Clinical performance and application of novel serum collection “Ser-Col” device in the practice of laboratory diagnosis of infection diseases and several other immunochemical tests
Hayk Harutyunyan , Alvard Hovhannisyan , Hamlet Torosyan , Gohar Hakobjanyan , Ani Grigoryan , Gayane Petrosyan , Mariam Movsisyan , Dennis Poland , Richard Monkel , Joan Lommen , Edouard Tuaillon , Konstantin Yenkoyan
{"title":"Clinical performance and application of novel serum collection “Ser-Col” device in the practice of laboratory diagnosis of infection diseases and several other immunochemical tests","authors":"Hayk Harutyunyan , Alvard Hovhannisyan , Hamlet Torosyan , Gohar Hakobjanyan , Ani Grigoryan , Gayane Petrosyan , Mariam Movsisyan , Dennis Poland , Richard Monkel , Joan Lommen , Edouard Tuaillon , Konstantin Yenkoyan","doi":"10.1016/j.cca.2024.119970","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Dried blood collection devices might be beneficial for diagnosing infectious diseases in areas far from the medical facilities and in lockdown situations. There are several reports on the efficacy of such applications for qualitative tests. Here we demonstrated the feasibility of a novel Ser-Col blood collection device as a standardized approach for qualitative and quantitative detection of infectious markers and several over immunochemical tests.</div></div><div><h3>Methods</h3><div>In the current study, we included 395 adult participants, 191 men and 204 women, with a median age of 41 years, as well as 75 children with a median age of 3 years. Serological status was determined by testing serum samples for three groups of infection diseases: hepatitis A and C, SARS-CoV-2, and herpes family viruses, as well as for thyroid peroxidase (TPO), prolactin, vitamin B12, and folate. Blood collected on the Ser-Col device (Labonovum) was eluted using an automated system (SCAUT Ser-Col automation, Blok System Supply) and manually. Ser-Col results were compared with serum sampled via standard venipuncture considered as the reference.</div></div><div><h3>Results</h3><div>High correlation coefficients (r = 0.95–0.99) were observed between serum samples collected with Ser-Col and via standard venipuncture for the following tests: anti-HCV, anti-SARS-CoV-2 IgG, anti-HSV-2 IgG, and anti-CMV IgM. Correlation coefficients between Ser-Col and standard venipuncture serum for anti-HSV-1 IgG, anti-CMV IgG, and anti-EBV tests were relatively low (r = 0.73–0.77). Correlation coefficients for anti-TPO, prolactin, vitamin B12, and folate were also characterized with high values (r = 0.97–0.99).</div></div><div><h3>Conclusions</h3><div>High accuracy and quantitative correlation were demonstrated between Ser-Col and samples collected by standard venipuncture. Hence, the Ser-Col blood collection device should be considered as a promising alternative for blood collection, storage, and transportation in both adult and pediatric populations.</div></div>","PeriodicalId":10205,"journal":{"name":"Clinica Chimica Acta","volume":null,"pages":null},"PeriodicalIF":3.2000,"publicationDate":"2024-09-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinica Chimica Acta","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S000989812402223X","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Dried blood collection devices might be beneficial for diagnosing infectious diseases in areas far from the medical facilities and in lockdown situations. There are several reports on the efficacy of such applications for qualitative tests. Here we demonstrated the feasibility of a novel Ser-Col blood collection device as a standardized approach for qualitative and quantitative detection of infectious markers and several over immunochemical tests.
Methods
In the current study, we included 395 adult participants, 191 men and 204 women, with a median age of 41 years, as well as 75 children with a median age of 3 years. Serological status was determined by testing serum samples for three groups of infection diseases: hepatitis A and C, SARS-CoV-2, and herpes family viruses, as well as for thyroid peroxidase (TPO), prolactin, vitamin B12, and folate. Blood collected on the Ser-Col device (Labonovum) was eluted using an automated system (SCAUT Ser-Col automation, Blok System Supply) and manually. Ser-Col results were compared with serum sampled via standard venipuncture considered as the reference.
Results
High correlation coefficients (r = 0.95–0.99) were observed between serum samples collected with Ser-Col and via standard venipuncture for the following tests: anti-HCV, anti-SARS-CoV-2 IgG, anti-HSV-2 IgG, and anti-CMV IgM. Correlation coefficients between Ser-Col and standard venipuncture serum for anti-HSV-1 IgG, anti-CMV IgG, and anti-EBV tests were relatively low (r = 0.73–0.77). Correlation coefficients for anti-TPO, prolactin, vitamin B12, and folate were also characterized with high values (r = 0.97–0.99).
Conclusions
High accuracy and quantitative correlation were demonstrated between Ser-Col and samples collected by standard venipuncture. Hence, the Ser-Col blood collection device should be considered as a promising alternative for blood collection, storage, and transportation in both adult and pediatric populations.
期刊介绍:
The Official Journal of the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)
Clinica Chimica Acta is a high-quality journal which publishes original Research Communications in the field of clinical chemistry and laboratory medicine, defined as the diagnostic application of chemistry, biochemistry, immunochemistry, biochemical aspects of hematology, toxicology, and molecular biology to the study of human disease in body fluids and cells.
The objective of the journal is to publish novel information leading to a better understanding of biological mechanisms of human diseases, their prevention, diagnosis, and patient management. Reports of an applied clinical character are also welcome. Papers concerned with normal metabolic processes or with constituents of normal cells or body fluids, such as reports of experimental or clinical studies in animals, are only considered when they are clearly and directly relevant to human disease. Evaluation of commercial products have a low priority for publication, unless they are novel or represent a technological breakthrough. Studies dealing with effects of drugs and natural products and studies dealing with the redox status in various diseases are not within the journal''s scope. Development and evaluation of novel analytical methodologies where applicable to diagnostic clinical chemistry and laboratory medicine, including point-of-care testing, and topics on laboratory management and informatics will also be considered. Studies focused on emerging diagnostic technologies and (big) data analysis procedures including digitalization, mobile Health, and artificial Intelligence applied to Laboratory Medicine are also of interest.