Ulrich Kaiser, Fabian Kück, Christian Stroszczynski, Ernst Michael Jung, Barbara Greiner
{"title":"Handheld ultrasound for portal vein evaluation in pediatric patients.","authors":"Ulrich Kaiser, Fabian Kück, Christian Stroszczynski, Ernst Michael Jung, Barbara Greiner","doi":"10.3233/CH-248105","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To assess diagnostic capabilities of a modern handheld ultrasound device (HUD) for portal vein (PV) evaluation in pediatric patients and to verify if age, body mass index (BMI) and object depth correlate to imagequality.</p><p><strong>Methods: </strong>45 patients (3.9 months- 17.9 years; 10.3 years±5.4) were examined using a HUD and cart-based high-end ultrasound system (HEUS). The intra-and extrahepatic PV was scanned using B-Mode and Color-coded Doppler sonography (CCDS). A five-point Likert scale was applied to evaluate image quality. Results were interpreted by two readers in consensus. Scores were compared between the scanners and correlated to age, BMI and skin-to-portal-vein-distance (SPVD). The influence on image quality and the difference between the two devices were analyzed using ordinal and Bayesian logistic regression models.</p><p><strong>Results: </strong>≥4 points (mild or no limitations) were achieved by the HUD in 60% for B-Mode and 56% for CCDS. There was a statistically significant negative correlation between the image quality achieved by the HUD and the SPVD for B-Mode and CCDS. The effect of BMI and SPVD on having≥4 points differed significantly between the HUD and HEUS for B-Mode and for CCDS.</p><p><strong>Conclusions: </strong>Sufficient to excellent image quality for portal vein assessment in pediatric patients was achieved by the HUD except for the most deep-lying structures. Compared to HEUS, diagnostic performance of the HUD is lower, based on a negative correlation with object depth.</p>","PeriodicalId":93943,"journal":{"name":"Clinical hemorheology and microcirculation","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical hemorheology and microcirculation","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3233/CH-248105","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: To assess diagnostic capabilities of a modern handheld ultrasound device (HUD) for portal vein (PV) evaluation in pediatric patients and to verify if age, body mass index (BMI) and object depth correlate to imagequality.
Methods: 45 patients (3.9 months- 17.9 years; 10.3 years±5.4) were examined using a HUD and cart-based high-end ultrasound system (HEUS). The intra-and extrahepatic PV was scanned using B-Mode and Color-coded Doppler sonography (CCDS). A five-point Likert scale was applied to evaluate image quality. Results were interpreted by two readers in consensus. Scores were compared between the scanners and correlated to age, BMI and skin-to-portal-vein-distance (SPVD). The influence on image quality and the difference between the two devices were analyzed using ordinal and Bayesian logistic regression models.
Results: ≥4 points (mild or no limitations) were achieved by the HUD in 60% for B-Mode and 56% for CCDS. There was a statistically significant negative correlation between the image quality achieved by the HUD and the SPVD for B-Mode and CCDS. The effect of BMI and SPVD on having≥4 points differed significantly between the HUD and HEUS for B-Mode and for CCDS.
Conclusions: Sufficient to excellent image quality for portal vein assessment in pediatric patients was achieved by the HUD except for the most deep-lying structures. Compared to HEUS, diagnostic performance of the HUD is lower, based on a negative correlation with object depth.