement by modifying the Rapid-O2, minute volumes were measured using the Rapid-O2.</p><p><strong>Results: </strong>The most appropriate inner catheter was 18 G. The insufflating pressures ranged from 97 (2-mm ID TTC) to 377 cmH2O (16 G) at 15 L/min. During expiration, similar negative pressures of 50 cmH2O were measured in the insufflation catheters at 15 L/min. At lung compliance of 100 ml/cmH2O, the minute volumes through a 2-mm ID and 14 G insufflation catheters were 7.0 and 5.37 L/min, respectively, at 15 L/min. The minute volumes were significantly greater in modified Rapid-O2.</p><p><strong>Conclusions: </strong>Modified Rapid-O2 provided sufficient minute volumes in adults using a 14 G or 2-mm ID insufflation catheter at 15 L/min, demonstrating its potential for ventilation in CICO events.</p>","PeriodicalId":17855,"journal":{"name":"Korean Journal of Anesthesiology","volume":null,"pages":null},"PeriodicalIF":4.2000,"publicationDate":"2024-09-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Korean Journal of Anesthesiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.4097/kja.24095","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ANESTHESIOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The Rapid-O2 oxygen insufflation device® (Rapid-O2) was designed primarily for rescue oxygenation in cannot intubate, cannot oxygenate (CICO) events; thus, hypercapnia is inevitable. Rapid-O2 was modified to enhance ventilation using the Venturi effect during expiration.

Methods: To determine the most effective combination of inner catheters (20 G, 18 G, 16 G, 14 G, and 2-mm inner diameter [ID] transtracheal catheter [TTC]) and insufflation catheters (16 G, 14 G, and 2-mm ID TTC) for achieving optimum ventilation, insufflating and expiratory flows were measured at an oxygen flow rate of 15 L/min. The insufflating and expiratory pressures were measured at 6-15 L/min. The flows and pressures were measured using a gas flow analyzer. The insufflating and expiratory times were measured using a trachea-lung model to obtain minute volumes. To assess the improvement by modifying the Rapid-O2, minute volumes were measured using the Rapid-O2.

Results: The most appropriate inner catheter was 18 G. The insufflating pressures ranged from 97 (2-mm ID TTC) to 377 cmH2O (16 G) at 15 L/min. During expiration, similar negative pressures of 50 cmH2O were measured in the insufflation catheters at 15 L/min. At lung compliance of 100 ml/cmH2O, the minute volumes through a 2-mm ID and 14 G insufflation catheters were 7.0 and 5.37 L/min, respectively, at 15 L/min. The minute volumes were significantly greater in modified Rapid-O2.

Conclusions: Modified Rapid-O2 provided sufficient minute volumes in adults using a 14 G or 2-mm ID insufflation catheter at 15 L/min, demonstrating its potential for ventilation in CICO events.

使用改良的 Rapid-O2 氧气充气装置进行经导管通气。
背景:Rapid-O2 氧气充注装置®(Rapid-O2)主要用于无法插管、无法吸氧(CICO)情况下的抢救性吸氧;因此,高碳酸血症不可避免。对 Rapid-O2 进行了改进,以便在呼气时利用文丘里效应加强通气:为了确定内导管(20 G、18 G、16 G、14 G 和 2 毫米内径的经气管导管 [TTC])和充气导管(16 G、14 G 和 2 毫米内径的经气管导管 [TTC])的最有效组合,以达到最佳通气效果,在氧气流速为 15 升/分钟时测量了充气和呼气流量。充气和呼气压力以 6-15 L/min 的速度测量。流量和压力使用气体流量分析仪进行测量。充气和呼气时间使用气管-肺模型进行测量,以获得分钟容量。为了评估改进 Rapid-O2 后的效果,使用 Rapid-O2 测量了分钟容量:最合适的内导管为 18 G。在 15 升/分钟的速度下,充气压力从 97(2 毫米内径 TTC)到 377 cmH2O(16 G)不等。呼气时,以 15 升/分钟的速度充气导管测得的负压类似,均为 50 cmH2O。在肺顺应性为 100 毫升/厘米水时,15 升/分钟时通过内径为 2 毫米和 14 G 的充气导管的分钟流量分别为 7.0 升/分钟和 5.37 升/分钟。结论:结论:使用 14 G 或 2 毫米内径充气导管,在 15 升/分钟的速度下,改良型 Rapid-O2 能为成人提供足够的分钟流量,这证明了其在 CICO 事件中的通气潜力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
6.20
自引率
6.90%
发文量
84
审稿时长
16 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信