MARPE expander activation load with different configurations of extender arms heights: in-vitro evaluation.

Q2 Medicine
Dental Press Journal of Orthodontics Pub Date : 2024-09-02 eCollection Date: 2024-01-01 DOI:10.1590/2177-6709.29.4.e242458.oar
Staell Ribeiro de Faria, Túlio Rodrigues de Andrade, Cristiane Barros André, Victor Angelo Martins Montalli, Jurandir Antonio Barbosa, Roberta Tarkany Basting
{"title":"MARPE expander activation load with different configurations of extender arms heights: in-vitro evaluation.","authors":"Staell Ribeiro de Faria, Túlio Rodrigues de Andrade, Cristiane Barros André, Victor Angelo Martins Montalli, Jurandir Antonio Barbosa, Roberta Tarkany Basting","doi":"10.1590/2177-6709.29.4.e242458.oar","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Evaluate the load mini-implants exert on the artificial bone when expanding the MARPE EX in three different extension arm configurations.</p><p><strong>Methods: </strong>A device simulating the human palate was fabricated and attached to a universal testing machine, for conducting tests with different MARPE expanders (n=5): non-adjustable/control (MARPE SL, Peclab) or with low, intermediate, and high extender arms (MARPE EX, Peclab). The expanders were manually activated until failure of the device occurred, and maximum load values were recorded. Load averages were also calculated for every five activations until the twentieth activation.</p><p><strong>Results: </strong>The generalized linear mixed model for repeated measures over time showed that there was significant increase in load with activations for all expanders (p=0.0004). Up to the twentieth activation, the expander with low extender arms presented higher load than the others, while the expander with high extender arms showed lower load values (p<0.05). There was no significant difference among expanders regarding the number of activations (p=0.0586), although there was a trend towards fewer activations until fracture for the control expander. It was observed that the higher the configuration, the lower the force the mini-implants delivered to the bone. The control expander provided a force magnitude similar to that of the adjustable expander when positioned at the intermediate height.</p><p><strong>Conclusions: </strong>The activation load of MARPE expanders is influenced by the type of presentation of the extensor arms, with higher configurations resulting in lower force delivered by the mini-implants to the bone.</p>","PeriodicalId":38720,"journal":{"name":"Dental Press Journal of Orthodontics","volume":"29 4","pages":"e242458"},"PeriodicalIF":0.0000,"publicationDate":"2024-09-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11368235/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Dental Press Journal of Orthodontics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1590/2177-6709.29.4.e242458.oar","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: Evaluate the load mini-implants exert on the artificial bone when expanding the MARPE EX in three different extension arm configurations.

Methods: A device simulating the human palate was fabricated and attached to a universal testing machine, for conducting tests with different MARPE expanders (n=5): non-adjustable/control (MARPE SL, Peclab) or with low, intermediate, and high extender arms (MARPE EX, Peclab). The expanders were manually activated until failure of the device occurred, and maximum load values were recorded. Load averages were also calculated for every five activations until the twentieth activation.

Results: The generalized linear mixed model for repeated measures over time showed that there was significant increase in load with activations for all expanders (p=0.0004). Up to the twentieth activation, the expander with low extender arms presented higher load than the others, while the expander with high extender arms showed lower load values (p<0.05). There was no significant difference among expanders regarding the number of activations (p=0.0586), although there was a trend towards fewer activations until fracture for the control expander. It was observed that the higher the configuration, the lower the force the mini-implants delivered to the bone. The control expander provided a force magnitude similar to that of the adjustable expander when positioned at the intermediate height.

Conclusions: The activation load of MARPE expanders is influenced by the type of presentation of the extensor arms, with higher configurations resulting in lower force delivered by the mini-implants to the bone.

MARPE扩张器的激活负荷与不同配置的扩张臂高度:体外评估。
目的:评估微型植入物在三种不同的扩展臂配置下扩展 MARPE EX 时对人工骨产生的负荷:评估MARPE EX以三种不同的扩展臂配置扩展时,微型植入物对人工骨产生的负荷:制作了一个模拟人腭的装置,并将其连接到万能试验机上,用不同的MARPE扩张器(n=5)进行测试:不可调/控制(MARPE SL,Peclab公司)或低、中、高扩展臂(MARPE EX,Peclab公司)。手动启动扩张器,直至装置失效,并记录最大负荷值。在第 20 次启动之前,每启动 5 次还会计算负荷平均值:重复测量时间的广义线性混合模型显示,所有扩张器的负荷量都随着激活次数的增加而显著增加(p=0.0004)。在第 20 次激活之前,伸展臂低的扩张器的负荷高于其他扩张器,而伸展臂高的扩张器的负荷值则较低(p结论:MARPE 扩张器的激活负荷与其他扩张器的负荷有明显差异:MARPE扩张器的激活负荷受伸展臂类型的影响,配置越高,微型植入体传递到骨的力越小。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Dental Press Journal of Orthodontics
Dental Press Journal of Orthodontics Dentistry-Orthodontics
CiteScore
2.00
自引率
0.00%
发文量
40
审稿时长
27 weeks
期刊介绍: The Dental Press Journal of Orthodontics publishes scientific research articles, significant reviews, clinical and technical case reports, brief communications, and other materials related to Orthodontics and Facial Orthopedics.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信