Significance and feasibility of air kerma length product and air kerma area product comparisons

IF 1.6 3区 物理与天体物理 Q2 NUCLEAR SCIENCE & TECHNOLOGY
Miloš Živanović , Maja Vojnić Kortmiš , Nikola Kržanović , Miloš Đaletić , Ivana Komatina
{"title":"Significance and feasibility of air kerma length product and air kerma area product comparisons","authors":"Miloš Živanović ,&nbsp;Maja Vojnić Kortmiš ,&nbsp;Nikola Kržanović ,&nbsp;Miloš Đaletić ,&nbsp;Ivana Komatina","doi":"10.1016/j.radmeas.2024.107291","DOIUrl":null,"url":null,"abstract":"<div><p>Air kerma length product and air kerma area product are special quantities used in diagnostic radiology. They are measured using special measurement devices – CT-chambers and KAP-meters, in order to calculate quantities related to patient exposure. Appropriate calibration of all measurement devices is of vital importance, and comparisons between calibration laboratories are necessary to prove competence.</p><p>It is usually considered adequate to participate in air kerma comparisons to prove capabilities for special quantities, but the literature shows that some problems in calibration procedure can remain unknown. The comparisons directly in terms of special quantities provide additional burden to laboratories, and require special transfer instruments, but they allow checking the whole calibration procedures.</p><p>This paper describes a comparison between two calibration laboratories in terms of both air kerma length product and air kerma area product. Both laboratories achieved good results for all radiation qualities, considering the measurement uncertainty. Transfer instruments’ linearity, field size dependence and energy dependence were investigated. Even though the metrological properties of the transfer instruments are worse than the ionization chambers, they can be taken into account by introducing additional measurement uncertainty, performing appropriate corrections or choosing calibration points for the comparison for the values of influence quantities where the transfer instrument response is relatively flat. These comparisons provide additional value to calibration laboratories, but there are still several challenges related to their organization and execution.</p></div>","PeriodicalId":21055,"journal":{"name":"Radiation Measurements","volume":"178 ","pages":"Article 107291"},"PeriodicalIF":1.6000,"publicationDate":"2024-09-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation Measurements","FirstCategoryId":"101","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1350448724002397","RegionNum":3,"RegionCategory":"物理与天体物理","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"NUCLEAR SCIENCE & TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Air kerma length product and air kerma area product are special quantities used in diagnostic radiology. They are measured using special measurement devices – CT-chambers and KAP-meters, in order to calculate quantities related to patient exposure. Appropriate calibration of all measurement devices is of vital importance, and comparisons between calibration laboratories are necessary to prove competence.

It is usually considered adequate to participate in air kerma comparisons to prove capabilities for special quantities, but the literature shows that some problems in calibration procedure can remain unknown. The comparisons directly in terms of special quantities provide additional burden to laboratories, and require special transfer instruments, but they allow checking the whole calibration procedures.

This paper describes a comparison between two calibration laboratories in terms of both air kerma length product and air kerma area product. Both laboratories achieved good results for all radiation qualities, considering the measurement uncertainty. Transfer instruments’ linearity, field size dependence and energy dependence were investigated. Even though the metrological properties of the transfer instruments are worse than the ionization chambers, they can be taken into account by introducing additional measurement uncertainty, performing appropriate corrections or choosing calibration points for the comparison for the values of influence quantities where the transfer instrument response is relatively flat. These comparisons provide additional value to calibration laboratories, but there are still several challenges related to their organization and execution.

空气气膜长度乘积和空气气膜面积乘积比较的意义和可行性
空气介质长度积和空气介质面积积是放射诊断中使用的特殊数量。它们通过特殊的测量设备--CT-腔室和 KAP 计进行测量,以计算与患者照射有关的数量。对所有测量设备进行适当的校准至关重要,校准实验室之间的比较是证明能力的必要手段。本文介绍了两个校准实验室在空气热玛长度积和空气热玛面积积方面的比较。考虑到测量的不确定性,两个实验室在所有辐射质量方面都取得了良好的结果。对转移仪器的线性、场大小依赖性和能量依赖性进行了研究。尽管转移仪器的计量特性比电离室差,但可以通过引入额外的测量不确定性、进行适当的修正或为转移仪器响应相对平缓的影响量值的比较选择校准点来考虑这些特性。这些比对为校准实验室提供了额外的价值,但在其组织和执行方面仍存在一些挑战。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Radiation Measurements
Radiation Measurements 工程技术-核科学技术
CiteScore
4.10
自引率
20.00%
发文量
116
审稿时长
48 days
期刊介绍: The journal seeks to publish papers that present advances in the following areas: spontaneous and stimulated luminescence (including scintillating materials, thermoluminescence, and optically stimulated luminescence); electron spin resonance of natural and synthetic materials; the physics, design and performance of radiation measurements (including computational modelling such as electronic transport simulations); the novel basic aspects of radiation measurement in medical physics. Studies of energy-transfer phenomena, track physics and microdosimetry are also of interest to the journal. Applications relevant to the journal, particularly where they present novel detection techniques, novel analytical approaches or novel materials, include: personal dosimetry (including dosimetric quantities, active/electronic and passive monitoring techniques for photon, neutron and charged-particle exposures); environmental dosimetry (including methodological advances and predictive models related to radon, but generally excluding local survey results of radon where the main aim is to establish the radiation risk to populations); cosmic and high-energy radiation measurements (including dosimetry, space radiation effects, and single event upsets); dosimetry-based archaeological and Quaternary dating; dosimetry-based approaches to thermochronometry; accident and retrospective dosimetry (including activation detectors), and dosimetry and measurements related to medical applications.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信