Mark S Goodin, Chihiro Miyagi, Barry D Kuban, Christine R Flick, Anthony R Polakowski, Jamshid H Karimov, Kiyotaka Fukamachi
{"title":"Improving hydraulic performance of the left atrial assist device using computational fluid dynamics.","authors":"Mark S Goodin, Chihiro Miyagi, Barry D Kuban, Christine R Flick, Anthony R Polakowski, Jamshid H Karimov, Kiyotaka Fukamachi","doi":"10.1111/aor.14850","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The left atrial assist device (LAAD) is a novel continuous-flow pump designed to treat patients with heart failure with preserved ejection fraction, a growing type of heart failure, but with limited device-treatment options. The LAAD is implanted in the mitral plane and pumps blood from the left atrium into the left ventricle. The purpose of this study was to refine the initial design of the LAAD, using results from computational fluid dynamics (CFD) analyses to inform changes that could improve hydraulic performance and flow patterns within the LAAD.</p><p><strong>Methods: </strong>The initial design and three variations were simulated, exploring changes to the primary impeller blades, the housing shape, and the number, size, and curvature of the diffuser vanes. Several pump rotational speeds and flow rates spanning the intended range of use were modeled.</p><p><strong>Results: </strong>Guided by the insight gained from each design iteration, the final design incorporated impeller blades with improved alignment relative to the incoming flow and wider, more curved diffuser vanes that better aligned with the approaching flow from the volute. These design adjustments reduced flow separation within the impeller and diffuser regions. In vitro testing confirmed the CFD predicted improvement in the hydraulic performance of the revised LAAD flow path design.</p><p><strong>Conclusions: </strong>The CFD results from this study provided insight into the key pump design-related parameters that can be adjusted to improve the LAAD's hydraulic performance and internal flow patterns. This work also provided a foundation for future studies assessing the LAAD's biocompatibility under clinical conditions.</p>","PeriodicalId":8450,"journal":{"name":"Artificial organs","volume":null,"pages":null},"PeriodicalIF":2.2000,"publicationDate":"2024-09-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Artificial organs","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/aor.14850","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The left atrial assist device (LAAD) is a novel continuous-flow pump designed to treat patients with heart failure with preserved ejection fraction, a growing type of heart failure, but with limited device-treatment options. The LAAD is implanted in the mitral plane and pumps blood from the left atrium into the left ventricle. The purpose of this study was to refine the initial design of the LAAD, using results from computational fluid dynamics (CFD) analyses to inform changes that could improve hydraulic performance and flow patterns within the LAAD.
Methods: The initial design and three variations were simulated, exploring changes to the primary impeller blades, the housing shape, and the number, size, and curvature of the diffuser vanes. Several pump rotational speeds and flow rates spanning the intended range of use were modeled.
Results: Guided by the insight gained from each design iteration, the final design incorporated impeller blades with improved alignment relative to the incoming flow and wider, more curved diffuser vanes that better aligned with the approaching flow from the volute. These design adjustments reduced flow separation within the impeller and diffuser regions. In vitro testing confirmed the CFD predicted improvement in the hydraulic performance of the revised LAAD flow path design.
Conclusions: The CFD results from this study provided insight into the key pump design-related parameters that can be adjusted to improve the LAAD's hydraulic performance and internal flow patterns. This work also provided a foundation for future studies assessing the LAAD's biocompatibility under clinical conditions.
期刊介绍:
Artificial Organs is the official peer reviewed journal of The International Federation for Artificial Organs (Members of the Federation are: The American Society for Artificial Internal Organs, The European Society for Artificial Organs, and The Japanese Society for Artificial Organs), The International Faculty for Artificial Organs, the International Society for Rotary Blood Pumps, The International Society for Pediatric Mechanical Cardiopulmonary Support, and the Vienna International Workshop on Functional Electrical Stimulation. Artificial Organs publishes original research articles dealing with developments in artificial organs applications and treatment modalities and their clinical applications worldwide. Membership in the Societies listed above is not a prerequisite for publication. Articles are published without charge to the author except for color figures and excess page charges as noted.